Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study
BACKGROUND: Atrial fibrillation (AF) is the most prevalent cardiac dysrhythmia with high morbidity and mortality rate. Evidence shows that in every three patients with AF, one is asymptomatic. The asymptomatic and paroxysmal nature of AF is the reason for unsatisfactory and delayed detection using t...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8606080/ https://www.ncbi.nlm.nih.gov/pubmed/34800984 http://dx.doi.org/10.1186/s12872-021-02363-1 |
_version_ | 1784602278186975232 |
---|---|
author | Fu, Wenxia Li, Ruogu |
author_facet | Fu, Wenxia Li, Ruogu |
author_sort | Fu, Wenxia |
collection | PubMed |
description | BACKGROUND: Atrial fibrillation (AF) is the most prevalent cardiac dysrhythmia with high morbidity and mortality rate. Evidence shows that in every three patients with AF, one is asymptomatic. The asymptomatic and paroxysmal nature of AF is the reason for unsatisfactory and delayed detection using traditional instruments. Research indicates that wearing a dynamic electrocardiogram (ECG) recorder can guide accurate and safe analysis, interpretation, and distinction of AF from normal sinus rhythm. This is also achievable in an upright position and after exercises, assisted by an artificial intelligence (AI) algorithm. METHODS: This study enrolled 114 participants from the outpatient registry of our institution from June 24, 2020 to July 24, 2020. Participants were tested with a wearable dynamic ECG recorder and 12-lead ECG in a supine, an upright position and after exercises for 60 s. RESULTS: Of the 114 subjects enrolled in the study, 61 had normal sinus rhythm and 53 had AF. The number of cases that could not be determined by the wristband of dynamic ECG recorder was two, one and one respectively. Case results that were not clinically objective were defined as false-negative or false-positive. Results for diagnostic accuracy, sensitivity, and specificity tested by wearable dynamic ECG recorders in a supine position were 94.74% (95% CI% 88.76–97.80%), 88.68% (95% CI 77.06–95.07%), and 100% (95% CI 92.91–100%), respectively. Meanwhile, the diagnostic accuracy, sensitivity and specificity in an upright position were 97.37% (95% CI 92.21–99.44%), 94.34% (95% CI 84.03–98.65%), and 100% (95% CI 92.91–100%), respectively. Similar results as those of the upright position were obtained after exercise. CONCLUSION: The widely accessible wearable dynamic ECG recorder integrated with an AI algorithm can efficiently detect AF in different postures and after exercises. As such, this tool holds great promise as a useful and user-friendly screening method for timely AF diagnosis in at-risk individuals. |
format | Online Article Text |
id | pubmed-8606080 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-86060802021-11-22 Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study Fu, Wenxia Li, Ruogu BMC Cardiovasc Disord Research Article BACKGROUND: Atrial fibrillation (AF) is the most prevalent cardiac dysrhythmia with high morbidity and mortality rate. Evidence shows that in every three patients with AF, one is asymptomatic. The asymptomatic and paroxysmal nature of AF is the reason for unsatisfactory and delayed detection using traditional instruments. Research indicates that wearing a dynamic electrocardiogram (ECG) recorder can guide accurate and safe analysis, interpretation, and distinction of AF from normal sinus rhythm. This is also achievable in an upright position and after exercises, assisted by an artificial intelligence (AI) algorithm. METHODS: This study enrolled 114 participants from the outpatient registry of our institution from June 24, 2020 to July 24, 2020. Participants were tested with a wearable dynamic ECG recorder and 12-lead ECG in a supine, an upright position and after exercises for 60 s. RESULTS: Of the 114 subjects enrolled in the study, 61 had normal sinus rhythm and 53 had AF. The number of cases that could not be determined by the wristband of dynamic ECG recorder was two, one and one respectively. Case results that were not clinically objective were defined as false-negative or false-positive. Results for diagnostic accuracy, sensitivity, and specificity tested by wearable dynamic ECG recorders in a supine position were 94.74% (95% CI% 88.76–97.80%), 88.68% (95% CI 77.06–95.07%), and 100% (95% CI 92.91–100%), respectively. Meanwhile, the diagnostic accuracy, sensitivity and specificity in an upright position were 97.37% (95% CI 92.21–99.44%), 94.34% (95% CI 84.03–98.65%), and 100% (95% CI 92.91–100%), respectively. Similar results as those of the upright position were obtained after exercise. CONCLUSION: The widely accessible wearable dynamic ECG recorder integrated with an AI algorithm can efficiently detect AF in different postures and after exercises. As such, this tool holds great promise as a useful and user-friendly screening method for timely AF diagnosis in at-risk individuals. BioMed Central 2021-11-20 /pmc/articles/PMC8606080/ /pubmed/34800984 http://dx.doi.org/10.1186/s12872-021-02363-1 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Article Fu, Wenxia Li, Ruogu Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study |
title | Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study |
title_full | Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study |
title_fullStr | Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study |
title_full_unstemmed | Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study |
title_short | Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study |
title_sort | diagnostic performance of a wearing dynamic ecg recorder for atrial fibrillation screening: the huami heart study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8606080/ https://www.ncbi.nlm.nih.gov/pubmed/34800984 http://dx.doi.org/10.1186/s12872-021-02363-1 |
work_keys_str_mv | AT fuwenxia diagnosticperformanceofawearingdynamicecgrecorderforatrialfibrillationscreeningthehuamiheartstudy AT liruogu diagnosticperformanceofawearingdynamicecgrecorderforatrialfibrillationscreeningthehuamiheartstudy |