In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study

INTRODUCTION: The most currently recommended method for sterilization of orthodontic pliers is the autoclave, while peracetic acid has also been shown to be effective in the chemical sterilization process. OBJECTIVE: This study sought to compare the corrosive effects of peracetic acid and autoclave...

Descripción completa

Detalles Bibliográficos
Autores principales: NUNES, Livia Kelly Ferraz, CARNEIRO, Érika Lopes, PALOMARES, Nathalia Barbosa, SENNA, Lilian, SAMPAIO-FILHO, Helio, MIGUEL, José Augusto Mendes
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dental Press International 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8576861/
https://www.ncbi.nlm.nih.gov/pubmed/35640079
http://dx.doi.org/10.1590/2177-6709.26.5.e2119353.oar
_version_ 1784595963225047040
author NUNES, Livia Kelly Ferraz
CARNEIRO, Érika Lopes
PALOMARES, Nathalia Barbosa
SENNA, Lilian
SAMPAIO-FILHO, Helio
MIGUEL, José Augusto Mendes
author_facet NUNES, Livia Kelly Ferraz
CARNEIRO, Érika Lopes
PALOMARES, Nathalia Barbosa
SENNA, Lilian
SAMPAIO-FILHO, Helio
MIGUEL, José Augusto Mendes
author_sort NUNES, Livia Kelly Ferraz
collection PubMed
description INTRODUCTION: The most currently recommended method for sterilization of orthodontic pliers is the autoclave, while peracetic acid has also been shown to be effective in the chemical sterilization process. OBJECTIVE: This study sought to compare the corrosive effects of peracetic acid and autoclave sterilization process of orthodontic pliers. METHODS: Four active tungsten carbide (WC) stainless steel tie-cutting pliers from the manufacturers Quinelato (Rio Claro, SP, Brazil) and ICE (Cajamar, SP, Brazil) were selected. The active ends of the pliers were sectioned, and six active tips were obtained and distributed into the following groups: 1) control group (no sterilization); 2) AC group (two active pliers tips submitted to 100 autoclave sterilization cycles); and 3) AP group (two active pliers tips submitted to 100 cycles of sterilization by immersion in 2% peracetic acid solution for 30 minutes). RESULTS: Chemical analysis using X-ray dispersive energy spectroscopy showed that after autoclave sterilization, only the ICE pliers presented oxidation corrosion (Δ[O] = +24.5%; Δ[Fe] = +5.8%; Δ[WC] = -1.9%). In comparison, following peracetic acid sterilization, both manufacturers ICE (Δ[O] = +1.8%; Δ[Fe] = +18.0%; Δ[WC] = -1.1%) and Quinelato (Δ[O] = +5.3%; Δ[Fe] = -10.4%; Δ[WC] = -15.2%) showed corrosion. The morphological analysis revealed that peracetic acid caused a pitting and localized corrosion in both brands, while the autoclave caused uniform surface corrosion on the ICE pliers. CONCLUSION: Autoclave application was the sterilization method that generated less corrosive damage to the orthodontic cutting pliers, when compared to the immersion in 2% peracetic acid.
format Online
Article
Text
id pubmed-8576861
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Dental Press International
record_format MEDLINE/PubMed
spelling pubmed-85768612021-11-18 In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study NUNES, Livia Kelly Ferraz CARNEIRO, Érika Lopes PALOMARES, Nathalia Barbosa SENNA, Lilian SAMPAIO-FILHO, Helio MIGUEL, José Augusto Mendes Dental Press J Orthod Original Article INTRODUCTION: The most currently recommended method for sterilization of orthodontic pliers is the autoclave, while peracetic acid has also been shown to be effective in the chemical sterilization process. OBJECTIVE: This study sought to compare the corrosive effects of peracetic acid and autoclave sterilization process of orthodontic pliers. METHODS: Four active tungsten carbide (WC) stainless steel tie-cutting pliers from the manufacturers Quinelato (Rio Claro, SP, Brazil) and ICE (Cajamar, SP, Brazil) were selected. The active ends of the pliers were sectioned, and six active tips were obtained and distributed into the following groups: 1) control group (no sterilization); 2) AC group (two active pliers tips submitted to 100 autoclave sterilization cycles); and 3) AP group (two active pliers tips submitted to 100 cycles of sterilization by immersion in 2% peracetic acid solution for 30 minutes). RESULTS: Chemical analysis using X-ray dispersive energy spectroscopy showed that after autoclave sterilization, only the ICE pliers presented oxidation corrosion (Δ[O] = +24.5%; Δ[Fe] = +5.8%; Δ[WC] = -1.9%). In comparison, following peracetic acid sterilization, both manufacturers ICE (Δ[O] = +1.8%; Δ[Fe] = +18.0%; Δ[WC] = -1.1%) and Quinelato (Δ[O] = +5.3%; Δ[Fe] = -10.4%; Δ[WC] = -15.2%) showed corrosion. The morphological analysis revealed that peracetic acid caused a pitting and localized corrosion in both brands, while the autoclave caused uniform surface corrosion on the ICE pliers. CONCLUSION: Autoclave application was the sterilization method that generated less corrosive damage to the orthodontic cutting pliers, when compared to the immersion in 2% peracetic acid. Dental Press International 2021-10-25 /pmc/articles/PMC8576861/ /pubmed/35640079 http://dx.doi.org/10.1590/2177-6709.26.5.e2119353.oar Text en https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License
spellingShingle Original Article
NUNES, Livia Kelly Ferraz
CARNEIRO, Érika Lopes
PALOMARES, Nathalia Barbosa
SENNA, Lilian
SAMPAIO-FILHO, Helio
MIGUEL, José Augusto Mendes
In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study
title In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study
title_full In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study
title_fullStr In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study
title_full_unstemmed In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study
title_short In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study
title_sort in vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8576861/
https://www.ncbi.nlm.nih.gov/pubmed/35640079
http://dx.doi.org/10.1590/2177-6709.26.5.e2119353.oar
work_keys_str_mv AT nunesliviakellyferraz invitrocomparisonofperaceticacidandautoclavesterilizationinthecorrosionoforthodonticpliersapilotstudy
AT carneiroerikalopes invitrocomparisonofperaceticacidandautoclavesterilizationinthecorrosionoforthodonticpliersapilotstudy
AT palomaresnathaliabarbosa invitrocomparisonofperaceticacidandautoclavesterilizationinthecorrosionoforthodonticpliersapilotstudy
AT sennalilian invitrocomparisonofperaceticacidandautoclavesterilizationinthecorrosionoforthodonticpliersapilotstudy
AT sampaiofilhohelio invitrocomparisonofperaceticacidandautoclavesterilizationinthecorrosionoforthodonticpliersapilotstudy
AT migueljoseaugustomendes invitrocomparisonofperaceticacidandautoclavesterilizationinthecorrosionoforthodonticpliersapilotstudy