Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer

PURPOSE: The purpose of this study was to compare visual field results of the COMPASS fundus perimeter (CMP) and the Humphrey Field Analyzer (HFA) in the same eyes; to compare structure-function concordance between circumpapillary retinal nerve fiber layer (Cp-RNFL) profiles and the two perimetry re...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Ping, Nguyen, Bao N., Turpin, Andrew, McKendrick, Allison M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8556560/
https://www.ncbi.nlm.nih.gov/pubmed/34694332
http://dx.doi.org/10.1167/tvst.10.12.33
_version_ 1784592192868712448
author Liu, Ping
Nguyen, Bao N.
Turpin, Andrew
McKendrick, Allison M.
author_facet Liu, Ping
Nguyen, Bao N.
Turpin, Andrew
McKendrick, Allison M.
author_sort Liu, Ping
collection PubMed
description PURPOSE: The purpose of this study was to compare visual field results of the COMPASS fundus perimeter (CMP) and the Humphrey Field Analyzer (HFA) in the same eyes; to compare structure-function concordance between circumpapillary retinal nerve fiber layer (Cp-RNFL) profiles and the two perimetry results; and to evaluate whether differences between the two results reflect postulated advantages of real-time eye movement compensation during perimetry. METHODS: We retrospectively analyzed 24-2 visual field data measured with CMP and HFA together with Cp-RNFL optical coherence tomography (OCT) scan data from 95 eyes of 65 people with glaucoma. We defined visual field locations with total deviation (TD) less than −5 dB as defective. The CMP and HFA fields were compared on measures of: spatial extent (number of defective locations); depth (TD values); and sharpness of scotomata edges (maximum TD difference between defective locations and their neighbors). Structure-function concordance between Cp-RNFL profile and respective visual field was also compared. RESULTS: Compared to the HFA, scotomata measured by CMP were of reduced spatial extent (mean difference = −3.14 locations, p < 0.001), greater depth (median TD of CMP = −17 dB versus HFA = −13 dB, p = 0.029) and steeper edges (median of maximum TD difference of CMP = 10.6 dB versus HFA = 6 dB, p < 0.001). Structure-function concordance between Cp-RNFL profile and either visual field were comparable despite the reduced scotoma spatial extent measured by CMP. CONCLUSIONS: Glaucomatous visual fields measured by CMP displayed characteristics consistent with expected effects of using real-time eye movement compensation technology compared to the widely used HFA. TRANSLATIONAL RELEVANCE: Glaucomatous visual field defects measured by the CMP are more localized, deeper, and steeper than those of the HFA.
format Online
Article
Text
id pubmed-8556560
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-85565602021-11-09 Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer Liu, Ping Nguyen, Bao N. Turpin, Andrew McKendrick, Allison M. Transl Vis Sci Technol Article PURPOSE: The purpose of this study was to compare visual field results of the COMPASS fundus perimeter (CMP) and the Humphrey Field Analyzer (HFA) in the same eyes; to compare structure-function concordance between circumpapillary retinal nerve fiber layer (Cp-RNFL) profiles and the two perimetry results; and to evaluate whether differences between the two results reflect postulated advantages of real-time eye movement compensation during perimetry. METHODS: We retrospectively analyzed 24-2 visual field data measured with CMP and HFA together with Cp-RNFL optical coherence tomography (OCT) scan data from 95 eyes of 65 people with glaucoma. We defined visual field locations with total deviation (TD) less than −5 dB as defective. The CMP and HFA fields were compared on measures of: spatial extent (number of defective locations); depth (TD values); and sharpness of scotomata edges (maximum TD difference between defective locations and their neighbors). Structure-function concordance between Cp-RNFL profile and respective visual field was also compared. RESULTS: Compared to the HFA, scotomata measured by CMP were of reduced spatial extent (mean difference = −3.14 locations, p < 0.001), greater depth (median TD of CMP = −17 dB versus HFA = −13 dB, p = 0.029) and steeper edges (median of maximum TD difference of CMP = 10.6 dB versus HFA = 6 dB, p < 0.001). Structure-function concordance between Cp-RNFL profile and either visual field were comparable despite the reduced scotoma spatial extent measured by CMP. CONCLUSIONS: Glaucomatous visual fields measured by CMP displayed characteristics consistent with expected effects of using real-time eye movement compensation technology compared to the widely used HFA. TRANSLATIONAL RELEVANCE: Glaucomatous visual field defects measured by the CMP are more localized, deeper, and steeper than those of the HFA. The Association for Research in Vision and Ophthalmology 2021-10-25 /pmc/articles/PMC8556560/ /pubmed/34694332 http://dx.doi.org/10.1167/tvst.10.12.33 Text en Copyright 2021 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
spellingShingle Article
Liu, Ping
Nguyen, Bao N.
Turpin, Andrew
McKendrick, Allison M.
Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer
title Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer
title_full Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer
title_fullStr Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer
title_full_unstemmed Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer
title_short Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer
title_sort increased depth, reduced extent, and sharpened edges of visual field defects measured by compass fundus perimeter compared to humphrey field analyzer
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8556560/
https://www.ncbi.nlm.nih.gov/pubmed/34694332
http://dx.doi.org/10.1167/tvst.10.12.33
work_keys_str_mv AT liuping increaseddepthreducedextentandsharpenededgesofvisualfielddefectsmeasuredbycompassfundusperimetercomparedtohumphreyfieldanalyzer
AT nguyenbaon increaseddepthreducedextentandsharpenededgesofvisualfielddefectsmeasuredbycompassfundusperimetercomparedtohumphreyfieldanalyzer
AT turpinandrew increaseddepthreducedextentandsharpenededgesofvisualfielddefectsmeasuredbycompassfundusperimetercomparedtohumphreyfieldanalyzer
AT mckendrickallisonm increaseddepthreducedextentandsharpenededgesofvisualfielddefectsmeasuredbycompassfundusperimetercomparedtohumphreyfieldanalyzer