Cargando…

Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study

BACKGROUND: To evaluate the correlation and agreement between superb micro-vascular imaging (SMI) mode and the contrast-enhanced ultrasound (CEUS) mode for the ablative completeness and the volumes of ablation lesions to determine the clinical application value of SMI in follow-up after radiofrequen...

Descripción completa

Detalles Bibliográficos
Autores principales: Lan, Yu, Li, Nan, Song, Qing, Zhang, Ming-bo, Luo, Yu-kun, Zhang, Yan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8609811/
https://www.ncbi.nlm.nih.gov/pubmed/34809604
http://dx.doi.org/10.1186/s12880-021-00697-y
_version_ 1784602987011768320
author Lan, Yu
Li, Nan
Song, Qing
Zhang, Ming-bo
Luo, Yu-kun
Zhang, Yan
author_facet Lan, Yu
Li, Nan
Song, Qing
Zhang, Ming-bo
Luo, Yu-kun
Zhang, Yan
author_sort Lan, Yu
collection PubMed
description BACKGROUND: To evaluate the correlation and agreement between superb micro-vascular imaging (SMI) mode and the contrast-enhanced ultrasound (CEUS) mode for the ablative completeness and the volumes of ablation lesions to determine the clinical application value of SMI in follow-up after radiofrequency ablation. METHODS: From April 2020 to June 2020, two radiologists used SMI and CEUS mode to measure the volume of the ablation lesion. We use intra-class correlation coefficient (ICC), scatter plots and Bland–Altman plots to evaluate the correlation and agreement of the two techniques. In addition, intra- and inter-observer reliability in volume measurement of ablation lesions with SMI mode was assessed. RESULTS: SMI mode and CEUS mode have good agreement in the evaluation of ablative completeness. The ICC was 0.876 and 0.928 of reader A and reader B between SMI mode and CEUS mode in terms of ablation lesions volume measurement. There was a strong correlation between the two modes in both reader A and reader B (r(A) = 0.808; r(B) = 0.882). The ICC was 0.836 for the inter-observer reliability of SMI technique. The scatter plot showed a good linear relation (r = 0.715). In the Bland–Altman plot, 4.35% (1/23) of the points was outside the 95% limits of agreement. The ICC was 0.965 for the intra-observer reliability of SMI technique, the scatter plot also showed a strong linear correlation (r = 0.965). In the Bland–Altman plot, 8.70% (2/23) of the points was outside the 95% limits of agreement. CONCLUSIONS: SMI and CEUS have good agreement and correlation in the ablation volume measurement. SMI technology is expected to be applied as an alternative to CEUS in the clinical follow-up of ablation lesions.
format Online
Article
Text
id pubmed-8609811
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-86098112021-11-29 Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study Lan, Yu Li, Nan Song, Qing Zhang, Ming-bo Luo, Yu-kun Zhang, Yan BMC Med Imaging Research BACKGROUND: To evaluate the correlation and agreement between superb micro-vascular imaging (SMI) mode and the contrast-enhanced ultrasound (CEUS) mode for the ablative completeness and the volumes of ablation lesions to determine the clinical application value of SMI in follow-up after radiofrequency ablation. METHODS: From April 2020 to June 2020, two radiologists used SMI and CEUS mode to measure the volume of the ablation lesion. We use intra-class correlation coefficient (ICC), scatter plots and Bland–Altman plots to evaluate the correlation and agreement of the two techniques. In addition, intra- and inter-observer reliability in volume measurement of ablation lesions with SMI mode was assessed. RESULTS: SMI mode and CEUS mode have good agreement in the evaluation of ablative completeness. The ICC was 0.876 and 0.928 of reader A and reader B between SMI mode and CEUS mode in terms of ablation lesions volume measurement. There was a strong correlation between the two modes in both reader A and reader B (r(A) = 0.808; r(B) = 0.882). The ICC was 0.836 for the inter-observer reliability of SMI technique. The scatter plot showed a good linear relation (r = 0.715). In the Bland–Altman plot, 4.35% (1/23) of the points was outside the 95% limits of agreement. The ICC was 0.965 for the intra-observer reliability of SMI technique, the scatter plot also showed a strong linear correlation (r = 0.965). In the Bland–Altman plot, 8.70% (2/23) of the points was outside the 95% limits of agreement. CONCLUSIONS: SMI and CEUS have good agreement and correlation in the ablation volume measurement. SMI technology is expected to be applied as an alternative to CEUS in the clinical follow-up of ablation lesions. BioMed Central 2021-11-22 /pmc/articles/PMC8609811/ /pubmed/34809604 http://dx.doi.org/10.1186/s12880-021-00697-y Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Lan, Yu
Li, Nan
Song, Qing
Zhang, Ming-bo
Luo, Yu-kun
Zhang, Yan
Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study
title Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study
title_full Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study
title_fullStr Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study
title_full_unstemmed Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study
title_short Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study
title_sort correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8609811/
https://www.ncbi.nlm.nih.gov/pubmed/34809604
http://dx.doi.org/10.1186/s12880-021-00697-y
work_keys_str_mv AT lanyu correlationandagreementbetweensuperbmicrovascularimagingandcontrastenhancedultrasoundforassessingradiofrequencyablationtreatmentofthyroidnodulesapreliminarystudy
AT linan correlationandagreementbetweensuperbmicrovascularimagingandcontrastenhancedultrasoundforassessingradiofrequencyablationtreatmentofthyroidnodulesapreliminarystudy
AT songqing correlationandagreementbetweensuperbmicrovascularimagingandcontrastenhancedultrasoundforassessingradiofrequencyablationtreatmentofthyroidnodulesapreliminarystudy
AT zhangmingbo correlationandagreementbetweensuperbmicrovascularimagingandcontrastenhancedultrasoundforassessingradiofrequencyablationtreatmentofthyroidnodulesapreliminarystudy
AT luoyukun correlationandagreementbetweensuperbmicrovascularimagingandcontrastenhancedultrasoundforassessingradiofrequencyablationtreatmentofthyroidnodulesapreliminarystudy
AT zhangyan correlationandagreementbetweensuperbmicrovascularimagingandcontrastenhancedultrasoundforassessingradiofrequencyablationtreatmentofthyroidnodulesapreliminarystudy