Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
BACKGROUND: Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). CASE SUMMARY: A 60-year-old male with symptomatic, severe ischa...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8665680/ https://www.ncbi.nlm.nih.gov/pubmed/34909577 http://dx.doi.org/10.1093/ehjcr/ytab491 |
_version_ | 1784614059398660096 |
---|---|
author | Uppal, Abhimanyu Kathuria, Sanjeev Shah, Bhushan Trehan, Vijay |
author_facet | Uppal, Abhimanyu Kathuria, Sanjeev Shah, Bhushan Trehan, Vijay |
author_sort | Uppal, Abhimanyu |
collection | PubMed |
description | BACKGROUND: Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). CASE SUMMARY: A 60-year-old male with symptomatic, severe ischaemic cardiomyopathy, and atrial fibrillation following a prior anterior wall myocardial infarction received a dual-chamber ICD with 7F-RiataST ventricular lead as a primary prevention strategy against sudden cardiac death in 2008. In 2017, a pulse generator replacement was performed for elective replacement indicator status. At that time, CE was noted in the ventricular lead but the electrical lead parameters were normal, hence lead replacement was decided against and the patient was closely followed up thereafter. Four years later, the patient presented with multiple ICD shocks within 48 h. Implantable cardioverter-defibrillator interrogation showed noise on the ventricular electrogram (EGM) channel that was detected as ventricular fibrillation (VF) episodes, triggering inappropriate ICD therapy (five ICD detected VF events within 24 h triggering three antitachycardia pacing therapies and one shock). Lead impedance and R-wave amplitude were within normal range in supine position but dramatically worsened in sitting posture. A new ventricular lead was implanted and the old lead abandoned. The patient has not experienced any device therapy in the follow-up period. DISCUSSION: An electrically inert CE of Riata ICD leads needs close follow-up because an ELF may occur even after several years. A careful analysis of EGMs including postural changes in lead parameters can aid in detection and better characterization of underlying electrical dysfunction following CE. |
format | Online Article Text |
id | pubmed-8665680 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-86656802021-12-13 Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report Uppal, Abhimanyu Kathuria, Sanjeev Shah, Bhushan Trehan, Vijay Eur Heart J Case Rep Case Report BACKGROUND: Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). CASE SUMMARY: A 60-year-old male with symptomatic, severe ischaemic cardiomyopathy, and atrial fibrillation following a prior anterior wall myocardial infarction received a dual-chamber ICD with 7F-RiataST ventricular lead as a primary prevention strategy against sudden cardiac death in 2008. In 2017, a pulse generator replacement was performed for elective replacement indicator status. At that time, CE was noted in the ventricular lead but the electrical lead parameters were normal, hence lead replacement was decided against and the patient was closely followed up thereafter. Four years later, the patient presented with multiple ICD shocks within 48 h. Implantable cardioverter-defibrillator interrogation showed noise on the ventricular electrogram (EGM) channel that was detected as ventricular fibrillation (VF) episodes, triggering inappropriate ICD therapy (five ICD detected VF events within 24 h triggering three antitachycardia pacing therapies and one shock). Lead impedance and R-wave amplitude were within normal range in supine position but dramatically worsened in sitting posture. A new ventricular lead was implanted and the old lead abandoned. The patient has not experienced any device therapy in the follow-up period. DISCUSSION: An electrically inert CE of Riata ICD leads needs close follow-up because an ELF may occur even after several years. A careful analysis of EGMs including postural changes in lead parameters can aid in detection and better characterization of underlying electrical dysfunction following CE. Oxford University Press 2021-12-07 /pmc/articles/PMC8665680/ /pubmed/34909577 http://dx.doi.org/10.1093/ehjcr/ytab491 Text en © The Author(s) 2021. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com |
spellingShingle | Case Report Uppal, Abhimanyu Kathuria, Sanjeev Shah, Bhushan Trehan, Vijay Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report |
title | Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report |
title_full | Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report |
title_fullStr | Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report |
title_full_unstemmed | Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report |
title_short | Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report |
title_sort | riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8665680/ https://www.ncbi.nlm.nih.gov/pubmed/34909577 http://dx.doi.org/10.1093/ehjcr/ytab491 |
work_keys_str_mv | AT uppalabhimanyu riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport AT kathuriasanjeev riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport AT shahbhushan riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport AT trehanvijay riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport |