Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series

BACKGROUND: The aim of this study was to assess the feasibility and diagnostic accuracy of an optimized (111)Indium‐diethylenetriamine‐penta‐acetic‐acid single‐photon‐emission computed tomography (CT) ((111)In‐DTPA SPECT‐CT) examination in patients with suspected intrathecal drug delivery (ITDD) fai...

Descripción completa

Detalles Bibliográficos
Autores principales: Delhaas, Elmar M., van Assema, Daniëlle M.E., Fröberg, Alida C., Zwezerijnen, Ben G.J.C., Harhangi, Biswadjiet S., Frankema, Sander P.G., Huygen, Frank J.P.M., van der Lugt, Aad
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley & Sons, Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8596779/
https://www.ncbi.nlm.nih.gov/pubmed/32946171
http://dx.doi.org/10.1111/ner.13275
_version_ 1784600463041101824
author Delhaas, Elmar M.
van Assema, Daniëlle M.E.
Fröberg, Alida C.
Zwezerijnen, Ben G.J.C.
Harhangi, Biswadjiet S.
Frankema, Sander P.G.
Huygen, Frank J.P.M.
van der Lugt, Aad
author_facet Delhaas, Elmar M.
van Assema, Daniëlle M.E.
Fröberg, Alida C.
Zwezerijnen, Ben G.J.C.
Harhangi, Biswadjiet S.
Frankema, Sander P.G.
Huygen, Frank J.P.M.
van der Lugt, Aad
author_sort Delhaas, Elmar M.
collection PubMed
description BACKGROUND: The aim of this study was to assess the feasibility and diagnostic accuracy of an optimized (111)Indium‐diethylenetriamine‐penta‐acetic‐acid single‐photon‐emission computed tomography (CT) ((111)In‐DTPA SPECT‐CT) examination in patients with suspected intrathecal drug delivery (ITDD) failure. MATERIALS AND METHODS: Retrospective analysis of routinely collected observational data from a case series of patients in the setting of the academic Center for Pain Medicine, Departments of Radiology and Nuclear Medicine and Neurosurgery. Twenty‐seven patients participated between January 2014 and January 2019. Thirty‐six optimized examinations including standardized pump flow rate with additional SPECT‐CT imaging and a stepwise standardized analysis were performed. A 10 mL mixture of medication and 20 MBq 111In‐DTPA was injected into the pump reservoir. Planar and SPECT‐CT images were acquired at 24, 48, and 72 hours (h) after injection and at 96 hours and/or seven days, if needed. All images were reassessed by the first two authors using an optimized procedure. RESULTS AND CONCLUSIONS: Twenty‐two abnormalities were identified in 21 examinations, with these abnormalities consisting of leakage (n = 7), spinal catheter obstruction (n = 7), and cerebrospinal fluid flow obstruction (n = 8). Interventions (n = 19) confirmed the cause of ITDD failure. A false‐positive finding at follow‐up (n = 1) and a false‐negative finding (n = 1) were encountered. Sensitivity was 95% (20/21) and the specificity 93% (14/15). A significant difference (p < 0.001) was found between the accuracy of the conventical and the optimized analysis. The optimized (111)In‐DTPA SPECT‐CT examination is a powerful diagnostic tool for detecting the cause of ITDD failure.
format Online
Article
Text
id pubmed-8596779
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley & Sons, Inc.
record_format MEDLINE/PubMed
spelling pubmed-85967792021-11-22 Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series Delhaas, Elmar M. van Assema, Daniëlle M.E. Fröberg, Alida C. Zwezerijnen, Ben G.J.C. Harhangi, Biswadjiet S. Frankema, Sander P.G. Huygen, Frank J.P.M. van der Lugt, Aad Neuromodulation INTRATHECAL DRUG DELIVERY BACKGROUND: The aim of this study was to assess the feasibility and diagnostic accuracy of an optimized (111)Indium‐diethylenetriamine‐penta‐acetic‐acid single‐photon‐emission computed tomography (CT) ((111)In‐DTPA SPECT‐CT) examination in patients with suspected intrathecal drug delivery (ITDD) failure. MATERIALS AND METHODS: Retrospective analysis of routinely collected observational data from a case series of patients in the setting of the academic Center for Pain Medicine, Departments of Radiology and Nuclear Medicine and Neurosurgery. Twenty‐seven patients participated between January 2014 and January 2019. Thirty‐six optimized examinations including standardized pump flow rate with additional SPECT‐CT imaging and a stepwise standardized analysis were performed. A 10 mL mixture of medication and 20 MBq 111In‐DTPA was injected into the pump reservoir. Planar and SPECT‐CT images were acquired at 24, 48, and 72 hours (h) after injection and at 96 hours and/or seven days, if needed. All images were reassessed by the first two authors using an optimized procedure. RESULTS AND CONCLUSIONS: Twenty‐two abnormalities were identified in 21 examinations, with these abnormalities consisting of leakage (n = 7), spinal catheter obstruction (n = 7), and cerebrospinal fluid flow obstruction (n = 8). Interventions (n = 19) confirmed the cause of ITDD failure. A false‐positive finding at follow‐up (n = 1) and a false‐negative finding (n = 1) were encountered. Sensitivity was 95% (20/21) and the specificity 93% (14/15). A significant difference (p < 0.001) was found between the accuracy of the conventical and the optimized analysis. The optimized (111)In‐DTPA SPECT‐CT examination is a powerful diagnostic tool for detecting the cause of ITDD failure. John Wiley & Sons, Inc. 2020-09-18 2021-10 /pmc/articles/PMC8596779/ /pubmed/32946171 http://dx.doi.org/10.1111/ner.13275 Text en © 2020 The Authors. Neuromodulation: Technology at the Neural Interface published by Wiley Periodicals LLC on behalf of International Neuromodulation Society. https://creativecommons.org/licenses/by-nc/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle INTRATHECAL DRUG DELIVERY
Delhaas, Elmar M.
van Assema, Daniëlle M.E.
Fröberg, Alida C.
Zwezerijnen, Ben G.J.C.
Harhangi, Biswadjiet S.
Frankema, Sander P.G.
Huygen, Frank J.P.M.
van der Lugt, Aad
Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series
title Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series
title_full Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series
title_fullStr Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series
title_full_unstemmed Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series
title_short Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single‐Institution Case Series
title_sort isotopic scintigraphy in intrathecal drug delivery failure: a single‐institution case series
topic INTRATHECAL DRUG DELIVERY
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8596779/
https://www.ncbi.nlm.nih.gov/pubmed/32946171
http://dx.doi.org/10.1111/ner.13275
work_keys_str_mv AT delhaaselmarm isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries
AT vanassemadanielleme isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries
AT frobergalidac isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries
AT zwezerijnenbengjc isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries
AT harhangibiswadjiets isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries
AT frankemasanderpg isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries
AT huygenfrankjpm isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries
AT vanderlugtaad isotopicscintigraphyinintrathecaldrugdeliveryfailureasingleinstitutioncaseseries