Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand

BACKGROUND: A comparison study is crucial before launching a new medical device; therefore, we compared the Mission Ultra Hb Testing System with the Sysmex XN‐3000 automated hematology analyzer in Thai adult males and non‐pregnant adult females. METHODS: Parallel studies were conducted using discard...

Descripción completa

Detalles Bibliográficos
Autores principales: Chantkran, Wittawat, Jamnarnwej, Pitipat, Sritanabutr, Pipat, Arnutti, Pasra
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8529128/
https://www.ncbi.nlm.nih.gov/pubmed/34427961
http://dx.doi.org/10.1002/jcla.23976
_version_ 1784586402161229824
author Chantkran, Wittawat
Jamnarnwej, Pitipat
Sritanabutr, Pipat
Arnutti, Pasra
author_facet Chantkran, Wittawat
Jamnarnwej, Pitipat
Sritanabutr, Pipat
Arnutti, Pasra
author_sort Chantkran, Wittawat
collection PubMed
description BACKGROUND: A comparison study is crucial before launching a new medical device; therefore, we compared the Mission Ultra Hb Testing System with the Sysmex XN‐3000 automated hematology analyzer in Thai adult males and non‐pregnant adult females. METHODS: Parallel studies were conducted using discarded venous K2‐ethylenediaminetetraacetic acid samples from participants requiring hematological investigations. According to the World Health Organization criteria, the participants were categorized as overall, anemia, and non‐anemia for analysis. RESULTS: Three hundred participants were included in this study. In all participants, near‐perfect correlation and agreement were observed between the two methods for Hb measurement (r = 0.963, p < 0.001) with an interclass correlation coefficient (ICC) of 0.981 (95% confidence interval [CI]: 0.976–0.985) and Hct measurement (r = 0.941, p < 0.001) with an ICC of 0.965 (95% CI: 0.956–0.972). The sensitivity and specificity of the device in detecting anemia were 86.2% (95% CI: 79.7–91.2) and 98.6% (95% CI: 95.2–99.8), respectively. The area under the curve was 0.976 (95% CI: 0.963–0.989). The device showed average biases of 0.76 g/dl (95% limits of agreement [LOA]: −1.03 to 2.54) for Hb measurement and −2.73% (95% LOA: −9.28 to 3.82) for Hct measurement in all participants. CONCLUSION: Agreement between the Mission Ultra Hb Testing System and Sysmex XN‐3000 was observed. The device was excellent for detecting anemia. However, the essential evidence showing biases of the Hb and Hct measurements obtained from the device was revealed. Laboratory interpretation should be carefully performed, particularly at the near cut‐off values.
format Online
Article
Text
id pubmed-8529128
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-85291282021-10-27 Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand Chantkran, Wittawat Jamnarnwej, Pitipat Sritanabutr, Pipat Arnutti, Pasra J Clin Lab Anal Research Articles BACKGROUND: A comparison study is crucial before launching a new medical device; therefore, we compared the Mission Ultra Hb Testing System with the Sysmex XN‐3000 automated hematology analyzer in Thai adult males and non‐pregnant adult females. METHODS: Parallel studies were conducted using discarded venous K2‐ethylenediaminetetraacetic acid samples from participants requiring hematological investigations. According to the World Health Organization criteria, the participants were categorized as overall, anemia, and non‐anemia for analysis. RESULTS: Three hundred participants were included in this study. In all participants, near‐perfect correlation and agreement were observed between the two methods for Hb measurement (r = 0.963, p < 0.001) with an interclass correlation coefficient (ICC) of 0.981 (95% confidence interval [CI]: 0.976–0.985) and Hct measurement (r = 0.941, p < 0.001) with an ICC of 0.965 (95% CI: 0.956–0.972). The sensitivity and specificity of the device in detecting anemia were 86.2% (95% CI: 79.7–91.2) and 98.6% (95% CI: 95.2–99.8), respectively. The area under the curve was 0.976 (95% CI: 0.963–0.989). The device showed average biases of 0.76 g/dl (95% limits of agreement [LOA]: −1.03 to 2.54) for Hb measurement and −2.73% (95% LOA: −9.28 to 3.82) for Hct measurement in all participants. CONCLUSION: Agreement between the Mission Ultra Hb Testing System and Sysmex XN‐3000 was observed. The device was excellent for detecting anemia. However, the essential evidence showing biases of the Hb and Hct measurements obtained from the device was revealed. Laboratory interpretation should be carefully performed, particularly at the near cut‐off values. John Wiley and Sons Inc. 2021-08-24 /pmc/articles/PMC8529128/ /pubmed/34427961 http://dx.doi.org/10.1002/jcla.23976 Text en © 2021 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Chantkran, Wittawat
Jamnarnwej, Pitipat
Sritanabutr, Pipat
Arnutti, Pasra
Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand
title Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand
title_full Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand
title_fullStr Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand
title_full_unstemmed Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand
title_short Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand
title_sort evaluation of point‐of‐care testing device for anemia detection: a cross‐sectional method comparison study from thailand
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8529128/
https://www.ncbi.nlm.nih.gov/pubmed/34427961
http://dx.doi.org/10.1002/jcla.23976
work_keys_str_mv AT chantkranwittawat evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand
AT jamnarnwejpitipat evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand
AT sritanabutrpipat evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand
AT arnuttipasra evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand