Cargando…

Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software

Background: Myocardial blood flow (MBF) and flow reserve (MFR) examination, especially useful in the diagnosis of multivessel coronary artery disease (CAD), can be assessed with a cadmium-zinc-telluride (CZT) SPECT gamma camera, as an alternative to the expensive and less available PET. However, stu...

Descripción completa

Detalles Bibliográficos
Autores principales: Cichocki, Paweł, Błaszczyk, Michał, Cygulska, Kamila, Filipczak, Krzysztof, Adamczewski, Zbigniew, Kuśmierek, Jacek, Lipiec, Piotr, Kasprzak, Jarosław Damian, Płachcińska, Anna
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8624690/
https://www.ncbi.nlm.nih.gov/pubmed/34834516
http://dx.doi.org/10.3390/jpm11111164
_version_ 1784606236073787392
author Cichocki, Paweł
Błaszczyk, Michał
Cygulska, Kamila
Filipczak, Krzysztof
Adamczewski, Zbigniew
Kuśmierek, Jacek
Lipiec, Piotr
Kasprzak, Jarosław Damian
Płachcińska, Anna
author_facet Cichocki, Paweł
Błaszczyk, Michał
Cygulska, Kamila
Filipczak, Krzysztof
Adamczewski, Zbigniew
Kuśmierek, Jacek
Lipiec, Piotr
Kasprzak, Jarosław Damian
Płachcińska, Anna
author_sort Cichocki, Paweł
collection PubMed
description Background: Myocardial blood flow (MBF) and flow reserve (MFR) examination, especially useful in the diagnosis of multivessel coronary artery disease (CAD), can be assessed with a cadmium-zinc-telluride (CZT) SPECT gamma camera, as an alternative to the expensive and less available PET. However, study processing is not free from subjective factors. Therefore, this paper aims to evaluate intra- and interobserver repeatability of MBF and MFR values obtained by the same operator and two independent operators. Methods: This study included 57 adult patients. MBF and MFR were assessed using a Discovery NM530c camera in a two-day, rest/dipyridamople protocol, using (99m)Tc-MIBI. Data were processed using Corridor4DM software, twice by one operator and once by another operator. Results: The repeatability of the assessed values was quite good in the whole myocardium, LAD and LCX vascular territories, but was poor in the RCA territory. Conclusions: The poor repeatability of MBF and MFR in RCA vascular territory can be explained by poor automatic orientation of the heart axis during post-processing and a so-called “cardiac creep” phenomenon. Better automatic heart orientation and introduction of automatic motion correction is likely to drastically improve this repeatability. In the present state of the software, PET is better for patients requiring assessment of MFR in the RCA territory.
format Online
Article
Text
id pubmed-8624690
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-86246902021-11-27 Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software Cichocki, Paweł Błaszczyk, Michał Cygulska, Kamila Filipczak, Krzysztof Adamczewski, Zbigniew Kuśmierek, Jacek Lipiec, Piotr Kasprzak, Jarosław Damian Płachcińska, Anna J Pers Med Article Background: Myocardial blood flow (MBF) and flow reserve (MFR) examination, especially useful in the diagnosis of multivessel coronary artery disease (CAD), can be assessed with a cadmium-zinc-telluride (CZT) SPECT gamma camera, as an alternative to the expensive and less available PET. However, study processing is not free from subjective factors. Therefore, this paper aims to evaluate intra- and interobserver repeatability of MBF and MFR values obtained by the same operator and two independent operators. Methods: This study included 57 adult patients. MBF and MFR were assessed using a Discovery NM530c camera in a two-day, rest/dipyridamople protocol, using (99m)Tc-MIBI. Data were processed using Corridor4DM software, twice by one operator and once by another operator. Results: The repeatability of the assessed values was quite good in the whole myocardium, LAD and LCX vascular territories, but was poor in the RCA territory. Conclusions: The poor repeatability of MBF and MFR in RCA vascular territory can be explained by poor automatic orientation of the heart axis during post-processing and a so-called “cardiac creep” phenomenon. Better automatic heart orientation and introduction of automatic motion correction is likely to drastically improve this repeatability. In the present state of the software, PET is better for patients requiring assessment of MFR in the RCA territory. MDPI 2021-11-09 /pmc/articles/PMC8624690/ /pubmed/34834516 http://dx.doi.org/10.3390/jpm11111164 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Cichocki, Paweł
Błaszczyk, Michał
Cygulska, Kamila
Filipczak, Krzysztof
Adamczewski, Zbigniew
Kuśmierek, Jacek
Lipiec, Piotr
Kasprzak, Jarosław Damian
Płachcińska, Anna
Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software
title Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software
title_full Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software
title_fullStr Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software
title_full_unstemmed Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software
title_short Inter- and Intraobserver Repeatability of Myocardial Flow Reserve Values Determined with SPECT Study Using a Discovery NM530c Camera and Corridor 4DM Software
title_sort inter- and intraobserver repeatability of myocardial flow reserve values determined with spect study using a discovery nm530c camera and corridor 4dm software
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8624690/
https://www.ncbi.nlm.nih.gov/pubmed/34834516
http://dx.doi.org/10.3390/jpm11111164
work_keys_str_mv AT cichockipaweł interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT błaszczykmichał interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT cygulskakamila interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT filipczakkrzysztof interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT adamczewskizbigniew interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT kusmierekjacek interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT lipiecpiotr interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT kasprzakjarosławdamian interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware
AT płachcinskaanna interandintraobserverrepeatabilityofmyocardialflowreservevaluesdeterminedwithspectstudyusingadiscoverynm530ccameraandcorridor4dmsoftware