Cargando…

Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures

BACKGROUND: Bispectral index (BIS) monitoring is a widely used non-invasive method to monitor the depth of anesthesia. However, in the event of surgeries requiring a frontal approach, placement of the electrode may be impossible at the designated area to achieve a proper BIS measurement. METHODS: We...

Descripción completa

Detalles Bibliográficos
Autores principales: Harada, Hideki, Muta, Seiya, Kakuma, Tatsuyuki, Ukeda, Misa, Ota, So, Hirata, Maiko, Fujioka, Hiroshi, Nakashima, Osamu, Dietel, Barbara, Tauchi, Miyuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8530286/
https://www.ncbi.nlm.nih.gov/pubmed/34673803
http://dx.doi.org/10.1371/journal.pone.0258647
_version_ 1784586638419034112
author Harada, Hideki
Muta, Seiya
Kakuma, Tatsuyuki
Ukeda, Misa
Ota, So
Hirata, Maiko
Fujioka, Hiroshi
Nakashima, Osamu
Dietel, Barbara
Tauchi, Miyuki
author_facet Harada, Hideki
Muta, Seiya
Kakuma, Tatsuyuki
Ukeda, Misa
Ota, So
Hirata, Maiko
Fujioka, Hiroshi
Nakashima, Osamu
Dietel, Barbara
Tauchi, Miyuki
author_sort Harada, Hideki
collection PubMed
description BACKGROUND: Bispectral index (BIS) monitoring is a widely used non-invasive method to monitor the depth of anesthesia. However, in the event of surgeries requiring a frontal approach, placement of the electrode may be impossible at the designated area to achieve a proper BIS measurement. METHODS: We developed an investigational interface device to connect needle-electrodes to BIS sensors. The safety and clinical performance were investigated in patients who underwent surgery. Direct BIS values from a disposable BIS electrode and indirect values via the interface device were simultaneously recorded from the same areas of electrode placement in a single patient. The agreement between the direct and indirect BIS values was statistically analyzed. RESULTS: The interface device with a silver electrode demonstrated sufficient electric conduction to transmit electroencephalogram signals. The overall BIS curves were similar to those of direct BIS monitoring. Direct and indirect BIS values from 18 patients were statistically analyzed using a linear mixed model and a significant concordance was confirmed (indirect BIS = 7.0405 + 0.8286 * direct BIS, p<0.0001). Most observed data (2582/2787 data points, 92.64%) had BIS unit differences of 10 or less. CONCLUSIONS: The interface device provides an opportunity for intraoperative BIS monitoring of patients, whose clinical situation does not permit the placement of conventional adhesive sensors at the standard location.
format Online
Article
Text
id pubmed-8530286
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-85302862021-10-22 Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures Harada, Hideki Muta, Seiya Kakuma, Tatsuyuki Ukeda, Misa Ota, So Hirata, Maiko Fujioka, Hiroshi Nakashima, Osamu Dietel, Barbara Tauchi, Miyuki PLoS One Research Article BACKGROUND: Bispectral index (BIS) monitoring is a widely used non-invasive method to monitor the depth of anesthesia. However, in the event of surgeries requiring a frontal approach, placement of the electrode may be impossible at the designated area to achieve a proper BIS measurement. METHODS: We developed an investigational interface device to connect needle-electrodes to BIS sensors. The safety and clinical performance were investigated in patients who underwent surgery. Direct BIS values from a disposable BIS electrode and indirect values via the interface device were simultaneously recorded from the same areas of electrode placement in a single patient. The agreement between the direct and indirect BIS values was statistically analyzed. RESULTS: The interface device with a silver electrode demonstrated sufficient electric conduction to transmit electroencephalogram signals. The overall BIS curves were similar to those of direct BIS monitoring. Direct and indirect BIS values from 18 patients were statistically analyzed using a linear mixed model and a significant concordance was confirmed (indirect BIS = 7.0405 + 0.8286 * direct BIS, p<0.0001). Most observed data (2582/2787 data points, 92.64%) had BIS unit differences of 10 or less. CONCLUSIONS: The interface device provides an opportunity for intraoperative BIS monitoring of patients, whose clinical situation does not permit the placement of conventional adhesive sensors at the standard location. Public Library of Science 2021-10-21 /pmc/articles/PMC8530286/ /pubmed/34673803 http://dx.doi.org/10.1371/journal.pone.0258647 Text en © 2021 Harada et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Harada, Hideki
Muta, Seiya
Kakuma, Tatsuyuki
Ukeda, Misa
Ota, So
Hirata, Maiko
Fujioka, Hiroshi
Nakashima, Osamu
Dietel, Barbara
Tauchi, Miyuki
Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures
title Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures
title_full Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures
title_fullStr Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures
title_full_unstemmed Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures
title_short Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures
title_sort accuracy of bis monitoring using a novel interface device connecting conventional needle-electrodes and bis sensors during frontal neurosurgical procedures
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8530286/
https://www.ncbi.nlm.nih.gov/pubmed/34673803
http://dx.doi.org/10.1371/journal.pone.0258647
work_keys_str_mv AT haradahideki accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT mutaseiya accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT kakumatatsuyuki accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT ukedamisa accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT otaso accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT hiratamaiko accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT fujiokahiroshi accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT nakashimaosamu accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT dietelbarbara accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures
AT tauchimiyuki accuracyofbismonitoringusinganovelinterfacedeviceconnectingconventionalneedleelectrodesandbissensorsduringfrontalneurosurgicalprocedures