Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand
BACKGROUND: A comparison study is crucial before launching a new medical device; therefore, we compared the Mission Ultra Hb Testing System with the Sysmex XN‐3000 automated hematology analyzer in Thai adult males and non‐pregnant adult females. METHODS: Parallel studies were conducted using discard...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8529128/ https://www.ncbi.nlm.nih.gov/pubmed/34427961 http://dx.doi.org/10.1002/jcla.23976 |
_version_ | 1784586402161229824 |
---|---|
author | Chantkran, Wittawat Jamnarnwej, Pitipat Sritanabutr, Pipat Arnutti, Pasra |
author_facet | Chantkran, Wittawat Jamnarnwej, Pitipat Sritanabutr, Pipat Arnutti, Pasra |
author_sort | Chantkran, Wittawat |
collection | PubMed |
description | BACKGROUND: A comparison study is crucial before launching a new medical device; therefore, we compared the Mission Ultra Hb Testing System with the Sysmex XN‐3000 automated hematology analyzer in Thai adult males and non‐pregnant adult females. METHODS: Parallel studies were conducted using discarded venous K2‐ethylenediaminetetraacetic acid samples from participants requiring hematological investigations. According to the World Health Organization criteria, the participants were categorized as overall, anemia, and non‐anemia for analysis. RESULTS: Three hundred participants were included in this study. In all participants, near‐perfect correlation and agreement were observed between the two methods for Hb measurement (r = 0.963, p < 0.001) with an interclass correlation coefficient (ICC) of 0.981 (95% confidence interval [CI]: 0.976–0.985) and Hct measurement (r = 0.941, p < 0.001) with an ICC of 0.965 (95% CI: 0.956–0.972). The sensitivity and specificity of the device in detecting anemia were 86.2% (95% CI: 79.7–91.2) and 98.6% (95% CI: 95.2–99.8), respectively. The area under the curve was 0.976 (95% CI: 0.963–0.989). The device showed average biases of 0.76 g/dl (95% limits of agreement [LOA]: −1.03 to 2.54) for Hb measurement and −2.73% (95% LOA: −9.28 to 3.82) for Hct measurement in all participants. CONCLUSION: Agreement between the Mission Ultra Hb Testing System and Sysmex XN‐3000 was observed. The device was excellent for detecting anemia. However, the essential evidence showing biases of the Hb and Hct measurements obtained from the device was revealed. Laboratory interpretation should be carefully performed, particularly at the near cut‐off values. |
format | Online Article Text |
id | pubmed-8529128 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-85291282021-10-27 Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand Chantkran, Wittawat Jamnarnwej, Pitipat Sritanabutr, Pipat Arnutti, Pasra J Clin Lab Anal Research Articles BACKGROUND: A comparison study is crucial before launching a new medical device; therefore, we compared the Mission Ultra Hb Testing System with the Sysmex XN‐3000 automated hematology analyzer in Thai adult males and non‐pregnant adult females. METHODS: Parallel studies were conducted using discarded venous K2‐ethylenediaminetetraacetic acid samples from participants requiring hematological investigations. According to the World Health Organization criteria, the participants were categorized as overall, anemia, and non‐anemia for analysis. RESULTS: Three hundred participants were included in this study. In all participants, near‐perfect correlation and agreement were observed between the two methods for Hb measurement (r = 0.963, p < 0.001) with an interclass correlation coefficient (ICC) of 0.981 (95% confidence interval [CI]: 0.976–0.985) and Hct measurement (r = 0.941, p < 0.001) with an ICC of 0.965 (95% CI: 0.956–0.972). The sensitivity and specificity of the device in detecting anemia were 86.2% (95% CI: 79.7–91.2) and 98.6% (95% CI: 95.2–99.8), respectively. The area under the curve was 0.976 (95% CI: 0.963–0.989). The device showed average biases of 0.76 g/dl (95% limits of agreement [LOA]: −1.03 to 2.54) for Hb measurement and −2.73% (95% LOA: −9.28 to 3.82) for Hct measurement in all participants. CONCLUSION: Agreement between the Mission Ultra Hb Testing System and Sysmex XN‐3000 was observed. The device was excellent for detecting anemia. However, the essential evidence showing biases of the Hb and Hct measurements obtained from the device was revealed. Laboratory interpretation should be carefully performed, particularly at the near cut‐off values. John Wiley and Sons Inc. 2021-08-24 /pmc/articles/PMC8529128/ /pubmed/34427961 http://dx.doi.org/10.1002/jcla.23976 Text en © 2021 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Articles Chantkran, Wittawat Jamnarnwej, Pitipat Sritanabutr, Pipat Arnutti, Pasra Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand |
title | Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand |
title_full | Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand |
title_fullStr | Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand |
title_full_unstemmed | Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand |
title_short | Evaluation of point‐of‐care testing device for anemia detection: A cross‐sectional method comparison study from Thailand |
title_sort | evaluation of point‐of‐care testing device for anemia detection: a cross‐sectional method comparison study from thailand |
topic | Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8529128/ https://www.ncbi.nlm.nih.gov/pubmed/34427961 http://dx.doi.org/10.1002/jcla.23976 |
work_keys_str_mv | AT chantkranwittawat evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand AT jamnarnwejpitipat evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand AT sritanabutrpipat evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand AT arnuttipasra evaluationofpointofcaretestingdeviceforanemiadetectionacrosssectionalmethodcomparisonstudyfromthailand |