Cargando…

Acromioclavicular joint reconstruction implants have differing ability to restore horizontal and vertical plane stability

PURPOSE: Persistent acromioclavicular joint (ACJ) instability following high grade injuries causes significant symptoms. The importance of horizontal plane stability is increasingly recognised. There is little evidence of the ability of current implant methods to restore native ACJ stability in the...

Descripción completa

Detalles Bibliográficos
Autores principales: Alkoheji, Mohamed, El-Daou, Hadi, Lee, Jillian, Carlos, Adrian, Di Mascio, Livio, Amis, Andrew A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8595167/
https://www.ncbi.nlm.nih.gov/pubmed/34436636
http://dx.doi.org/10.1007/s00167-021-06700-x
Descripción
Sumario:PURPOSE: Persistent acromioclavicular joint (ACJ) instability following high grade injuries causes significant symptoms. The importance of horizontal plane stability is increasingly recognised. There is little evidence of the ability of current implant methods to restore native ACJ stability in the vertical and horizontal planes. The purpose of this work was to measure the ability of three implant reconstructions to restore native ACJ stability. METHODS: Three groups of nine fresh-frozen shoulders each were mounted into a robotic testing system. The scapula was stationary and the robot displaced the clavicle to measure native anterior, posterior, superior and inferior (A, P, S, I) stability at 50 N force. The ACJ capsule, conoid and trapezoid ligaments were transected and the ACJ was reconstructed using one of three commercially available systems. Two systems (tape loop + screw and tape loop + button) wrapped a tape around the clavicle and coracoid, the third system (sutures + buttons) passed directly through tunnels in the clavicle and coracoid. The stabilities were remeasured. The data for A, P, S, I stability and ranges of A–P and S–I stability were analyzed by ANOVA and repeated-measures Student t tests with Bonferroni correction, to contrast each reconstruction stability versus the native ACJ data for that set of nine specimens, and examined contrasts among the reconstructions. RESULTS: All three reconstructions restored the range of A–P stability to that of the native ACJ. However, the coracoid loop devices shifted the clavicle anteriorly. For S–I stability, only the sutures + buttons reconstruction did not differ significantly from native ligament restraint. CONCLUSIONS: Only the sutures + buttons reconstruction, that passed directly through tunnels in the clavicle and coracoid, restored all stability measures (A, P, S, I) to the native values, while the tape implants wrapped around the bones anteriorised the clavicle. These findings show differing abilities among reconstructions to restore native stability in horizontal and vertical planes. (300 words)