Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays

The objective of this study was to evaluate the interrater and intrarater reliability of electrical impedance myography (EIM) using handheld sensors of different sizes. Electrical impedance myography of the biceps brachii muscle of twenty healthy individuals was performed by two raters using both la...

Descripción completa

Detalles Bibliográficos
Autores principales: Hu, Huijing, Lo, Wai Leung Ambrose, Wang, Xiaoyun, Li, Le, Zhou, Ping
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8577940/
https://www.ncbi.nlm.nih.gov/pubmed/34765103
http://dx.doi.org/10.1155/2021/7296322
_version_ 1784596167412154368
author Hu, Huijing
Lo, Wai Leung Ambrose
Wang, Xiaoyun
Li, Le
Zhou, Ping
author_facet Hu, Huijing
Lo, Wai Leung Ambrose
Wang, Xiaoyun
Li, Le
Zhou, Ping
author_sort Hu, Huijing
collection PubMed
description The objective of this study was to evaluate the interrater and intrarater reliability of electrical impedance myography (EIM) using handheld sensors of different sizes. Electrical impedance myography of the biceps brachii muscle of twenty healthy individuals was performed by two raters using both large and small sensors. The procedures were also repeated 5 to 8 days after the first recording session. The repeatability of the resistance, reactance, and phase angle at two different current frequencies (50 and 100 kHz) was assessed by the intraclass correlation coefficient (ICC). The ICCs of the large sensor were higher than those of the small sensor for both the intrarater and interrater reliabilities. High-frequency current tended to improve the ICC for the small sensor. These results indicate reasonable repeatability of the handheld electrode arrays for EIM measurements. The findings suggest that electrode array should be selected appropriately according to the size of the tested muscle.
format Online
Article
Text
id pubmed-8577940
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-85779402021-11-10 Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays Hu, Huijing Lo, Wai Leung Ambrose Wang, Xiaoyun Li, Le Zhou, Ping J Healthc Eng Research Article The objective of this study was to evaluate the interrater and intrarater reliability of electrical impedance myography (EIM) using handheld sensors of different sizes. Electrical impedance myography of the biceps brachii muscle of twenty healthy individuals was performed by two raters using both large and small sensors. The procedures were also repeated 5 to 8 days after the first recording session. The repeatability of the resistance, reactance, and phase angle at two different current frequencies (50 and 100 kHz) was assessed by the intraclass correlation coefficient (ICC). The ICCs of the large sensor were higher than those of the small sensor for both the intrarater and interrater reliabilities. High-frequency current tended to improve the ICC for the small sensor. These results indicate reasonable repeatability of the handheld electrode arrays for EIM measurements. The findings suggest that electrode array should be selected appropriately according to the size of the tested muscle. Hindawi 2021-11-02 /pmc/articles/PMC8577940/ /pubmed/34765103 http://dx.doi.org/10.1155/2021/7296322 Text en Copyright © 2021 Huijing Hu et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Hu, Huijing
Lo, Wai Leung Ambrose
Wang, Xiaoyun
Li, Le
Zhou, Ping
Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays
title Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays
title_full Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays
title_fullStr Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays
title_full_unstemmed Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays
title_short Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays
title_sort interrater and intrarater reliability of electrical impedance myography: a comparison between large and small handheld electrode arrays
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8577940/
https://www.ncbi.nlm.nih.gov/pubmed/34765103
http://dx.doi.org/10.1155/2021/7296322
work_keys_str_mv AT huhuijing interraterandintraraterreliabilityofelectricalimpedancemyographyacomparisonbetweenlargeandsmallhandheldelectrodearrays
AT lowaileungambrose interraterandintraraterreliabilityofelectricalimpedancemyographyacomparisonbetweenlargeandsmallhandheldelectrodearrays
AT wangxiaoyun interraterandintraraterreliabilityofelectricalimpedancemyographyacomparisonbetweenlargeandsmallhandheldelectrodearrays
AT lile interraterandintraraterreliabilityofelectricalimpedancemyographyacomparisonbetweenlargeandsmallhandheldelectrodearrays
AT zhouping interraterandintraraterreliabilityofelectricalimpedancemyographyacomparisonbetweenlargeandsmallhandheldelectrodearrays