Carbon Dioxide Elimination After Sodium Bicarbonate Administration as a Novel Method to Assess Cardiac Output: A Pilot Study

Introduction Cardiac output/pulmonary blood flow measurement is an important way to assess patients during the perioperative period, as well as patients who are critically ill. Current methods of assessing cardiac output have limitations. One indicator of cardiac output may be the expired carbon dio...

Descripción completa

Detalles Bibliográficos
Autores principales: Keidan, Ilan, Arzillo, Samantha, Vasilopoulos, Terrie, Ben-Menachem, Erez, Gravenstein, Nikolaus, McGough, Edward
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8574226/
https://www.ncbi.nlm.nih.gov/pubmed/34765374
http://dx.doi.org/10.7759/cureus.18621
Descripción
Sumario:Introduction Cardiac output/pulmonary blood flow measurement is an important way to assess patients during the perioperative period, as well as patients who are critically ill. Current methods of assessing cardiac output have limitations. One indicator of cardiac output may be the expired carbon dioxide (CO(2)) partial pressure response to intravenous sodium bicarbonate (IVSB), which is rapidly converted to CO(2). Methods We conducted an initial evaluation of the relationship between expired CO(2) partial pressure and blood flow after a bolus of IVSB. To assess this relationship, we used a cardiopulmonary bypass circuit with predetermined blood flows in a laboratory trial and then assessed 18 patients undergoing surgery requiring cardiopulmonary bypass. Results For the laboratory portion of this pilot study, higher peak expired CO(2), faster time to reach peak, higher area under the curve, and greater kurtosis of peak were observed at higher cardiac output flow rates, and higher mean expired CO(2) was significantly associated with higher flow rates (p < 0.001). In the human study, higher mean (p = 0.023) and peak expired CO(2) (p = 0.028) were both significantly associated with higher cardiac output flow rates. Conclusions This technique may be a way to intermittently assess cardiac output or improve accuracy when used in conjunction with other continuous output monitors.