Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model

Background: Extracorporeal membrane oxygenation (ECMO) provides cardiopulmonary support for critically ill patients. Portable ECMO devices can be applied in both in-hospital and out-of-hospital emergency conditions. We evaluated the safety and biocompatibility of a novel centrifugal pump and ECMO de...

Descripción completa

Detalles Bibliográficos
Autores principales: Gao, Sizhe, Wang, Weining, Qi, Jiachen, Liu, Gang, Wang, Jian, Yan, Shujie, Teng, Yuan, Zhou, Chun, Wang, Qian, Yan, Weidong, Zhang, Qiaoni, Liu, Youjun, Gao, Bin, Ji, Bingyang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8542924/
https://www.ncbi.nlm.nih.gov/pubmed/34708051
http://dx.doi.org/10.3389/fmed.2021.712205
_version_ 1784589531640496128
author Gao, Sizhe
Wang, Weining
Qi, Jiachen
Liu, Gang
Wang, Jian
Yan, Shujie
Teng, Yuan
Zhou, Chun
Wang, Qian
Yan, Weidong
Zhang, Qiaoni
Liu, Youjun
Gao, Bin
Ji, Bingyang
author_facet Gao, Sizhe
Wang, Weining
Qi, Jiachen
Liu, Gang
Wang, Jian
Yan, Shujie
Teng, Yuan
Zhou, Chun
Wang, Qian
Yan, Weidong
Zhang, Qiaoni
Liu, Youjun
Gao, Bin
Ji, Bingyang
author_sort Gao, Sizhe
collection PubMed
description Background: Extracorporeal membrane oxygenation (ECMO) provides cardiopulmonary support for critically ill patients. Portable ECMO devices can be applied in both in-hospital and out-of-hospital emergency conditions. We evaluated the safety and biocompatibility of a novel centrifugal pump and ECMO device of the OASSIST ECMO System (Jiangsu STMed Technologies Co., Suzhou, China) in a 168-h ovine ECMO model. Methods: The portable OASSIST ECMO system consists of the control console, the pump drive, and the disposable centrifugal pump. Ten healthy sheep were used to evaluate the OASSIST ECMO system. Five were supported on veno-venous ECMO and five on veno-arterial ECMO, each for 168 h. The systemic anticoagulation was achieved by continuous heparin infusion to maintain the activated clotting time (ACT) between 220 and 250 s. The rotary speed was set at 3,200–3,500 rpm. The ECMO configurations and ACT were recorded every 6 hours (h). The free hemoglobin (fHb), complete blood count, and coagulation action test were monitored, at the 6th h and every 24 h after the initiation of the ECMO. The dissection of the pump head and oxygenator were conducted to explore thrombosis. Results: Ten sheep successfully completed the study duration without device-related accidents. The pumps ran stably, and the ECMO flow ranged from 1.6 ± 0.1 to 2.0 ± 0.11 L/min in the V-V group, and from 1.8 ± 0.1 to 2.4 ± 0.14 L/min in the V-A group. The anticoagulation was well-performed. The ACT was maintained at 239.78 ± 36.31 s, no major bleeding or thrombosis was observed during the ECMO run or in the autopsy. 3/5 in the V-A group and 4/5 in the V-V group developed small thrombus in the bearing pedestal. No obvious thrombus formed in the oxygenator was observed. The hemolytic blood damage was not significant. The average fHb was 0.17 ± 0.12 g/L. Considering hemodilution, the hemoglobin, white blood cell, and platelets didn't reduce during the ECMO runs. Conclusions: The OASSIST ECMO system shows satisfactory safety and biocompatibility for the 168-h preclinical evaluation in the ovine model. The OASSIST ECMO system is promising to be applied in clinical conditions in the future.
format Online
Article
Text
id pubmed-8542924
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-85429242021-10-26 Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model Gao, Sizhe Wang, Weining Qi, Jiachen Liu, Gang Wang, Jian Yan, Shujie Teng, Yuan Zhou, Chun Wang, Qian Yan, Weidong Zhang, Qiaoni Liu, Youjun Gao, Bin Ji, Bingyang Front Med (Lausanne) Medicine Background: Extracorporeal membrane oxygenation (ECMO) provides cardiopulmonary support for critically ill patients. Portable ECMO devices can be applied in both in-hospital and out-of-hospital emergency conditions. We evaluated the safety and biocompatibility of a novel centrifugal pump and ECMO device of the OASSIST ECMO System (Jiangsu STMed Technologies Co., Suzhou, China) in a 168-h ovine ECMO model. Methods: The portable OASSIST ECMO system consists of the control console, the pump drive, and the disposable centrifugal pump. Ten healthy sheep were used to evaluate the OASSIST ECMO system. Five were supported on veno-venous ECMO and five on veno-arterial ECMO, each for 168 h. The systemic anticoagulation was achieved by continuous heparin infusion to maintain the activated clotting time (ACT) between 220 and 250 s. The rotary speed was set at 3,200–3,500 rpm. The ECMO configurations and ACT were recorded every 6 hours (h). The free hemoglobin (fHb), complete blood count, and coagulation action test were monitored, at the 6th h and every 24 h after the initiation of the ECMO. The dissection of the pump head and oxygenator were conducted to explore thrombosis. Results: Ten sheep successfully completed the study duration without device-related accidents. The pumps ran stably, and the ECMO flow ranged from 1.6 ± 0.1 to 2.0 ± 0.11 L/min in the V-V group, and from 1.8 ± 0.1 to 2.4 ± 0.14 L/min in the V-A group. The anticoagulation was well-performed. The ACT was maintained at 239.78 ± 36.31 s, no major bleeding or thrombosis was observed during the ECMO run or in the autopsy. 3/5 in the V-A group and 4/5 in the V-V group developed small thrombus in the bearing pedestal. No obvious thrombus formed in the oxygenator was observed. The hemolytic blood damage was not significant. The average fHb was 0.17 ± 0.12 g/L. Considering hemodilution, the hemoglobin, white blood cell, and platelets didn't reduce during the ECMO runs. Conclusions: The OASSIST ECMO system shows satisfactory safety and biocompatibility for the 168-h preclinical evaluation in the ovine model. The OASSIST ECMO system is promising to be applied in clinical conditions in the future. Frontiers Media S.A. 2021-10-11 /pmc/articles/PMC8542924/ /pubmed/34708051 http://dx.doi.org/10.3389/fmed.2021.712205 Text en Copyright © 2021 Gao, Wang, Qi, Liu, Wang, Yan, Teng, Zhou, Wang, Yan, Zhang, Liu, Gao and Ji. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Medicine
Gao, Sizhe
Wang, Weining
Qi, Jiachen
Liu, Gang
Wang, Jian
Yan, Shujie
Teng, Yuan
Zhou, Chun
Wang, Qian
Yan, Weidong
Zhang, Qiaoni
Liu, Youjun
Gao, Bin
Ji, Bingyang
Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model
title Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model
title_full Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model
title_fullStr Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model
title_full_unstemmed Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model
title_short Safety and Efficacy of a Novel Centrifugal Pump and Driving Devices of the OASSIST ECMO System: A Preclinical Evaluation in the Ovine Model
title_sort safety and efficacy of a novel centrifugal pump and driving devices of the oassist ecmo system: a preclinical evaluation in the ovine model
topic Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8542924/
https://www.ncbi.nlm.nih.gov/pubmed/34708051
http://dx.doi.org/10.3389/fmed.2021.712205
work_keys_str_mv AT gaosizhe safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT wangweining safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT qijiachen safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT liugang safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT wangjian safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT yanshujie safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT tengyuan safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT zhouchun safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT wangqian safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT yanweidong safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT zhangqiaoni safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT liuyoujun safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT gaobin safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel
AT jibingyang safetyandefficacyofanovelcentrifugalpumpanddrivingdevicesoftheoassistecmosystemapreclinicalevaluationintheovinemodel