Cargando…
The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants
Background and purpose — We noted several adverse events in patients in whom the first version of the STRYDE limb-lengthening nail (NuVasive Specialized Orthopaedics, San Diego, CA) had been implanted. Pain, osteolysis, periosteal reactions, and cortical hypertrophy at the nail junction were noted....
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Taylor & Francis
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8519523/ https://www.ncbi.nlm.nih.gov/pubmed/34102950 http://dx.doi.org/10.1080/17453674.2021.1927506 |
_version_ | 1784584467828965376 |
---|---|
author | Jellesen, Morten Stendahl Lomholt, Trine Nybo Hansen, Rikke Quist Mathiesen, Troels Gundlach, Carsten Kold, Søren Nygaard, Tobias Mikuzis, Mindaugas Olesen, Ulrik Kähler Rölfing, Jan Duedal |
author_facet | Jellesen, Morten Stendahl Lomholt, Trine Nybo Hansen, Rikke Quist Mathiesen, Troels Gundlach, Carsten Kold, Søren Nygaard, Tobias Mikuzis, Mindaugas Olesen, Ulrik Kähler Rölfing, Jan Duedal |
author_sort | Jellesen, Morten Stendahl |
collection | PubMed |
description | Background and purpose — We noted several adverse events in patients in whom the first version of the STRYDE limb-lengthening nail (NuVasive Specialized Orthopaedics, San Diego, CA) had been implanted. Pain, osteolysis, periosteal reactions, and cortical hypertrophy at the nail junction were noted. Here, we present the analysis of 23 retrieved STRYDE implants. Materials and methods — We undertook visual inspection of the retrieved nails and screws, mechanical evaluation of the junction, micro-CT analyses, microscopic inspection of the bushing, screws, screw holes, and separated parts of the implants. Positive material identification (PMI) and energy-dispersive X-ray spectroscopy (EDS) were used to analyze the chemical composition. The hardness of the material was also investigated. Results — 20/23 retrieved nails had visible signs of corrosion, i.e., discoloration at the telescopic junction. Micro-CT verified corrosion attacks in 12/12 scanned bushings. Corrosion, predominantly mechanically assisted crevice corrosion, was observed at the locking screws and screw holes in 20/23 nails. Biological material inside the nail was observed in addition to oozing from the junction of 2 nails during hardware removal, which was experimentally reproducible. Notably, the mechanical construction of the bushing changed from PRECICE P2 to STRYDE nails. Interpretation — STRYDE nails are not hermetically sealed, and liquid can pass the bushing. Biodur 108 itself is corrosion resistant; however, mechanically assisted crevice corrosion of the bushing, locking screws, and screw holes may be aggravated due to manufacturing aiming for increased strength and hardness of the alloy. Observing several adverse events, we recently published a nationwide cross-sectional analysis of all 30 STRYDE limb- lengthening nails (NuVasive, Specialized Orthopedics, San Diego, CA) that were implanted in Denmark (Rölfing et al. 2021a). 27/30 STRYDE nails have now been removed and we present data from metallurgical analysis of 23 of the retrieved implants. |
format | Online Article Text |
id | pubmed-8519523 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Taylor & Francis |
record_format | MEDLINE/PubMed |
spelling | pubmed-85195232021-10-16 The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants Jellesen, Morten Stendahl Lomholt, Trine Nybo Hansen, Rikke Quist Mathiesen, Troels Gundlach, Carsten Kold, Søren Nygaard, Tobias Mikuzis, Mindaugas Olesen, Ulrik Kähler Rölfing, Jan Duedal Acta Orthop Research Article Background and purpose — We noted several adverse events in patients in whom the first version of the STRYDE limb-lengthening nail (NuVasive Specialized Orthopaedics, San Diego, CA) had been implanted. Pain, osteolysis, periosteal reactions, and cortical hypertrophy at the nail junction were noted. Here, we present the analysis of 23 retrieved STRYDE implants. Materials and methods — We undertook visual inspection of the retrieved nails and screws, mechanical evaluation of the junction, micro-CT analyses, microscopic inspection of the bushing, screws, screw holes, and separated parts of the implants. Positive material identification (PMI) and energy-dispersive X-ray spectroscopy (EDS) were used to analyze the chemical composition. The hardness of the material was also investigated. Results — 20/23 retrieved nails had visible signs of corrosion, i.e., discoloration at the telescopic junction. Micro-CT verified corrosion attacks in 12/12 scanned bushings. Corrosion, predominantly mechanically assisted crevice corrosion, was observed at the locking screws and screw holes in 20/23 nails. Biological material inside the nail was observed in addition to oozing from the junction of 2 nails during hardware removal, which was experimentally reproducible. Notably, the mechanical construction of the bushing changed from PRECICE P2 to STRYDE nails. Interpretation — STRYDE nails are not hermetically sealed, and liquid can pass the bushing. Biodur 108 itself is corrosion resistant; however, mechanically assisted crevice corrosion of the bushing, locking screws, and screw holes may be aggravated due to manufacturing aiming for increased strength and hardness of the alloy. Observing several adverse events, we recently published a nationwide cross-sectional analysis of all 30 STRYDE limb- lengthening nails (NuVasive, Specialized Orthopedics, San Diego, CA) that were implanted in Denmark (Rölfing et al. 2021a). 27/30 STRYDE nails have now been removed and we present data from metallurgical analysis of 23 of the retrieved implants. Taylor & Francis 2021-06-08 /pmc/articles/PMC8519523/ /pubmed/34102950 http://dx.doi.org/10.1080/17453674.2021.1927506 Text en © 2021 The Author(s). Published by Taylor & Francis on behalf of the Nordic Orthopedic Federation. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Jellesen, Morten Stendahl Lomholt, Trine Nybo Hansen, Rikke Quist Mathiesen, Troels Gundlach, Carsten Kold, Søren Nygaard, Tobias Mikuzis, Mindaugas Olesen, Ulrik Kähler Rölfing, Jan Duedal The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants |
title | The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants |
title_full | The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants |
title_fullStr | The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants |
title_full_unstemmed | The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants |
title_short | The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants |
title_sort | stryde limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8519523/ https://www.ncbi.nlm.nih.gov/pubmed/34102950 http://dx.doi.org/10.1080/17453674.2021.1927506 |
work_keys_str_mv | AT jellesenmortenstendahl thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT lomholttrinenybo thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT hansenrikkequist thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT mathiesentroels thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT gundlachcarsten thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT koldsøren thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT nygaardtobias thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT mikuzismindaugas thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT olesenulrikkahler thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT rolfingjanduedal thestrydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT jellesenmortenstendahl strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT lomholttrinenybo strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT hansenrikkequist strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT mathiesentroels strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT gundlachcarsten strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT koldsøren strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT nygaardtobias strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT mikuzismindaugas strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT olesenulrikkahler strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants AT rolfingjanduedal strydelimblengtheningnailissusceptibletomechanicallyassistedcrevicecorrosionananalysisof23retrievedimplants |