1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood
BACKGROUND: Therapeutic drug monitoring (TDM) is paramount to optimize the safety and efficacy of vancomycin (VAN). In children, TDM is challenged by difficulty in obtaining venous samples, impeding timely sampling. We assessed the ability of volumetric absorptive microsampling (VAMS) as a novel, wh...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8643864/ http://dx.doi.org/10.1093/ofid/ofab466.1298 |
_version_ | 1784609952226082816 |
---|---|
author | Downes, Kevin J Tam, Derrick Sharova, Anna Vedar, Christina Fitzgerald, Julie C Jawad, Abbas F Moorthy, Ganesh S Zuppa, Athena F |
author_facet | Downes, Kevin J Tam, Derrick Sharova, Anna Vedar, Christina Fitzgerald, Julie C Jawad, Abbas F Moorthy, Ganesh S Zuppa, Athena F |
author_sort | Downes, Kevin J |
collection | PubMed |
description | BACKGROUND: Therapeutic drug monitoring (TDM) is paramount to optimize the safety and efficacy of vancomycin (VAN). In children, TDM is challenged by difficulty in obtaining venous samples, impeding timely sampling. We assessed the ability of volumetric absorptive microsampling (VAMS) as a novel, whole blood sampling technique to predict plasma VAN concentrations in plasma. METHODS: We conducted a prospective pilot study among critically ill children prescribed VAN for clinical care. Coincident with VAN TDM in plasma (P), we collected 20 µL of capillary whole blood (C) and venous/arterial whole blood (V) using VAMS. Paired VAMS-P samples drawn >5 mins apart and VAMS samples with over- or under-loaded filter tip on visual inspection were excluded. Plasma concentrations were measured via chemiluminescent immunoassay in the Chemistry Laboratory. VAMS C and V concentrations were measured using LC/MS in the Bioanalytic Core Laboratory. Plasma concentrations were predicted from whole blood VAMS with Passing-Bablok regression using 3 methods: 1) uncorrected VAMS measures, 2) hematocrit-corrected VAMS, and 3) lab-corrected VAMS (Figure 1). We then assessed bias, imprecision, and accuracy of plasma predictions from VAMS (C and V) as compared to coincident P concentrations for each technique (Figure 1). Figure 1. Methods for relating whole blood vancomycin concentrations collected via VAMS to plasma concentrations and measure to evaluate predictive performance. [Image: see text] RESULTS: Paired samples were collected from 31 enrolled subjects (Figure 2), with a median age of 3.3 years (range 0.1-17.9). Measured P concentrations ranged from 4.6 - 54.9 mg/L. 11 C samples (29%) and 3 V samples (10%) were excluded due to collection issues. Prediction results are shown in Figure 3. The 3 prediction techniques had similar performance characteristics, with each method displaying minimal bias (-0.4-2.0%) and reasonable imprecision (13.7-20.2%). The accuracy of prediction of P concentrations using VAMS was better for V than C samples. Figure 2. Flow diagram from sample collection to evaluation. [Image: see text] Abbreviations: C-P, capillary VAMS-plasma; V-P, venous/arterial VAMS-plasma; VAMS, volumetric absorptive microsampling. Figure 3. Performance of 3 techniques to predict plasma vancomycin concentrations using whole blood collected via VAMS. [Image: see text] CONCLUSION: Our pilot highlights the challenges of using VAMS for TDM. Sample collection issues were common. When VAMS is used, education on collection techniques is imperative. The predictive performance of VAMS was modest and V sampling had higher accuracy than C, although our sample size was small. Larger studies will be needed to further evaluate the predictive performance of the regression equations derived by our study. DISCLOSURES: Kevin J. Downes, MD, Merck, Inc. (Grant/Research Support) |
format | Online Article Text |
id | pubmed-8643864 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-86438642021-12-06 1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood Downes, Kevin J Tam, Derrick Sharova, Anna Vedar, Christina Fitzgerald, Julie C Jawad, Abbas F Moorthy, Ganesh S Zuppa, Athena F Open Forum Infect Dis Poster Abstracts BACKGROUND: Therapeutic drug monitoring (TDM) is paramount to optimize the safety and efficacy of vancomycin (VAN). In children, TDM is challenged by difficulty in obtaining venous samples, impeding timely sampling. We assessed the ability of volumetric absorptive microsampling (VAMS) as a novel, whole blood sampling technique to predict plasma VAN concentrations in plasma. METHODS: We conducted a prospective pilot study among critically ill children prescribed VAN for clinical care. Coincident with VAN TDM in plasma (P), we collected 20 µL of capillary whole blood (C) and venous/arterial whole blood (V) using VAMS. Paired VAMS-P samples drawn >5 mins apart and VAMS samples with over- or under-loaded filter tip on visual inspection were excluded. Plasma concentrations were measured via chemiluminescent immunoassay in the Chemistry Laboratory. VAMS C and V concentrations were measured using LC/MS in the Bioanalytic Core Laboratory. Plasma concentrations were predicted from whole blood VAMS with Passing-Bablok regression using 3 methods: 1) uncorrected VAMS measures, 2) hematocrit-corrected VAMS, and 3) lab-corrected VAMS (Figure 1). We then assessed bias, imprecision, and accuracy of plasma predictions from VAMS (C and V) as compared to coincident P concentrations for each technique (Figure 1). Figure 1. Methods for relating whole blood vancomycin concentrations collected via VAMS to plasma concentrations and measure to evaluate predictive performance. [Image: see text] RESULTS: Paired samples were collected from 31 enrolled subjects (Figure 2), with a median age of 3.3 years (range 0.1-17.9). Measured P concentrations ranged from 4.6 - 54.9 mg/L. 11 C samples (29%) and 3 V samples (10%) were excluded due to collection issues. Prediction results are shown in Figure 3. The 3 prediction techniques had similar performance characteristics, with each method displaying minimal bias (-0.4-2.0%) and reasonable imprecision (13.7-20.2%). The accuracy of prediction of P concentrations using VAMS was better for V than C samples. Figure 2. Flow diagram from sample collection to evaluation. [Image: see text] Abbreviations: C-P, capillary VAMS-plasma; V-P, venous/arterial VAMS-plasma; VAMS, volumetric absorptive microsampling. Figure 3. Performance of 3 techniques to predict plasma vancomycin concentrations using whole blood collected via VAMS. [Image: see text] CONCLUSION: Our pilot highlights the challenges of using VAMS for TDM. Sample collection issues were common. When VAMS is used, education on collection techniques is imperative. The predictive performance of VAMS was modest and V sampling had higher accuracy than C, although our sample size was small. Larger studies will be needed to further evaluate the predictive performance of the regression equations derived by our study. DISCLOSURES: Kevin J. Downes, MD, Merck, Inc. (Grant/Research Support) Oxford University Press 2021-12-04 /pmc/articles/PMC8643864/ http://dx.doi.org/10.1093/ofid/ofab466.1298 Text en © The Author(s) 2021. Published by Oxford University Press on behalf of Infectious Diseases Society of America. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Poster Abstracts Downes, Kevin J Tam, Derrick Sharova, Anna Vedar, Christina Fitzgerald, Julie C Jawad, Abbas F Moorthy, Ganesh S Zuppa, Athena F 1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood |
title | 1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood |
title_full | 1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood |
title_fullStr | 1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood |
title_full_unstemmed | 1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood |
title_short | 1104. Comparison of Antibiotic Sampling Techniques: Predicting Plasma Vancomycin Concentrations Using Volumetric Absorptive Microsampling (VAMS) from Capillary and Venous/Arterial Whole Blood |
title_sort | 1104. comparison of antibiotic sampling techniques: predicting plasma vancomycin concentrations using volumetric absorptive microsampling (vams) from capillary and venous/arterial whole blood |
topic | Poster Abstracts |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8643864/ http://dx.doi.org/10.1093/ofid/ofab466.1298 |
work_keys_str_mv | AT downeskevinj 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood AT tamderrick 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood AT sharovaanna 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood AT vedarchristina 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood AT fitzgeraldjuliec 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood AT jawadabbasf 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood AT moorthyganeshs 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood AT zuppaathenaf 1104comparisonofantibioticsamplingtechniquespredictingplasmavancomycinconcentrationsusingvolumetricabsorptivemicrosamplingvamsfromcapillaryandvenousarterialwholeblood |