Pump position and thrombosis in ventricular assist devices: Correlation of radiographs and CT data
Malpositioning of left ventricular assist devices (LVAD) is a risk factor for thrombosis, but its identification from clinical imaging remains challenging. X-rays and CT scans were analyzed and parameters identified that correlated to pump thrombosis. Retrospective imaging data of patients (n = 115)...
Autores principales: | Aigner, Philipp, Schlöglhofer, Thomas, Plunger, Lea Carmen, Beitzke, Dietrich, Wielandner, Alice, Schima, Heinrich, Zimpfer, Daniel, Moscato, Francesco |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8581720/ https://www.ncbi.nlm.nih.gov/pubmed/34088235 http://dx.doi.org/10.1177/03913988211017552 |
Ejemplares similares
-
The influence of left ventricular assist device inflow cannula position on thrombosis risk
por: Ghodrati, Mojgan, et al.
Publicado: (2020) -
Left atrial appendage occlusion in ventricular assist device patients to decrease thromboembolic events: A computer simulation study
por: Ghodrati-Misek, Mojgan, et al.
Publicado: (2022) -
Inflow cannula position as risk factor for stroke in patients with HeartMate 3 left ventricular assist devices
por: Schlöglhofer, Thomas, et al.
Publicado: (2022) -
Validation of Intrinsic Left Ventricular Assist Device Data Tracking Algorithm for Early Recognition of Centrifugal Flow Pump Thrombosis
por: Gross, Christoph, et al.
Publicado: (2022) -
Effects of the atrium on intraventricular flow patterns during mechanical circulatory support
por: Ghodrati, Mojgan, et al.
Publicado: (2021)