Cargando…

Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India

Objectives This study was performed to estimate diagnostic accuracy of the two commercially available point-of-care tests to identify poor glycemic control defined by HbA1c levels, with high-performance liquid chromatography (HPLC) as a reference. Settings The study was carried at two locations, gen...

Descripción completa

Detalles Bibliográficos
Autores principales: Khadanga, Sagar, Singh, Gyanendra, Pakhare, Abhijit P, Joshi, Rajnish
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8513731/
https://www.ncbi.nlm.nih.gov/pubmed/34660113
http://dx.doi.org/10.7759/cureus.17920
_version_ 1784583261921476608
author Khadanga, Sagar
Singh, Gyanendra
Pakhare, Abhijit P
Joshi, Rajnish
author_facet Khadanga, Sagar
Singh, Gyanendra
Pakhare, Abhijit P
Joshi, Rajnish
author_sort Khadanga, Sagar
collection PubMed
description Objectives This study was performed to estimate diagnostic accuracy of the two commercially available point-of-care tests to identify poor glycemic control defined by HbA1c levels, with high-performance liquid chromatography (HPLC) as a reference. Settings The study was carried at two locations, general medical outpatient department of a teaching medical college in Bhopal (urban), and a primary health care centre in a rural area in the state of Madhya Pradesh, India. Participants All individuals with diabetes mellitus who presented to the health care facility for assessment of glycemic control. We compared HbA1c estimated from two index tests (Hemocue Hb 501, Sweden; SD Biosensor, South Korea) from capillary blood samples with HPLC performed from venous blood, as a reference standard. Primary and secondary outcome measures Diagnostic properties of index tests such as sensitivity, specificity, positive and negative predictive value and diagnostic accuracy for identifying poor glycemic control were primary outcome measures. Lin’s concordance correlation coefficient (CCC) was secondary outcome measure. Results Out of 114 patients, all received reference standard - 103 patients received Hemocue A1C test and 110 patients received SD Biosensor test. Overall both the index tests had similar diagnostic accuracy estimates. The area under the Receiver Operating Curve for SDA1c device was 0.935 (95% CI 0.886-0.983), and for Hemocue device was 0.938 (95% CI 0.893-0.984). The Hemocue device HbA1c value of above 7.0 (positive) correctly predicted poor glycemic control 92% times (81.58% for SD device). There were 4 vs. 11 device failures and 14 vs. 12 failures with SD and Hemocue, respectively. Ambient air temperatures were no different for the device test failures. Conclusions Commercially available point-of-care tests evaluated in this study are comparable and an acceptable alternative to HPLC-based measurements for the assessment of glycemic control. Tests and device failure rates of both the index tests are similar.
format Online
Article
Text
id pubmed-8513731
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-85137312021-10-15 Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India Khadanga, Sagar Singh, Gyanendra Pakhare, Abhijit P Joshi, Rajnish Cureus Endocrinology/Diabetes/Metabolism Objectives This study was performed to estimate diagnostic accuracy of the two commercially available point-of-care tests to identify poor glycemic control defined by HbA1c levels, with high-performance liquid chromatography (HPLC) as a reference. Settings The study was carried at two locations, general medical outpatient department of a teaching medical college in Bhopal (urban), and a primary health care centre in a rural area in the state of Madhya Pradesh, India. Participants All individuals with diabetes mellitus who presented to the health care facility for assessment of glycemic control. We compared HbA1c estimated from two index tests (Hemocue Hb 501, Sweden; SD Biosensor, South Korea) from capillary blood samples with HPLC performed from venous blood, as a reference standard. Primary and secondary outcome measures Diagnostic properties of index tests such as sensitivity, specificity, positive and negative predictive value and diagnostic accuracy for identifying poor glycemic control were primary outcome measures. Lin’s concordance correlation coefficient (CCC) was secondary outcome measure. Results Out of 114 patients, all received reference standard - 103 patients received Hemocue A1C test and 110 patients received SD Biosensor test. Overall both the index tests had similar diagnostic accuracy estimates. The area under the Receiver Operating Curve for SDA1c device was 0.935 (95% CI 0.886-0.983), and for Hemocue device was 0.938 (95% CI 0.893-0.984). The Hemocue device HbA1c value of above 7.0 (positive) correctly predicted poor glycemic control 92% times (81.58% for SD device). There were 4 vs. 11 device failures and 14 vs. 12 failures with SD and Hemocue, respectively. Ambient air temperatures were no different for the device test failures. Conclusions Commercially available point-of-care tests evaluated in this study are comparable and an acceptable alternative to HPLC-based measurements for the assessment of glycemic control. Tests and device failure rates of both the index tests are similar. Cureus 2021-09-13 /pmc/articles/PMC8513731/ /pubmed/34660113 http://dx.doi.org/10.7759/cureus.17920 Text en Copyright © 2021, Khadanga et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Endocrinology/Diabetes/Metabolism
Khadanga, Sagar
Singh, Gyanendra
Pakhare, Abhijit P
Joshi, Rajnish
Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India
title Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India
title_full Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India
title_fullStr Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India
title_full_unstemmed Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India
title_short Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India
title_sort diagnostic accuracy of point-of-care tests measuring glycosylated haemoglobin (hba1c) for glycemic control: a field study in india
topic Endocrinology/Diabetes/Metabolism
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8513731/
https://www.ncbi.nlm.nih.gov/pubmed/34660113
http://dx.doi.org/10.7759/cureus.17920
work_keys_str_mv AT khadangasagar diagnosticaccuracyofpointofcaretestsmeasuringglycosylatedhaemoglobinhba1cforglycemiccontrolafieldstudyinindia
AT singhgyanendra diagnosticaccuracyofpointofcaretestsmeasuringglycosylatedhaemoglobinhba1cforglycemiccontrolafieldstudyinindia
AT pakhareabhijitp diagnosticaccuracyofpointofcaretestsmeasuringglycosylatedhaemoglobinhba1cforglycemiccontrolafieldstudyinindia
AT joshirajnish diagnosticaccuracyofpointofcaretestsmeasuringglycosylatedhaemoglobinhba1cforglycemiccontrolafieldstudyinindia