Cargando…

The use of imageless navigation to quantify cutting error in total knee arthroplasty

PURPOSE: Navigated total knee arthroplasty (TKA) improves implant alignment by providing feedback on resection parameters based on femoral and tibial cutting guide positions. However, saw blade thickness, deflection, and cutting guide motion may lead to final bone cuts differing from planned resecti...

Descripción completa

Detalles Bibliográficos
Autores principales: Schwarzkopf, Ran, Meftah, Morteza, Marwin, Scott E., Zabat, Michelle A., Muir, Jeffrey M., Lamb, Iain R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8645113/
https://www.ncbi.nlm.nih.gov/pubmed/34863317
http://dx.doi.org/10.1186/s43019-021-00125-z
_version_ 1784610241882619904
author Schwarzkopf, Ran
Meftah, Morteza
Marwin, Scott E.
Zabat, Michelle A.
Muir, Jeffrey M.
Lamb, Iain R.
author_facet Schwarzkopf, Ran
Meftah, Morteza
Marwin, Scott E.
Zabat, Michelle A.
Muir, Jeffrey M.
Lamb, Iain R.
author_sort Schwarzkopf, Ran
collection PubMed
description PURPOSE: Navigated total knee arthroplasty (TKA) improves implant alignment by providing feedback on resection parameters based on femoral and tibial cutting guide positions. However, saw blade thickness, deflection, and cutting guide motion may lead to final bone cuts differing from planned resections, potentially contributing to suboptimal component alignment. We used an imageless navigation device to intraoperatively quantify the magnitude of error between planned and actual resections, hypothesizing final bone cuts will differ from planned alignment. MATERIALS AND METHODS: A retrospective study including 60 consecutive patients undergoing primary TKA using a novel imageless navigation device was conducted. Device measurements of resection parameters were obtained via attachment of optical trackers to femoral and tibial cutting guides prior to resection. Following resection, optical trackers were placed directly on the bone cut surface and measurements were recorded. Cutting guide and bone resection measurements of both femoral and tibial varus/valgus, femoral flexion, tibial slope angles, and both femoral and tibial medial and lateral resection depths were compared using a Student's t-test. RESULTS: Femoral cutting guide position differed from the actual cut by an average 0.6 ± 0.5° (p = 0.85) in the varus/valgus angle and 1.0 ± 1.0° (p = 0.003) in the flexion/extension angle. The difference between planned and actual cut measurements for medial and lateral femoral resection depth was 1.1 ± 1.1 mm (p = 0.32) and 1.2 ± 1.0 mm (p = 0.067), respectively. Planned cut measurements based on tibial guide position differed from the actual cut by an average of 0.9 ± 0.8° (p = 0.63) in the varus/valgus angle and 1.1 ± 1.0° (p = 0.95) in slope angle. Measurement of medial and lateral tibial resection depth differed by an average of 0.1 ± 1.8 mm (p = 0.78) and 0.2 ± 2.1 mm (p = 0.85), respectively. CONCLUSIONS: Significant discrepancies between planned and actual femoral bone resection were demonstrated for flexion/extension angle, likely the result of cutting error. Our data highlights the importance of cut verification postresection to confirm planned resections are achieved, and suggests imageless navigation may be a source of feedback that would allow surgeons to intraoperatively adjust resections to achieve optimal implant alignment.
format Online
Article
Text
id pubmed-8645113
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-86451132021-12-13 The use of imageless navigation to quantify cutting error in total knee arthroplasty Schwarzkopf, Ran Meftah, Morteza Marwin, Scott E. Zabat, Michelle A. Muir, Jeffrey M. Lamb, Iain R. Knee Surg Relat Res Research Article PURPOSE: Navigated total knee arthroplasty (TKA) improves implant alignment by providing feedback on resection parameters based on femoral and tibial cutting guide positions. However, saw blade thickness, deflection, and cutting guide motion may lead to final bone cuts differing from planned resections, potentially contributing to suboptimal component alignment. We used an imageless navigation device to intraoperatively quantify the magnitude of error between planned and actual resections, hypothesizing final bone cuts will differ from planned alignment. MATERIALS AND METHODS: A retrospective study including 60 consecutive patients undergoing primary TKA using a novel imageless navigation device was conducted. Device measurements of resection parameters were obtained via attachment of optical trackers to femoral and tibial cutting guides prior to resection. Following resection, optical trackers were placed directly on the bone cut surface and measurements were recorded. Cutting guide and bone resection measurements of both femoral and tibial varus/valgus, femoral flexion, tibial slope angles, and both femoral and tibial medial and lateral resection depths were compared using a Student's t-test. RESULTS: Femoral cutting guide position differed from the actual cut by an average 0.6 ± 0.5° (p = 0.85) in the varus/valgus angle and 1.0 ± 1.0° (p = 0.003) in the flexion/extension angle. The difference between planned and actual cut measurements for medial and lateral femoral resection depth was 1.1 ± 1.1 mm (p = 0.32) and 1.2 ± 1.0 mm (p = 0.067), respectively. Planned cut measurements based on tibial guide position differed from the actual cut by an average of 0.9 ± 0.8° (p = 0.63) in the varus/valgus angle and 1.1 ± 1.0° (p = 0.95) in slope angle. Measurement of medial and lateral tibial resection depth differed by an average of 0.1 ± 1.8 mm (p = 0.78) and 0.2 ± 2.1 mm (p = 0.85), respectively. CONCLUSIONS: Significant discrepancies between planned and actual femoral bone resection were demonstrated for flexion/extension angle, likely the result of cutting error. Our data highlights the importance of cut verification postresection to confirm planned resections are achieved, and suggests imageless navigation may be a source of feedback that would allow surgeons to intraoperatively adjust resections to achieve optimal implant alignment. BioMed Central 2021-12-04 /pmc/articles/PMC8645113/ /pubmed/34863317 http://dx.doi.org/10.1186/s43019-021-00125-z Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research Article
Schwarzkopf, Ran
Meftah, Morteza
Marwin, Scott E.
Zabat, Michelle A.
Muir, Jeffrey M.
Lamb, Iain R.
The use of imageless navigation to quantify cutting error in total knee arthroplasty
title The use of imageless navigation to quantify cutting error in total knee arthroplasty
title_full The use of imageless navigation to quantify cutting error in total knee arthroplasty
title_fullStr The use of imageless navigation to quantify cutting error in total knee arthroplasty
title_full_unstemmed The use of imageless navigation to quantify cutting error in total knee arthroplasty
title_short The use of imageless navigation to quantify cutting error in total knee arthroplasty
title_sort use of imageless navigation to quantify cutting error in total knee arthroplasty
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8645113/
https://www.ncbi.nlm.nih.gov/pubmed/34863317
http://dx.doi.org/10.1186/s43019-021-00125-z
work_keys_str_mv AT schwarzkopfran theuseofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT meftahmorteza theuseofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT marwinscotte theuseofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT zabatmichellea theuseofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT muirjeffreym theuseofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT lambiainr theuseofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT schwarzkopfran useofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT meftahmorteza useofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT marwinscotte useofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT zabatmichellea useofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT muirjeffreym useofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty
AT lambiainr useofimagelessnavigationtoquantifycuttingerrorintotalkneearthroplasty