A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults
BACKGROUND: Although knee extensors are essential in daily activities (e.g. walking, climbing stairs), knee extensor strength is often not measured in clinical settings. Existing devices to test muscle strength are not always suitable to accurately measure the high forces of this muscle group. There...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8602994/ https://www.ncbi.nlm.nih.gov/pubmed/34798859 http://dx.doi.org/10.1186/s12891-021-04848-8 |
_version_ | 1784601681798889472 |
---|---|
author | Hartog, Johanneke Dijkstra, Sandra Fleer, Joke van der Harst, Pim Mariani, Massimo A. van der Woude, Lucas H. V. |
author_facet | Hartog, Johanneke Dijkstra, Sandra Fleer, Joke van der Harst, Pim Mariani, Massimo A. van der Woude, Lucas H. V. |
author_sort | Hartog, Johanneke |
collection | PubMed |
description | BACKGROUND: Although knee extensors are essential in daily activities (e.g. walking, climbing stairs), knee extensor strength is often not measured in clinical settings. Existing devices to test muscle strength are not always suitable to accurately measure the high forces of this muscle group. Therefore, a device to test muscle strength that is convenient, feasible, reliable, and valid in clinical settings is required. This study evaluated the reliability, responsiveness, and level of discomfort of the newly developed Q-Force ӀӀ (i.e. a portable device to measure isometric knee extensor strength) in healthy middle-aged and elderly adults. METHODS: Participants (n = 22) conducted two standardized test sessions on the Q-Force ӀӀ (five to ten days apart). Each session consisted of one familiarisation trial followed by three trials of peak isometric knee extension per each leg. Per trial, peak and mean knee extension force (N) and torque (Nm) were measured at 90° flexion. The level of discomfort was determined using a visual analog scale (VAS: 0-100). Intra Class Correlation (ICC, model: two-way mixed with absolute agreement), Standard Error of Measurement (SEM), and minimal detectable change (MDC) were determined. A repeated measures ANOVA was used to determine between-test variation. RESULTS: Excellent test-retest (ICC > 0.95) and inter-trial (ICC > 0.91) reliability for both legs were shown. No significant differences were found in peak and mean knee forces and torques between test and retest of both legs, indicating good test-retest reliability (P-value range: 0.360-0.538; F(1,21) range: 0.4-0.9). The SEM of the peak and mean forces and torques ranged from 28.0 to 30.4 N (6.0-6.8%) and from 9.2 to 10.4 Nm (6.4-7.7%), respectively. The MDC for these outcomes ranged respectively from 77.6 to 84.1 N (16.5-18.8%) and from 25.5 to 28.9 Nm (17.6-21.4%). The level of discomfort was low (median range: 7-10, IQR: 4-18). CONCLUSION: The portable Q-Force ӀӀ is a comfortable, responsive, and relatively cheap device with excellent test-retest reliability. This device would be potentially suitable to measure isometric knee extensor strength in clinical settings. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12891-021-04848-8. |
format | Online Article Text |
id | pubmed-8602994 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-86029942021-11-19 A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults Hartog, Johanneke Dijkstra, Sandra Fleer, Joke van der Harst, Pim Mariani, Massimo A. van der Woude, Lucas H. V. BMC Musculoskelet Disord Research BACKGROUND: Although knee extensors are essential in daily activities (e.g. walking, climbing stairs), knee extensor strength is often not measured in clinical settings. Existing devices to test muscle strength are not always suitable to accurately measure the high forces of this muscle group. Therefore, a device to test muscle strength that is convenient, feasible, reliable, and valid in clinical settings is required. This study evaluated the reliability, responsiveness, and level of discomfort of the newly developed Q-Force ӀӀ (i.e. a portable device to measure isometric knee extensor strength) in healthy middle-aged and elderly adults. METHODS: Participants (n = 22) conducted two standardized test sessions on the Q-Force ӀӀ (five to ten days apart). Each session consisted of one familiarisation trial followed by three trials of peak isometric knee extension per each leg. Per trial, peak and mean knee extension force (N) and torque (Nm) were measured at 90° flexion. The level of discomfort was determined using a visual analog scale (VAS: 0-100). Intra Class Correlation (ICC, model: two-way mixed with absolute agreement), Standard Error of Measurement (SEM), and minimal detectable change (MDC) were determined. A repeated measures ANOVA was used to determine between-test variation. RESULTS: Excellent test-retest (ICC > 0.95) and inter-trial (ICC > 0.91) reliability for both legs were shown. No significant differences were found in peak and mean knee forces and torques between test and retest of both legs, indicating good test-retest reliability (P-value range: 0.360-0.538; F(1,21) range: 0.4-0.9). The SEM of the peak and mean forces and torques ranged from 28.0 to 30.4 N (6.0-6.8%) and from 9.2 to 10.4 Nm (6.4-7.7%), respectively. The MDC for these outcomes ranged respectively from 77.6 to 84.1 N (16.5-18.8%) and from 25.5 to 28.9 Nm (17.6-21.4%). The level of discomfort was low (median range: 7-10, IQR: 4-18). CONCLUSION: The portable Q-Force ӀӀ is a comfortable, responsive, and relatively cheap device with excellent test-retest reliability. This device would be potentially suitable to measure isometric knee extensor strength in clinical settings. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12891-021-04848-8. BioMed Central 2021-11-19 /pmc/articles/PMC8602994/ /pubmed/34798859 http://dx.doi.org/10.1186/s12891-021-04848-8 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Hartog, Johanneke Dijkstra, Sandra Fleer, Joke van der Harst, Pim Mariani, Massimo A. van der Woude, Lucas H. V. A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults |
title | A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults |
title_full | A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults |
title_fullStr | A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults |
title_full_unstemmed | A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults |
title_short | A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults |
title_sort | portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the q-force ӏӏ in healthy adults |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8602994/ https://www.ncbi.nlm.nih.gov/pubmed/34798859 http://dx.doi.org/10.1186/s12891-021-04848-8 |
work_keys_str_mv | AT hartogjohanneke aportableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT dijkstrasandra aportableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT fleerjoke aportableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT vanderharstpim aportableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT marianimassimoa aportableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT vanderwoudelucashv aportableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT hartogjohanneke portableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT dijkstrasandra portableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT fleerjoke portableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT vanderharstpim portableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT marianimassimoa portableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults AT vanderwoudelucashv portableisometrickneeextensorstrengthtestingdevicetestretestreliabilityandminimaldetectablechangescoresoftheqforceӏӏinhealthyadults |