Iterative reconstruction improves image quality and reduces radiation dose in trauma protocols; A human cadaver study

BACKGROUND: Radiation-related cancer risk is an object of concern in CT of trauma patients, as these represent a young population. Different radiation reducing methods, including iterative reconstruction (IR), and spilt bolus techniques have been introduced in the recent years in different large sca...

Descripción completa

Detalles Bibliográficos
Autores principales: Godt, Johannes Clemens, Johansen, Cathrine K, Martinsen, Anne Catrine T, Schulz, Anselm, Brøgger, Helga M, Jensen, Kristin, Stray-Pedersen, Arne, Dormagen, Johann Baptist
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8619783/
https://www.ncbi.nlm.nih.gov/pubmed/34840815
http://dx.doi.org/10.1177/20584601211055389
Descripción
Sumario:BACKGROUND: Radiation-related cancer risk is an object of concern in CT of trauma patients, as these represent a young population. Different radiation reducing methods, including iterative reconstruction (IR), and spilt bolus techniques have been introduced in the recent years in different large scale trauma centers. PURPOSE: To compare image quality in human cadaver exposed to thoracoabdominal computed tomography using IR and standard filtered back-projection (FBP) at different dose levels. MATERIAL AND METHODS: Ten cadavers were scanned at full dose and a dose reduction in CTDIvol of 5 mGy (low dose 1) and 7.5 mGy (low dose 2) on a Siemens Definition Flash 128-slice computed tomography scanner. Low dose images were reconstructed with FBP and Sinogram affirmed iterative reconstruction (SAFIRE) level 2 and 4. Quantitative image quality was analyzed by comparison of contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR). Qualitative image quality was evaluated by use of visual grading regression (VGR) by four radiologists. RESULTS: Readers preferred SAFIRE reconstructed images over FBP at a dose reduction of 40% (low dose 1) and 56% (low dose 2), with significant difference in overall impression of image quality. CNR and SNR showed significant improvement for images reconstructed with SAFIRE 2 and 4 compared to FBP at both low dose levels. CONCLUSIONS: Iterative image reconstruction, SAFIRE 2 and 4, resulted in equal or improved image quality at a dose reduction of up to 56% compared to full dose FBP and may be used a strong radiation reduction tool in the young trauma population.