Cargando…
RELIABILITY AND VALIDITY OF A STANDARDIZED ULTRASOUND EXAMINATION PROTOCOL TO QUANTIFY VASTUS LATERALIS MUSCLE
OBJECTIVE: To evaluate the reliability and validity of a standardized ultrasound examination protocol for measuring vastus lateralis muscle size. DESIGN: Prospective cohort study. SUBJECTS: Sixteen staff members of the university hospital of Heidelberg. METHODS: Muscle thickness, cross-sectional are...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Foundation for Rehabilitation Information
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8638746/ https://www.ncbi.nlm.nih.gov/pubmed/34121129 http://dx.doi.org/10.2340/16501977-2854 |
Sumario: | OBJECTIVE: To evaluate the reliability and validity of a standardized ultrasound examination protocol for measuring vastus lateralis muscle size. DESIGN: Prospective cohort study. SUBJECTS: Sixteen staff members of the university hospital of Heidelberg. METHODS: Muscle thickness, cross-sectional area and subcutaneous adipose tissue thickness were measurxsed at 3 standardized sites on the right and left vastus lateralis muscle. Ultrasound measurements were collected by 2 independent investigators on 2 different days and compared with magnetic resonance imaging measurements. RESULTS: Intraclass correlation coefficients (ICC) for intra- and inter-rater reliability showed very good closeness of agreement for all parameters (ICC = 0.929–0.994, p < 0.001). Muscle thickness and subcutaneous adipose tissue thickness ultrasound and magnetic resonance imaging measurements revealed good to very good closeness of agreement (ICC = 0.835–0.969, p < 0.001), whereas cross-sectional area showed only average closeness of agreement (ICC = 0.727, p < 0.001). A strong predictive positive correlation for ultrasound and magnetic resonance imaging-based measurements of cross-sectional area was found (R(²) = 0.793, p < 0.001). CONCLUSION: By standardization of an examination protocol, quantitative vastus lateralis muscle ultrasound proved to be a reliable method for assessing vastus lateralis muscle size. Furthermore, this protocol is valid for measuring muscle thickness and subcutaneous adipose tissue thickness, although there seems to be a systematic underestimation of cross-sectional area depending on subcutaneous adipose tissue thickness. |
---|