Cargando…
The Improved Method for Determination of Orotic Acid in Milk by Ultra-Fast Liquid Chromatography with Optimized Photodiode Array Detection
SIMPLE SUMMARY: Liquid chromatography for determination of orotic acid (OAc) in milk of ewes and cows is presented. Acetonitrile was added to collected milk samples; then, the resulting mixture was centrifuged. Water was added to the obtained supernatant. Finally, the obtained solution was injected...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8614319/ https://www.ncbi.nlm.nih.gov/pubmed/34827928 http://dx.doi.org/10.3390/ani11113196 |
Sumario: | SIMPLE SUMMARY: Liquid chromatography for determination of orotic acid (OAc) in milk of ewes and cows is presented. Acetonitrile was added to collected milk samples; then, the resulting mixture was centrifuged. Water was added to the obtained supernatant. Finally, the obtained solution was injected into two analytical columns. All analyses were performed at a column temperature of 35 °C. Suitable separation of OAc from endogenous species of milk can be achieved using the gradient elution program and UV detection. The current chromatographic procedure resulted in satisfactory precision, accuracy and sensitivity of OAc analyses in milk samples; OAc eluted at ca. 6.4 min. Our improved chromatographic method is suitable for routine determination of OAc in milk of sheep and cows. ABSTRACT: Ultra-fast liquid chromatography (UFLC) with a photodiode array detector (DAD) for simple and rapid determination of orotic acid (OAc) in milk of sheep and cows is described. Milk samples are treated with acetonitrile (1:1, v/v) and then centrifuged at 4 °C. To 1 mL of the obtained supernatant 9 mL of ultrapure water was added. Subsequently, 0.5–6 µL of the resulting solution was injected into the UFLC-DAD system. Separation and quantification of OAc in milk samples was achieved using two Kinetex C18 columns (1.7 µm, 150 mm × 2.1 mm, i.d., 100 Å; Phenomenex) fitted with a pre-column of 4 mm × 2 mm, i.d. (Phenomenex) containing C(18) packing material. All separations were performed at a column temperature of 35 °C while the ambient temperature was 21–24 °C. Satisfactory separation of OAc from endogenous species of milk can be achieved using the binary gradient elution program and UV detection at wavelengths 278 nm. Our original procedure resulted in suitable separation and quantification of OAc in milk samples; OAc eluted at 6.44 ± 0.03 min. The total run time of OAc analysis (including re-equilibration) was 27 min. As expected, the OAc peak was absent from the blank when the proposed gradient elution program and UV detection at 278 nm was used. The average recoveries of OAc standards added to milk samples were satisfactory (96.7–105.3%). The low inter-and intra-assay coefficient of variation derived from the measurements of OAc in cow and ovine milk samples (i.e., 0.784%, 1.283% and 0.710%, 1.221%, respectively) and in O-Ac standards (i.e., 0.377% and 0.294%, respectively), as well as high recoveries of OAc added to ovine and cows’ milk (~100%) and the low detection (0.04 ng) and quantification (0.12 ng) limits point to satisfactory accuracy, precision and sensitivity of the reported method. OAc concentrations in ovine milk samples were within the range from 25 to 36 mg/L, while OAc levels in cows’ milk samples was found in the range of 32–36 mg/L. Our original procedure is suitable for routine quantification of OAc in milk of ewes and cows. |
---|