Cargando…

Late Breakage of a Dual-Mobility Polyethylene Insert in a Revision Total Hip Arthroplasty: An Unusual Failure Mode

The use of dual-mobility cups has gained popularity in recent years. Thus, surgeons can expect an increase in known and new causes of failure. We report a previously undescribed form of a late intraprosthetic dislocation consisting of a complete breakage of a polyethylene mobile bearing that suffere...

Descripción completa

Detalles Bibliográficos
Autores principales: Rodríguez Pérez, Daniel, Agulló Ferre, José Luis, Del Carmen Rodríguez, Marcos, Tramunt Monsonet, Carles
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8551416/
https://www.ncbi.nlm.nih.gov/pubmed/34746346
http://dx.doi.org/10.1016/j.artd.2021.07.013
Descripción
Sumario:The use of dual-mobility cups has gained popularity in recent years. Thus, surgeons can expect an increase in known and new causes of failure. We report a previously undescribed form of a late intraprosthetic dislocation consisting of a complete breakage of a polyethylene mobile bearing that suffered a dislocation 3 months after its implantation. Two years later, he began feeling anterior groin pain and suffered gait changes. Computed tomography scan revealed an eccentric alignment of the mobile polyethylene bearing suggestive of poly wear. During the revision surgery, the polyethylene was found to be split in 2. Possible causes of this complication are proposed. Our case shows a previously unreported implant-specific complication, so surgeons can identify it.