Cargando…
Intraoperative and Postoperative Intraocular Lens Opacifications: Analysis of 42545 Cases
PURPOSE: To assess the types and causes of intraocular lens (IOL) turbidity in a tertiary eye center. Setting. Qingdao Eye Hospital of Shandong First Medical University, Qingdao, China. DESIGN: Retrospective case series. METHODS: Patients who underwent uncomplicated phacoemulsification and IOL impla...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8668301/ https://www.ncbi.nlm.nih.gov/pubmed/34912573 http://dx.doi.org/10.1155/2021/1285947 |
Sumario: | PURPOSE: To assess the types and causes of intraocular lens (IOL) turbidity in a tertiary eye center. Setting. Qingdao Eye Hospital of Shandong First Medical University, Qingdao, China. DESIGN: Retrospective case series. METHODS: Patients who underwent uncomplicated phacoemulsification and IOL implantation for cataract between January 2015 and December 2019 were included. Medical records were reviewed of participants with intraoperative or postoperative IOL opacification for clinical data, artificial crystal materials, and causes of the opacification. RESULTS: A total of 42545 IOLs were implanted in the five years, comprising 25471 (66.0%) hydrophilic IOLs, 11881 (27.9%) hydrophobic IOLs, and 2601 (6.1%) hydrophilic-hydrophobic acrylic IOLs. Among the operated eyes, 14 eyes (13 patients) experienced IOL opacification, which was permanent for 10 IOLs, including 7 (0.6%) hydrophilic IOLs (860UV) and 3 (0.2%) hydrophilic-hydrophobic acrylic IOLs (L-312). The mean interval between surgery and diagnosis of permanent opacification was 34.4 ± 18.4 (SD) months (range, 12 to 59 months). Permanent IOL clouding led to a statistically significant reduction in best corrected visual acuity (mean, 0.64 ± 0.4 logMAR; P < 0.004). Acute IOL clouding occurred in four eyes during the implantation of a hydrophilic-hydrophobic acrylic IOL of L-312, 809M, or 839M and returned to transparency several hours later. All four procedures were performed in winter, with the mean outside temperature being −5.75°C. CONCLUSIONS: The rate of IOL opacification was 0.03%. Both delayed postoperative and acute intraoperative opacifications occurred with various characteristics in IOLs made of different materials and designs. Clinicians should be aware of this risk for cataract surgery. |
---|