Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip

Existing methods for detecting cardiac markers are difficult to be applied in point-of-care testing (POCT) due to complex operation, long time consumption, and low sensitivity. Here, we report a snail-shaped microfluidic chip (SMC) for the multiplex detection of cTnI, CK-MB, and Myo with high sensit...

Descripción completa

Detalles Bibliográficos
Autores principales: Yin, Binfeng, Wan, Xinhua, Qian, Changcheng, Sohan, A. S. M. Muhtasim Fuad, Wang, Songbai, Zhou, Teng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8521045/
https://www.ncbi.nlm.nih.gov/pubmed/34671590
http://dx.doi.org/10.3389/fchem.2021.741058
_version_ 1784584815611215872
author Yin, Binfeng
Wan, Xinhua
Qian, Changcheng
Sohan, A. S. M. Muhtasim Fuad
Wang, Songbai
Zhou, Teng
author_facet Yin, Binfeng
Wan, Xinhua
Qian, Changcheng
Sohan, A. S. M. Muhtasim Fuad
Wang, Songbai
Zhou, Teng
author_sort Yin, Binfeng
collection PubMed
description Existing methods for detecting cardiac markers are difficult to be applied in point-of-care testing (POCT) due to complex operation, long time consumption, and low sensitivity. Here, we report a snail-shaped microfluidic chip (SMC) for the multiplex detection of cTnI, CK-MB, and Myo with high sensitivity and a short detection time. The SMC consists of a sandwich structure: a channel layer with a mixer and reaction zone, a reaction layer coated with capture antibodies, and a base layer. The opening or closing of the microchannels is realized by controlling the downward movement of the press-type mechanical valve. The chemiluminescence method was used as a signal readout, and the experimental conditions were optimized. SMC could detect cTnI, CK-MB, and Myo at concentrations as low as 1.02, 1.37, and 4.15. The SMC will be a promising platform for a simultaneous determination of multianalytes and shows a potential application in POCT.
format Online
Article
Text
id pubmed-8521045
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-85210452021-10-19 Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip Yin, Binfeng Wan, Xinhua Qian, Changcheng Sohan, A. S. M. Muhtasim Fuad Wang, Songbai Zhou, Teng Front Chem Chemistry Existing methods for detecting cardiac markers are difficult to be applied in point-of-care testing (POCT) due to complex operation, long time consumption, and low sensitivity. Here, we report a snail-shaped microfluidic chip (SMC) for the multiplex detection of cTnI, CK-MB, and Myo with high sensitivity and a short detection time. The SMC consists of a sandwich structure: a channel layer with a mixer and reaction zone, a reaction layer coated with capture antibodies, and a base layer. The opening or closing of the microchannels is realized by controlling the downward movement of the press-type mechanical valve. The chemiluminescence method was used as a signal readout, and the experimental conditions were optimized. SMC could detect cTnI, CK-MB, and Myo at concentrations as low as 1.02, 1.37, and 4.15. The SMC will be a promising platform for a simultaneous determination of multianalytes and shows a potential application in POCT. Frontiers Media S.A. 2021-10-04 /pmc/articles/PMC8521045/ /pubmed/34671590 http://dx.doi.org/10.3389/fchem.2021.741058 Text en Copyright © 2021 Yin, Wan, Qian, Sohan, Wang and Zhou. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Chemistry
Yin, Binfeng
Wan, Xinhua
Qian, Changcheng
Sohan, A. S. M. Muhtasim Fuad
Wang, Songbai
Zhou, Teng
Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip
title Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip
title_full Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip
title_fullStr Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip
title_full_unstemmed Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip
title_short Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip
title_sort point-of-care testing for multiple cardiac markers based on a snail-shaped microfluidic chip
topic Chemistry
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8521045/
https://www.ncbi.nlm.nih.gov/pubmed/34671590
http://dx.doi.org/10.3389/fchem.2021.741058
work_keys_str_mv AT yinbinfeng pointofcaretestingformultiplecardiacmarkersbasedonasnailshapedmicrofluidicchip
AT wanxinhua pointofcaretestingformultiplecardiacmarkersbasedonasnailshapedmicrofluidicchip
AT qianchangcheng pointofcaretestingformultiplecardiacmarkersbasedonasnailshapedmicrofluidicchip
AT sohanasmmuhtasimfuad pointofcaretestingformultiplecardiacmarkersbasedonasnailshapedmicrofluidicchip
AT wangsongbai pointofcaretestingformultiplecardiacmarkersbasedonasnailshapedmicrofluidicchip
AT zhouteng pointofcaretestingformultiplecardiacmarkersbasedonasnailshapedmicrofluidicchip