Effect of head size and rotation on taper corrosion in a hip simulator

AIMS: This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. METHODS: In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were...

Descripción completa

Detalles Bibliográficos
Autores principales: Wight, Christian M., Whyne, Cari M., Bogoch, Earl R., Zdero, Radovan, Chapman, Ryan M., van Citters, Douglas W., Walsh, William R., Schemitsch, Emil
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Editorial Society of Bone & Joint Surgery 2021
Materias:
Hip
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8636299/
https://www.ncbi.nlm.nih.gov/pubmed/34825826
http://dx.doi.org/10.1302/2633-1462.211.BJO-2021-0147.R1
_version_ 1784608507522187264
author Wight, Christian M.
Whyne, Cari M.
Bogoch, Earl R.
Zdero, Radovan
Chapman, Ryan M.
van Citters, Douglas W.
Walsh, William R.
Schemitsch, Emil
author_facet Wight, Christian M.
Whyne, Cari M.
Bogoch, Earl R.
Zdero, Radovan
Chapman, Ryan M.
van Citters, Douglas W.
Walsh, William R.
Schemitsch, Emil
author_sort Wight, Christian M.
collection PubMed
description AIMS: This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. METHODS: In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth. RESULTS: Overall, 36 mm heads had lower corrosion onset load (p = 0.009) and change in open circuit potential (OCP) during simulated gait with (p = 0.006) and without joint movement (p = 0.004). Discontinuing gait’s joint movement decreased corrosion currents (p = 0.042); however, wear testing showed no significant effect of joint movement on taper damage. In addition, 36 mm heads had greater corrosion area (p = 0.050), but no significant difference was found for maximum linear wear depth (p = 0.155). CONCLUSION: Larger heads are more susceptible to taper corrosion; however, not due to frictional torque as hypothesized. An alternative hypothesis of taper flexural rigidity differential is proposed. Further studies are necessary to investigate the clinical significance and underlying mechanism of this finding. Cite this article: Bone Jt Open 2021;2(11):1004–1016.
format Online
Article
Text
id pubmed-8636299
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The British Editorial Society of Bone & Joint Surgery
record_format MEDLINE/PubMed
spelling pubmed-86362992021-12-17 Effect of head size and rotation on taper corrosion in a hip simulator Wight, Christian M. Whyne, Cari M. Bogoch, Earl R. Zdero, Radovan Chapman, Ryan M. van Citters, Douglas W. Walsh, William R. Schemitsch, Emil Bone Jt Open Hip AIMS: This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. METHODS: In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth. RESULTS: Overall, 36 mm heads had lower corrosion onset load (p = 0.009) and change in open circuit potential (OCP) during simulated gait with (p = 0.006) and without joint movement (p = 0.004). Discontinuing gait’s joint movement decreased corrosion currents (p = 0.042); however, wear testing showed no significant effect of joint movement on taper damage. In addition, 36 mm heads had greater corrosion area (p = 0.050), but no significant difference was found for maximum linear wear depth (p = 0.155). CONCLUSION: Larger heads are more susceptible to taper corrosion; however, not due to frictional torque as hypothesized. An alternative hypothesis of taper flexural rigidity differential is proposed. Further studies are necessary to investigate the clinical significance and underlying mechanism of this finding. Cite this article: Bone Jt Open 2021;2(11):1004–1016. The British Editorial Society of Bone & Joint Surgery 2021-11-26 /pmc/articles/PMC8636299/ /pubmed/34825826 http://dx.doi.org/10.1302/2633-1462.211.BJO-2021-0147.R1 Text en © 2021 Author(s) et al. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND 4.0) licence, which permits the copying and redistribution of the work only, and provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc-nd/4.0/
spellingShingle Hip
Wight, Christian M.
Whyne, Cari M.
Bogoch, Earl R.
Zdero, Radovan
Chapman, Ryan M.
van Citters, Douglas W.
Walsh, William R.
Schemitsch, Emil
Effect of head size and rotation on taper corrosion in a hip simulator
title Effect of head size and rotation on taper corrosion in a hip simulator
title_full Effect of head size and rotation on taper corrosion in a hip simulator
title_fullStr Effect of head size and rotation on taper corrosion in a hip simulator
title_full_unstemmed Effect of head size and rotation on taper corrosion in a hip simulator
title_short Effect of head size and rotation on taper corrosion in a hip simulator
title_sort effect of head size and rotation on taper corrosion in a hip simulator
topic Hip
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8636299/
https://www.ncbi.nlm.nih.gov/pubmed/34825826
http://dx.doi.org/10.1302/2633-1462.211.BJO-2021-0147.R1
work_keys_str_mv AT wightchristianm effectofheadsizeandrotationontapercorrosioninahipsimulator
AT whynecarim effectofheadsizeandrotationontapercorrosioninahipsimulator
AT bogochearlr effectofheadsizeandrotationontapercorrosioninahipsimulator
AT zderoradovan effectofheadsizeandrotationontapercorrosioninahipsimulator
AT chapmanryanm effectofheadsizeandrotationontapercorrosioninahipsimulator
AT vancittersdouglasw effectofheadsizeandrotationontapercorrosioninahipsimulator
AT walshwilliamr effectofheadsizeandrotationontapercorrosioninahipsimulator
AT schemitschemil effectofheadsizeandrotationontapercorrosioninahipsimulator