A Study of Analytical and Clinical Sensitivity of Aptima SARS-CoV-2 Assay (Hologic) and Proposals of Complementary Tests for SARS-CoV-2 Detection in Low Viral Load Specimens

Early and accurate detection of SARS-CoV-2 is important for diagnosis and transmission control. The use of high-throughput and automated testing allows laboratories to better deliver diagnostic testing given manpower and resource limitations. We validated the clinical and analytical performance of t...

Descripción completa

Detalles Bibliográficos
Autores principales: La, My-Van, Koo, Seok Hwee, Jiang, Boran, Heng, Ying Xuan, Tan, Thean Yen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer US 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8669225/
https://www.ncbi.nlm.nih.gov/pubmed/34905087
http://dx.doi.org/10.1007/s00284-021-02730-3
Descripción
Sumario:Early and accurate detection of SARS-CoV-2 is important for diagnosis and transmission control. The use of high-throughput and automated testing allows laboratories to better deliver diagnostic testing given manpower and resource limitations. We validated the clinical and analytical performance of the Hologic Panther Aptima SARS-CoV-2 assay with an emphasis on detection of specimens with low viral loads. The clinical performance was evaluated using 245 clinical specimens, against a comparator PCR-based laboratory developed test (LDT). The analytical performance was determined by replicate testing of contrived samples in a ten-fold dilution series (C(T) values 32–42, based on LDT). The Aptima assay had 96.7% overall percent agreement, 100% negative percent agreement and 88.1% positive percent agreement. It was able to consistently detect SARS-CoV-2 in contrived samples with C(T) = 32 by LDT (calculated 2354 copies/mL). The 95% limit of detection of the Aptima assay was estimated to be at LDT C(T) = 33 (equivalent to 870 copies/mL). The relative light units (RLU) × 1000 for 52 true positive clinical specimens was 962.2 ± 181.5, and that for the 186 true negative specimens was 264.6 ± 14.3. The Aptima assay was a reliable method with a high overall percent agreement against our comparator LDT. We propose that samples reported as negative by the Aptima assay with RLU > 350 be tested by a secondary method, in order to improve detection of samples with very low viral loads. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00284-021-02730-3.