Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study

OBJECTIVES: The purpose of this study was twofold. First, the influence of a novel calcium-aware (Ca-aware) computed tomography (CT) reconstruction technique on coronary artery calcium (CAC) scores surrounded by a variety of tissues was assessed. Second, the performance of the Ca-aware reconstructio...

Descripción completa

Detalles Bibliográficos
Autores principales: van der Werf, Niels R., Booij, Ronald, Schmidt, Bernhard, Flohr, Thomas G., Leiner, Tim, de Groen, Joël J., Bos, Daniël, Budde, Ricardo P. J., Willemink, Martin J., Greuter, Marcel J. W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8589753/
https://www.ncbi.nlm.nih.gov/pubmed/34050386
http://dx.doi.org/10.1007/s00330-021-08076-5
_version_ 1784598798934212608
author van der Werf, Niels R.
Booij, Ronald
Schmidt, Bernhard
Flohr, Thomas G.
Leiner, Tim
de Groen, Joël J.
Bos, Daniël
Budde, Ricardo P. J.
Willemink, Martin J.
Greuter, Marcel J. W.
author_facet van der Werf, Niels R.
Booij, Ronald
Schmidt, Bernhard
Flohr, Thomas G.
Leiner, Tim
de Groen, Joël J.
Bos, Daniël
Budde, Ricardo P. J.
Willemink, Martin J.
Greuter, Marcel J. W.
author_sort van der Werf, Niels R.
collection PubMed
description OBJECTIVES: The purpose of this study was twofold. First, the influence of a novel calcium-aware (Ca-aware) computed tomography (CT) reconstruction technique on coronary artery calcium (CAC) scores surrounded by a variety of tissues was assessed. Second, the performance of the Ca-aware reconstruction technique on moving CAC was evaluated with a dynamic phantom. METHODS: An artificial coronary artery, containing two CAC of equal size and different densities (196 ± 3, 380 ± 2 mg hydroxyapatite cm(−3)), was moved in the center compartment of an anthropomorphic thorax phantom at different heart rates. The center compartment was filled with mixtures, which resembled fat, water, and soft tissue equivalent CT numbers. Raw data was acquired with a routine clinical CAC protocol, at 120 peak kilovolt (kVp). Subsequently, reduced tube voltage (100 kVp) and tin-filtration (150Sn kVp) acquisitions were performed. Raw data was reconstructed with a standard and a novel Ca-aware reconstruction technique. Agatston scores of all reconstructions were compared with the reference (120 kVp) and standard reconstruction technique, with relevant deviations defined as > 10%. RESULTS: For all heart rates, Agatston scores for CAC submerged in fat were comparable to the reference, for the reduced-kVp acquisition with Ca-aware reconstruction kernel. For water and soft tissue, medium-density Agatston scores were again comparable to the reference for all heart rates. Low-density Agatston scores showed relevant deviations, up to 15% and 23% for water and soft tissue, respectively. CONCLUSION: CT CAC scoring with varying surrounding materials and heart rates is feasible at patient-specific tube voltages with the novel Ca-aware reconstruction technique. KEY POINTS: • A dedicated calcium-aware reconstruction kernel results in similar Agatston scores for CAC surrounded by fatty materials regardless of CAC density and heart rate. • Application of a dedicated calcium-aware reconstruction kernel allows for radiation dose reduction. • Mass scores determined with CT underestimated physical mass.
format Online
Article
Text
id pubmed-8589753
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-85897532021-11-15 Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study van der Werf, Niels R. Booij, Ronald Schmidt, Bernhard Flohr, Thomas G. Leiner, Tim de Groen, Joël J. Bos, Daniël Budde, Ricardo P. J. Willemink, Martin J. Greuter, Marcel J. W. Eur Radiol Computed Tomography OBJECTIVES: The purpose of this study was twofold. First, the influence of a novel calcium-aware (Ca-aware) computed tomography (CT) reconstruction technique on coronary artery calcium (CAC) scores surrounded by a variety of tissues was assessed. Second, the performance of the Ca-aware reconstruction technique on moving CAC was evaluated with a dynamic phantom. METHODS: An artificial coronary artery, containing two CAC of equal size and different densities (196 ± 3, 380 ± 2 mg hydroxyapatite cm(−3)), was moved in the center compartment of an anthropomorphic thorax phantom at different heart rates. The center compartment was filled with mixtures, which resembled fat, water, and soft tissue equivalent CT numbers. Raw data was acquired with a routine clinical CAC protocol, at 120 peak kilovolt (kVp). Subsequently, reduced tube voltage (100 kVp) and tin-filtration (150Sn kVp) acquisitions were performed. Raw data was reconstructed with a standard and a novel Ca-aware reconstruction technique. Agatston scores of all reconstructions were compared with the reference (120 kVp) and standard reconstruction technique, with relevant deviations defined as > 10%. RESULTS: For all heart rates, Agatston scores for CAC submerged in fat were comparable to the reference, for the reduced-kVp acquisition with Ca-aware reconstruction kernel. For water and soft tissue, medium-density Agatston scores were again comparable to the reference for all heart rates. Low-density Agatston scores showed relevant deviations, up to 15% and 23% for water and soft tissue, respectively. CONCLUSION: CT CAC scoring with varying surrounding materials and heart rates is feasible at patient-specific tube voltages with the novel Ca-aware reconstruction technique. KEY POINTS: • A dedicated calcium-aware reconstruction kernel results in similar Agatston scores for CAC surrounded by fatty materials regardless of CAC density and heart rate. • Application of a dedicated calcium-aware reconstruction kernel allows for radiation dose reduction. • Mass scores determined with CT underestimated physical mass. Springer Berlin Heidelberg 2021-05-28 2021 /pmc/articles/PMC8589753/ /pubmed/34050386 http://dx.doi.org/10.1007/s00330-021-08076-5 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Computed Tomography
van der Werf, Niels R.
Booij, Ronald
Schmidt, Bernhard
Flohr, Thomas G.
Leiner, Tim
de Groen, Joël J.
Bos, Daniël
Budde, Ricardo P. J.
Willemink, Martin J.
Greuter, Marcel J. W.
Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study
title Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study
title_full Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study
title_fullStr Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study
title_full_unstemmed Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study
title_short Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study
title_sort evaluating a calcium-aware kernel for ct cac scoring with varying surrounding materials and heart rates: a dynamic phantom study
topic Computed Tomography
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8589753/
https://www.ncbi.nlm.nih.gov/pubmed/34050386
http://dx.doi.org/10.1007/s00330-021-08076-5
work_keys_str_mv AT vanderwerfnielsr evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT booijronald evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT schmidtbernhard evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT flohrthomasg evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT leinertim evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT degroenjoelj evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT bosdaniel evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT buddericardopj evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT willeminkmartinj evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy
AT greutermarceljw evaluatingacalciumawarekernelforctcacscoringwithvaryingsurroundingmaterialsandheartratesadynamicphantomstudy