Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes

PURPOSE: There is a considerable lack of awareness of slit-lamp measurement of anterior chamber depth (ACD) by the Redmond Smith method (SACD) in present day-to-day clinical practice, which may provide rapid assessment in pseudoexfoliation (PXF) when assessing for angle closure and planning for cata...

Descripción completa

Detalles Bibliográficos
Autores principales: Pathak Ray, Vanita, Ramesh, Smitha B, Rathi, Varsha
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8544031/
https://www.ncbi.nlm.nih.gov/pubmed/34427246
http://dx.doi.org/10.4103/ijo.IJO_3738_20
_version_ 1784589732805607424
author Pathak Ray, Vanita
Ramesh, Smitha B
Rathi, Varsha
author_facet Pathak Ray, Vanita
Ramesh, Smitha B
Rathi, Varsha
author_sort Pathak Ray, Vanita
collection PubMed
description PURPOSE: There is a considerable lack of awareness of slit-lamp measurement of anterior chamber depth (ACD) by the Redmond Smith method (SACD) in present day-to-day clinical practice, which may provide rapid assessment in pseudoexfoliation (PXF) when assessing for angle closure and planning for cataract surgery. This assumes importance not only in outreach clinics but also in the ongoing pandemic caused by the highly contagious novel coronavirus, where social distancing is advocated to contain the spread. We aimed to compare the axial ACD in PXF and normal patients by SACD, and its agreement with the anterior segment optical coherence tomography (ASOCT) and LenstarLS-900. METHODS: A prospective comparative observational study was done at a tertiary eye care hospital. A PXF group and a normal group of controls were recruited. All eyes were phakic with normal cornea. Any eye with previous intraocular/refractive surgery and cause of other secondary or uncontrolled glaucoma was excluded. SACD was measured clinically via slit-lamp method and also via ASOCT and Lenstar; agreement between the methodologies was plotted. RESULTS: Fifty patients were recruited in each group. Mean age was 66.82 ± 4.88 years in PXF patients and 65 ± 5.46 years in controls (P = 0.2). ACD was found to be greater in controls compared with the PXF patients; this difference was statistically significant (P < 0.001) across all methodologies. A good agreement with narrow 95% limits of agreement was found between these methodologies. CONCLUSION: Redmond Smith slit-lamp methodology of estimating the axial ACD is recommended as a rapid, quantifiable, noncontact screening technique during routine examination, especially in primary outreach centers, and is also advantageous during the ongoing pandemic by reducing expendable investigations.
format Online
Article
Text
id pubmed-8544031
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-85440312021-10-29 Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes Pathak Ray, Vanita Ramesh, Smitha B Rathi, Varsha Indian J Ophthalmol Special Focus, Glaucoma, Original Article PURPOSE: There is a considerable lack of awareness of slit-lamp measurement of anterior chamber depth (ACD) by the Redmond Smith method (SACD) in present day-to-day clinical practice, which may provide rapid assessment in pseudoexfoliation (PXF) when assessing for angle closure and planning for cataract surgery. This assumes importance not only in outreach clinics but also in the ongoing pandemic caused by the highly contagious novel coronavirus, where social distancing is advocated to contain the spread. We aimed to compare the axial ACD in PXF and normal patients by SACD, and its agreement with the anterior segment optical coherence tomography (ASOCT) and LenstarLS-900. METHODS: A prospective comparative observational study was done at a tertiary eye care hospital. A PXF group and a normal group of controls were recruited. All eyes were phakic with normal cornea. Any eye with previous intraocular/refractive surgery and cause of other secondary or uncontrolled glaucoma was excluded. SACD was measured clinically via slit-lamp method and also via ASOCT and Lenstar; agreement between the methodologies was plotted. RESULTS: Fifty patients were recruited in each group. Mean age was 66.82 ± 4.88 years in PXF patients and 65 ± 5.46 years in controls (P = 0.2). ACD was found to be greater in controls compared with the PXF patients; this difference was statistically significant (P < 0.001) across all methodologies. A good agreement with narrow 95% limits of agreement was found between these methodologies. CONCLUSION: Redmond Smith slit-lamp methodology of estimating the axial ACD is recommended as a rapid, quantifiable, noncontact screening technique during routine examination, especially in primary outreach centers, and is also advantageous during the ongoing pandemic by reducing expendable investigations. Wolters Kluwer - Medknow 2021-09 2021-08-25 /pmc/articles/PMC8544031/ /pubmed/34427246 http://dx.doi.org/10.4103/ijo.IJO_3738_20 Text en Copyright: © 2021 Indian Journal of Ophthalmology https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 4.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Special Focus, Glaucoma, Original Article
Pathak Ray, Vanita
Ramesh, Smitha B
Rathi, Varsha
Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes
title Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes
title_full Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes
title_fullStr Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes
title_full_unstemmed Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes
title_short Slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and Lenstar LS 900 in pseudoexfoliation and normal eyes
title_sort slit-lamp measurement of anterior chamber depth and its agreement with anterior segment optical coherence tomography and lenstar ls 900 in pseudoexfoliation and normal eyes
topic Special Focus, Glaucoma, Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8544031/
https://www.ncbi.nlm.nih.gov/pubmed/34427246
http://dx.doi.org/10.4103/ijo.IJO_3738_20
work_keys_str_mv AT pathakrayvanita slitlampmeasurementofanteriorchamberdepthanditsagreementwithanteriorsegmentopticalcoherencetomographyandlenstarls900inpseudoexfoliationandnormaleyes
AT rameshsmithab slitlampmeasurementofanteriorchamberdepthanditsagreementwithanteriorsegmentopticalcoherencetomographyandlenstarls900inpseudoexfoliationandnormaleyes
AT rathivarsha slitlampmeasurementofanteriorchamberdepthanditsagreementwithanteriorsegmentopticalcoherencetomographyandlenstarls900inpseudoexfoliationandnormaleyes