Cargando…
Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy
INTRODUCTION: Paraspinal stereotactic body radiotherapy (SBRT) involves risks of severe complications. We evaluated the safety of the paraspinal SBRT program in a large academic hospital by applying failure modes and effects analysis. METHODS: The analysis was conducted by a multidisciplinary commit...
Autores principales: | , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8664134/ https://www.ncbi.nlm.nih.gov/pubmed/34708910 http://dx.doi.org/10.1002/acm2.13455 |
_version_ | 1784613782200254464 |
---|---|
author | Lee, Sangkyu Lovelock, Dale Michael Kowalski, Alex Chapman, Kate Foley, Robert Gil, Mary Pastrana, Gerri Higginson, Daniel S. Yamada, Yoshiya Zhang, Lei Mechalakos, James Yorke, Ellen |
author_facet | Lee, Sangkyu Lovelock, Dale Michael Kowalski, Alex Chapman, Kate Foley, Robert Gil, Mary Pastrana, Gerri Higginson, Daniel S. Yamada, Yoshiya Zhang, Lei Mechalakos, James Yorke, Ellen |
author_sort | Lee, Sangkyu |
collection | PubMed |
description | INTRODUCTION: Paraspinal stereotactic body radiotherapy (SBRT) involves risks of severe complications. We evaluated the safety of the paraspinal SBRT program in a large academic hospital by applying failure modes and effects analysis. METHODS: The analysis was conducted by a multidisciplinary committee (two therapists, one dosimetrist, four physicists, and two radiation oncologists). The paraspinal SBRT workflow was segmented into four phases (simulation, treatment planning, delivery, and machine quality assurance (QA)). Each phase was further divided into a sequence of sub‐processes. Potential failure modes (PFM) were identified from each subprocess and scored in terms of the frequency of occurrence, severity and detectability, and a risk priority number (RPN). High‐risk PFMs were identified based on RPN and were studied for root causes using fault tree analysis. RESULTS: Our paraspinal SBRT process was characterized by eight simulations, 11 treatment planning, nine delivery, and two machine QA sub‐processes. There were 18, 29, 19, and eight PFMs identified from simulation, planning, treatment, and machine QA, respectively. The median RPN of the PFMs was 62.9 for simulation, 68.3 for planning, 52.9 for delivery, and 22.0 for machine QA. The three PFMs with the highest RPN were: previous radiotherapy outside the institution is not accurately evaluated (RPN: 293.3), incorrect registration between diagnostic magnetic resonance imaging and simulation computed tomography causing incorrect contours (273.0), and undetected patient movement before ExacTrac baseline (217.8). Remedies to the high RPN failures were implemented, including staff education, standardized magnetic resonance imaging acquisition parameters, and an image fusion process, and additional QA on beam steering. CONCLUSIONS: A paraspinal SBRT workflow in a large clinic was evaluated using a multidisciplinary and systematic risk analysis, which led to feasible solutions to key root causes. Treatment planning was a major source of PFMs that systematically affect the safety and quality of treatments. Accurate evaluation of external treatment records remains a challenge. |
format | Online Article Text |
id | pubmed-8664134 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-86641342021-12-21 Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy Lee, Sangkyu Lovelock, Dale Michael Kowalski, Alex Chapman, Kate Foley, Robert Gil, Mary Pastrana, Gerri Higginson, Daniel S. Yamada, Yoshiya Zhang, Lei Mechalakos, James Yorke, Ellen J Appl Clin Med Phys Radiation Oncology Physics INTRODUCTION: Paraspinal stereotactic body radiotherapy (SBRT) involves risks of severe complications. We evaluated the safety of the paraspinal SBRT program in a large academic hospital by applying failure modes and effects analysis. METHODS: The analysis was conducted by a multidisciplinary committee (two therapists, one dosimetrist, four physicists, and two radiation oncologists). The paraspinal SBRT workflow was segmented into four phases (simulation, treatment planning, delivery, and machine quality assurance (QA)). Each phase was further divided into a sequence of sub‐processes. Potential failure modes (PFM) were identified from each subprocess and scored in terms of the frequency of occurrence, severity and detectability, and a risk priority number (RPN). High‐risk PFMs were identified based on RPN and were studied for root causes using fault tree analysis. RESULTS: Our paraspinal SBRT process was characterized by eight simulations, 11 treatment planning, nine delivery, and two machine QA sub‐processes. There were 18, 29, 19, and eight PFMs identified from simulation, planning, treatment, and machine QA, respectively. The median RPN of the PFMs was 62.9 for simulation, 68.3 for planning, 52.9 for delivery, and 22.0 for machine QA. The three PFMs with the highest RPN were: previous radiotherapy outside the institution is not accurately evaluated (RPN: 293.3), incorrect registration between diagnostic magnetic resonance imaging and simulation computed tomography causing incorrect contours (273.0), and undetected patient movement before ExacTrac baseline (217.8). Remedies to the high RPN failures were implemented, including staff education, standardized magnetic resonance imaging acquisition parameters, and an image fusion process, and additional QA on beam steering. CONCLUSIONS: A paraspinal SBRT workflow in a large clinic was evaluated using a multidisciplinary and systematic risk analysis, which led to feasible solutions to key root causes. Treatment planning was a major source of PFMs that systematically affect the safety and quality of treatments. Accurate evaluation of external treatment records remains a challenge. John Wiley and Sons Inc. 2021-10-28 /pmc/articles/PMC8664134/ /pubmed/34708910 http://dx.doi.org/10.1002/acm2.13455 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Lee, Sangkyu Lovelock, Dale Michael Kowalski, Alex Chapman, Kate Foley, Robert Gil, Mary Pastrana, Gerri Higginson, Daniel S. Yamada, Yoshiya Zhang, Lei Mechalakos, James Yorke, Ellen Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy |
title | Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy |
title_full | Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy |
title_fullStr | Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy |
title_full_unstemmed | Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy |
title_short | Failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy |
title_sort | failure mode and effect analysis for linear accelerator‐based paraspinal stereotactic body radiotherapy |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8664134/ https://www.ncbi.nlm.nih.gov/pubmed/34708910 http://dx.doi.org/10.1002/acm2.13455 |
work_keys_str_mv | AT leesangkyu failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT lovelockdalemichael failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT kowalskialex failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT chapmankate failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT foleyrobert failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT gilmary failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT pastranagerri failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT higginsondaniels failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT yamadayoshiya failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT zhanglei failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT mechalakosjames failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy AT yorkeellen failuremodeandeffectanalysisforlinearacceleratorbasedparaspinalstereotacticbodyradiotherapy |