Investigation of the Number of Tests Required for Assaying Plasma Biomarkers Associated with Alzheimer’s Disease Using Immunomagnetic Reduction

INTRODUCTION: Concentrations of plasma biomarkers associated with Alzheimer’s disease have been reported to be as low as several tens of picograms/milliliter (pg/ml). However, in assays measuring these biomarkers, it is likely that repeated measurements are necessary to obtain reliable values. METHO...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Huei-Chun, Chen, Hsin-Hsien, Ho, Chia-Shin, Chang, Jui-Feng, Lin, Chia-Chun, Chiu, Ming-Jang, Chen, Ta-Fu, Hu, Chaur-Jong, Yan, Sui-Hing, Sun, Yu, Yang, Shieh-Yueh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Healthcare 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8571465/
https://www.ncbi.nlm.nih.gov/pubmed/34515952
http://dx.doi.org/10.1007/s40120-021-00280-1
Descripción
Sumario:INTRODUCTION: Concentrations of plasma biomarkers associated with Alzheimer’s disease have been reported to be as low as several tens of picograms/milliliter (pg/ml). However, in assays measuring these biomarkers, it is likely that repeated measurements are necessary to obtain reliable values. METHODS: We performed assays as a single test or as duplicate, quadruplicate, fivefold and tenfold repeated tests, on samples spiked with different concentrations of amyloid β 1–40 (Aβ(1–40); 1–1000 pg/ml), Aβ(1–42) (1–30,000 pg/ml) and total Tau protein (T-Tau; 0.1–1000 pg/ml), with the aim to to calculate the coefficients of variation (CVs). RESULTS: The results demonstrated common changes in the CVs with changes in the number of tests for a given sample: the CVs decreased with increases in the number of tests from one to ten. All CV values were distributed within the range of 0.35 to 15.5%; as such, the CV values were all lower than the acceptable value of 20%. CONCLUSION: Based on this study, a single assay of Aβ(1–40), Aβ(1–42) and T-Tau, respectively, provides reliable results in terms of the measurement of that plasma biomarker.