Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report

BACKGROUND: Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). CASE SUMMARY: A 60-year-old male with symptomatic, severe ischa...

Descripción completa

Detalles Bibliográficos
Autores principales: Uppal, Abhimanyu, Kathuria, Sanjeev, Shah, Bhushan, Trehan, Vijay
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8665680/
https://www.ncbi.nlm.nih.gov/pubmed/34909577
http://dx.doi.org/10.1093/ehjcr/ytab491
_version_ 1784614059398660096
author Uppal, Abhimanyu
Kathuria, Sanjeev
Shah, Bhushan
Trehan, Vijay
author_facet Uppal, Abhimanyu
Kathuria, Sanjeev
Shah, Bhushan
Trehan, Vijay
author_sort Uppal, Abhimanyu
collection PubMed
description BACKGROUND: Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). CASE SUMMARY: A 60-year-old male with symptomatic, severe ischaemic cardiomyopathy, and atrial fibrillation following a prior anterior wall myocardial infarction received a dual-chamber ICD with 7F-RiataST ventricular lead as a primary prevention strategy against sudden cardiac death in 2008. In 2017, a pulse generator replacement was performed for elective replacement indicator status. At that time, CE was noted in the ventricular lead but the electrical lead parameters were normal, hence lead replacement was decided against and the patient was closely followed up thereafter. Four years later, the patient presented with multiple ICD shocks within 48 h. Implantable cardioverter-defibrillator interrogation showed noise on the ventricular electrogram (EGM) channel that was detected as ventricular fibrillation (VF) episodes, triggering inappropriate ICD therapy (five ICD detected VF events within 24 h triggering three antitachycardia pacing therapies and one shock). Lead impedance and R-wave amplitude were within normal range in supine position but dramatically worsened in sitting posture. A new ventricular lead was implanted and the old lead abandoned. The patient has not experienced any device therapy in the follow-up period. DISCUSSION: An electrically inert CE of Riata ICD leads needs close follow-up because an ELF may occur even after several years. A careful analysis of EGMs including postural changes in lead parameters can aid in detection and better characterization of underlying electrical dysfunction following CE.
format Online
Article
Text
id pubmed-8665680
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-86656802021-12-13 Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report Uppal, Abhimanyu Kathuria, Sanjeev Shah, Bhushan Trehan, Vijay Eur Heart J Case Rep Case Report BACKGROUND: Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). CASE SUMMARY: A 60-year-old male with symptomatic, severe ischaemic cardiomyopathy, and atrial fibrillation following a prior anterior wall myocardial infarction received a dual-chamber ICD with 7F-RiataST ventricular lead as a primary prevention strategy against sudden cardiac death in 2008. In 2017, a pulse generator replacement was performed for elective replacement indicator status. At that time, CE was noted in the ventricular lead but the electrical lead parameters were normal, hence lead replacement was decided against and the patient was closely followed up thereafter. Four years later, the patient presented with multiple ICD shocks within 48 h. Implantable cardioverter-defibrillator interrogation showed noise on the ventricular electrogram (EGM) channel that was detected as ventricular fibrillation (VF) episodes, triggering inappropriate ICD therapy (five ICD detected VF events within 24 h triggering three antitachycardia pacing therapies and one shock). Lead impedance and R-wave amplitude were within normal range in supine position but dramatically worsened in sitting posture. A new ventricular lead was implanted and the old lead abandoned. The patient has not experienced any device therapy in the follow-up period. DISCUSSION: An electrically inert CE of Riata ICD leads needs close follow-up because an ELF may occur even after several years. A careful analysis of EGMs including postural changes in lead parameters can aid in detection and better characterization of underlying electrical dysfunction following CE. Oxford University Press 2021-12-07 /pmc/articles/PMC8665680/ /pubmed/34909577 http://dx.doi.org/10.1093/ehjcr/ytab491 Text en © The Author(s) 2021. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Case Report
Uppal, Abhimanyu
Kathuria, Sanjeev
Shah, Bhushan
Trehan, Vijay
Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
title Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
title_full Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
title_fullStr Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
title_full_unstemmed Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
title_short Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
title_sort riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8665680/
https://www.ncbi.nlm.nih.gov/pubmed/34909577
http://dx.doi.org/10.1093/ehjcr/ytab491
work_keys_str_mv AT uppalabhimanyu riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport
AT kathuriasanjeev riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport
AT shahbhushan riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport
AT trehanvijay riataleadfailuremanifestsaselectricalstormwithleadparametersshowinguniqueposturalvariationsacasereport