A Study of Analytical and Clinical Sensitivity of Aptima SARS-CoV-2 Assay (Hologic) and Proposals of Complementary Tests for SARS-CoV-2 Detection in Low Viral Load Specimens
Early and accurate detection of SARS-CoV-2 is important for diagnosis and transmission control. The use of high-throughput and automated testing allows laboratories to better deliver diagnostic testing given manpower and resource limitations. We validated the clinical and analytical performance of t...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer US
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8669225/ https://www.ncbi.nlm.nih.gov/pubmed/34905087 http://dx.doi.org/10.1007/s00284-021-02730-3 |
Sumario: | Early and accurate detection of SARS-CoV-2 is important for diagnosis and transmission control. The use of high-throughput and automated testing allows laboratories to better deliver diagnostic testing given manpower and resource limitations. We validated the clinical and analytical performance of the Hologic Panther Aptima SARS-CoV-2 assay with an emphasis on detection of specimens with low viral loads. The clinical performance was evaluated using 245 clinical specimens, against a comparator PCR-based laboratory developed test (LDT). The analytical performance was determined by replicate testing of contrived samples in a ten-fold dilution series (C(T) values 32–42, based on LDT). The Aptima assay had 96.7% overall percent agreement, 100% negative percent agreement and 88.1% positive percent agreement. It was able to consistently detect SARS-CoV-2 in contrived samples with C(T) = 32 by LDT (calculated 2354 copies/mL). The 95% limit of detection of the Aptima assay was estimated to be at LDT C(T) = 33 (equivalent to 870 copies/mL). The relative light units (RLU) × 1000 for 52 true positive clinical specimens was 962.2 ± 181.5, and that for the 186 true negative specimens was 264.6 ± 14.3. The Aptima assay was a reliable method with a high overall percent agreement against our comparator LDT. We propose that samples reported as negative by the Aptima assay with RLU > 350 be tested by a secondary method, in order to improve detection of samples with very low viral loads. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00284-021-02730-3. |
---|