Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis
BACKGROUND: Phase contrast (PC) cardiovascular magnetic resonance (CMR) in the ascending aorta (AAo) is widely used to calculate left ventricular (LV) stroke volume (SV). The accuracy of PC CMR may be altered by turbulent flow. Measurement of SV at another site is suggested in the presence of aortic...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8567621/ https://www.ncbi.nlm.nih.gov/pubmed/34732204 http://dx.doi.org/10.1186/s12968-021-00814-4 |
_version_ | 1784594273092501504 |
---|---|
author | Guzzetti, Ezequiel Racine, Hugo-Pierre Tastet, Lionel Shen, Mylène Larose, Eric Clavel, Marie-Annick Pibarot, Philippe Beaudoin, Jonathan |
author_facet | Guzzetti, Ezequiel Racine, Hugo-Pierre Tastet, Lionel Shen, Mylène Larose, Eric Clavel, Marie-Annick Pibarot, Philippe Beaudoin, Jonathan |
author_sort | Guzzetti, Ezequiel |
collection | PubMed |
description | BACKGROUND: Phase contrast (PC) cardiovascular magnetic resonance (CMR) in the ascending aorta (AAo) is widely used to calculate left ventricular (LV) stroke volume (SV). The accuracy of PC CMR may be altered by turbulent flow. Measurement of SV at another site is suggested in the presence of aortic stenosis, but very few data validates the accuracy or inaccuracy of PC in that setting. Our objective is to compare flow measurements obtained in the AAo and LV outflow tract (LVOT) in patients with aortic stenosis. METHODS: Retrospective analysis of patients with aortic stenosis who had CMR and echocardiography. Patients with mitral regurgitation were excluded. PC in the AAo and LVOT were acquired to derive SV. LV SV from end-systolic and end-diastolic tracings was used as the reference measure. A difference ≥ 10% between the volumetric method and PC derived SVs was considered discordant. Metrics of turbulence and jet eccentricity were assessed to explore the predictors of discordant measurements. RESULTS: We included 88 patients, 41% with bicuspid aortic valve. LVOT SV was concordant with the volumetric method in 79 (90%) patients vs 52 (59%) patients for AAo SV (p = 0.015). In multivariate analysis, aortic stenosis flow jet angle was a strong predictor of discordant measurement in the AAo (p = 0.003). Mathematical correction for the jet angle improved the concordance from 59 to 91%. Concordance was comparable in patients with bicuspid and trileaflet valves (57% and 62% concordance respectively; p = 0.11). Accuracy of SV measured in the LVOT was not influenced by jet eccentricity. For aortic regurgitation quantification, PC in the AAo had better correlation to volumetric assessments than LVOT PC. CONCLUSION: LVOT PC SV in patients with aortic stenosis and eccentric jet might be more accurate compared to the AAo SV. Mathematical correction for the jet angle in the AAo might be another alternative to improve accuracy. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12968-021-00814-4. |
format | Online Article Text |
id | pubmed-8567621 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-85676212021-11-04 Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis Guzzetti, Ezequiel Racine, Hugo-Pierre Tastet, Lionel Shen, Mylène Larose, Eric Clavel, Marie-Annick Pibarot, Philippe Beaudoin, Jonathan J Cardiovasc Magn Reson Research BACKGROUND: Phase contrast (PC) cardiovascular magnetic resonance (CMR) in the ascending aorta (AAo) is widely used to calculate left ventricular (LV) stroke volume (SV). The accuracy of PC CMR may be altered by turbulent flow. Measurement of SV at another site is suggested in the presence of aortic stenosis, but very few data validates the accuracy or inaccuracy of PC in that setting. Our objective is to compare flow measurements obtained in the AAo and LV outflow tract (LVOT) in patients with aortic stenosis. METHODS: Retrospective analysis of patients with aortic stenosis who had CMR and echocardiography. Patients with mitral regurgitation were excluded. PC in the AAo and LVOT were acquired to derive SV. LV SV from end-systolic and end-diastolic tracings was used as the reference measure. A difference ≥ 10% between the volumetric method and PC derived SVs was considered discordant. Metrics of turbulence and jet eccentricity were assessed to explore the predictors of discordant measurements. RESULTS: We included 88 patients, 41% with bicuspid aortic valve. LVOT SV was concordant with the volumetric method in 79 (90%) patients vs 52 (59%) patients for AAo SV (p = 0.015). In multivariate analysis, aortic stenosis flow jet angle was a strong predictor of discordant measurement in the AAo (p = 0.003). Mathematical correction for the jet angle improved the concordance from 59 to 91%. Concordance was comparable in patients with bicuspid and trileaflet valves (57% and 62% concordance respectively; p = 0.11). Accuracy of SV measured in the LVOT was not influenced by jet eccentricity. For aortic regurgitation quantification, PC in the AAo had better correlation to volumetric assessments than LVOT PC. CONCLUSION: LVOT PC SV in patients with aortic stenosis and eccentric jet might be more accurate compared to the AAo SV. Mathematical correction for the jet angle in the AAo might be another alternative to improve accuracy. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12968-021-00814-4. BioMed Central 2021-11-04 /pmc/articles/PMC8567621/ /pubmed/34732204 http://dx.doi.org/10.1186/s12968-021-00814-4 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Guzzetti, Ezequiel Racine, Hugo-Pierre Tastet, Lionel Shen, Mylène Larose, Eric Clavel, Marie-Annick Pibarot, Philippe Beaudoin, Jonathan Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis |
title | Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis |
title_full | Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis |
title_fullStr | Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis |
title_full_unstemmed | Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis |
title_short | Accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis |
title_sort | accuracy of stroke volume measurement with phase-contrast cardiovascular magnetic resonance in patients with aortic stenosis |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8567621/ https://www.ncbi.nlm.nih.gov/pubmed/34732204 http://dx.doi.org/10.1186/s12968-021-00814-4 |
work_keys_str_mv | AT guzzettiezequiel accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis AT racinehugopierre accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis AT tastetlionel accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis AT shenmylene accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis AT laroseeric accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis AT clavelmarieannick accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis AT pibarotphilippe accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis AT beaudoinjonathan accuracyofstrokevolumemeasurementwithphasecontrastcardiovascularmagneticresonanceinpatientswithaorticstenosis |