Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study
BACKGROUND: Measures of the sonorheometry based Quantra® viscoelastic hemostatic analyzer (HemoSonics, LCC, Charlottesville, VA, USA) were compared with corresponding results of the ROTEM® sigma device (Instrumentation Laboratory, Bedford, MA, USA). METHODS: In thirty-eight patients scheduled for el...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8555139/ https://www.ncbi.nlm.nih.gov/pubmed/34711167 http://dx.doi.org/10.1186/s12871-021-01469-5 |
_version_ | 1784591917552500736 |
---|---|
author | Baulig, Werner Akbas, Samira Schütt, Philipp K. Keul, Wolfgang Jovic, Marija Berdat, Pascal von Felten, Stefanie Steigmiller, Klaus Ganter, Michael Thomas Theusinger, Oliver M. |
author_facet | Baulig, Werner Akbas, Samira Schütt, Philipp K. Keul, Wolfgang Jovic, Marija Berdat, Pascal von Felten, Stefanie Steigmiller, Klaus Ganter, Michael Thomas Theusinger, Oliver M. |
author_sort | Baulig, Werner |
collection | PubMed |
description | BACKGROUND: Measures of the sonorheometry based Quantra® viscoelastic hemostatic analyzer (HemoSonics, LCC, Charlottesville, VA, USA) were compared with corresponding results of the ROTEM® sigma device (Instrumentation Laboratory, Bedford, MA, USA). METHODS: In thirty-eight patients scheduled for elective cardiac surgery between December 2018 and October 2019, blood samples were taken after induction of anesthesia (sample 1) and after heparin neutralization (sample 2) and measured on Quantra (QPlus® Cartridge) and ROTEM sigma (ROTEM® sigma complete + hep Cartridge). Clot times and clot stiffness values were recorded. Clot stiffness values of ROTEM amplitudes (A in mm) were converted to shear modulus (G) in hectoPascal (hPa): G (hPa) = (5 x A)/(100-A). Additionally, time-to-results was recorded. Spearman rank test correlation and Bland Altman analysis were performed. RESULTS: Clot stiffness parameters of the Quantra correlated strongly with corresponding measurements of the ROTEM with r = 0.93 and 0.94 for EXTEM A10 vs CS and r = 0.94 and 0.96 for FIBTEM A10 vs FCS for sample 1 and 2, respectively. Quantra clot time correlated strongly with ROTEM INTEM CT with r = 0.71 for sample 1 and r = 0.75 for sample 2. However, Bland Altman analysis showed no agreement in all compared assays of both methods. The median time to delivery of first and complete results was significantly shorter for Quantra (412 and 658 s) compared to ROTEM sigma (839 and 1290 s). CONCLUSIONS: The Quantra showed a strong correlation with the ROTEM sigma for determining clot times and clot stiffness and the parameters assess similar aspects of clot development. However, these parameters are not directly interchangeable and implicate that separate cut-off values need to be established for users of the Quantra device. Word count: 278. TRIAL REGISTRATION: The study was retrospectively registered with ClinicalTrials.gov (ID: NCT04210830) at December 20th 2019. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12871-021-01469-5. |
format | Online Article Text |
id | pubmed-8555139 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-85551392021-10-29 Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study Baulig, Werner Akbas, Samira Schütt, Philipp K. Keul, Wolfgang Jovic, Marija Berdat, Pascal von Felten, Stefanie Steigmiller, Klaus Ganter, Michael Thomas Theusinger, Oliver M. BMC Anesthesiol Research BACKGROUND: Measures of the sonorheometry based Quantra® viscoelastic hemostatic analyzer (HemoSonics, LCC, Charlottesville, VA, USA) were compared with corresponding results of the ROTEM® sigma device (Instrumentation Laboratory, Bedford, MA, USA). METHODS: In thirty-eight patients scheduled for elective cardiac surgery between December 2018 and October 2019, blood samples were taken after induction of anesthesia (sample 1) and after heparin neutralization (sample 2) and measured on Quantra (QPlus® Cartridge) and ROTEM sigma (ROTEM® sigma complete + hep Cartridge). Clot times and clot stiffness values were recorded. Clot stiffness values of ROTEM amplitudes (A in mm) were converted to shear modulus (G) in hectoPascal (hPa): G (hPa) = (5 x A)/(100-A). Additionally, time-to-results was recorded. Spearman rank test correlation and Bland Altman analysis were performed. RESULTS: Clot stiffness parameters of the Quantra correlated strongly with corresponding measurements of the ROTEM with r = 0.93 and 0.94 for EXTEM A10 vs CS and r = 0.94 and 0.96 for FIBTEM A10 vs FCS for sample 1 and 2, respectively. Quantra clot time correlated strongly with ROTEM INTEM CT with r = 0.71 for sample 1 and r = 0.75 for sample 2. However, Bland Altman analysis showed no agreement in all compared assays of both methods. The median time to delivery of first and complete results was significantly shorter for Quantra (412 and 658 s) compared to ROTEM sigma (839 and 1290 s). CONCLUSIONS: The Quantra showed a strong correlation with the ROTEM sigma for determining clot times and clot stiffness and the parameters assess similar aspects of clot development. However, these parameters are not directly interchangeable and implicate that separate cut-off values need to be established for users of the Quantra device. Word count: 278. TRIAL REGISTRATION: The study was retrospectively registered with ClinicalTrials.gov (ID: NCT04210830) at December 20th 2019. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12871-021-01469-5. BioMed Central 2021-10-28 /pmc/articles/PMC8555139/ /pubmed/34711167 http://dx.doi.org/10.1186/s12871-021-01469-5 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Baulig, Werner Akbas, Samira Schütt, Philipp K. Keul, Wolfgang Jovic, Marija Berdat, Pascal von Felten, Stefanie Steigmiller, Klaus Ganter, Michael Thomas Theusinger, Oliver M. Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study |
title | Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study |
title_full | Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study |
title_fullStr | Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study |
title_full_unstemmed | Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study |
title_short | Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery – a prospective observational study |
title_sort | comparison of the resonance sonorheometry based quantra® system with rotational thromboelastometry rotem® sigma in cardiac surgery – a prospective observational study |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8555139/ https://www.ncbi.nlm.nih.gov/pubmed/34711167 http://dx.doi.org/10.1186/s12871-021-01469-5 |
work_keys_str_mv | AT bauligwerner comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT akbassamira comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT schuttphilippk comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT keulwolfgang comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT jovicmarija comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT berdatpascal comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT vonfeltenstefanie comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT steigmillerklaus comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT gantermichaelthomas comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy AT theusingeroliverm comparisonoftheresonancesonorheometrybasedquantrasystemwithrotationalthromboelastometryrotemsigmaincardiacsurgeryaprospectiveobservationalstudy |