Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices

Despite improved hemocompatibility of left-ventricular assist devices (LVADs), assessment of blood damage remains mandatory in preclinical testing as standardized by ASTM-F1841. The most relevant test fluid is fresh, non-pooled human blood, but the limited volume of a standard donation requires sign...

Descripción completa

Detalles Bibliográficos
Autores principales: Woelke, Eva, Mager, Ilona, Schmitz-Rode, Thomas, Steinseifer, Ulrich, Clauser, Johanna C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8671281/
https://www.ncbi.nlm.nih.gov/pubmed/34431015
http://dx.doi.org/10.1007/s10439-021-02849-1
_version_ 1784615115720491008
author Woelke, Eva
Mager, Ilona
Schmitz-Rode, Thomas
Steinseifer, Ulrich
Clauser, Johanna C.
author_facet Woelke, Eva
Mager, Ilona
Schmitz-Rode, Thomas
Steinseifer, Ulrich
Clauser, Johanna C.
author_sort Woelke, Eva
collection PubMed
description Despite improved hemocompatibility of left-ventricular assist devices (LVADs), assessment of blood damage remains mandatory in preclinical testing as standardized by ASTM-F1841. The most relevant test fluid is fresh, non-pooled human blood, but the limited volume of a standard donation requires significantly smaller loops than those commonly used with animal blood. In a recent study with porcine blood, we verified a miniaturized test loop with only 160 mL for the ASTM-conform paired testing of at least two LVADs and a static reference. Here, we validated this mini test loop for standardized assessment of blood damage with one 450-mL single donation of fresh human blood. Blood damage was assessed for HeartMate 3 and BPX-80 in 9 experiments with heparinized human blood for 6 hours. We analyzed plasma free hemoglobin, von Willebrand factor (vWF) concentration and collagen-binding functionality and calculated indices of hemolysis and vWF-ratios. Overall, we observed less blood damage compared to our previous study; however, the differences in mean indices of hemolysis and in mean normalized vWF-ratio between BPX-80 and HeartMate 3 were consistent for human blood. Thus, our mini test loop proved to be valid for preclinical standardized assessment of blood damage with only 450 mL of fresh human blood.
format Online
Article
Text
id pubmed-8671281
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-86712812021-12-28 Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices Woelke, Eva Mager, Ilona Schmitz-Rode, Thomas Steinseifer, Ulrich Clauser, Johanna C. Ann Biomed Eng Original Article Despite improved hemocompatibility of left-ventricular assist devices (LVADs), assessment of blood damage remains mandatory in preclinical testing as standardized by ASTM-F1841. The most relevant test fluid is fresh, non-pooled human blood, but the limited volume of a standard donation requires significantly smaller loops than those commonly used with animal blood. In a recent study with porcine blood, we verified a miniaturized test loop with only 160 mL for the ASTM-conform paired testing of at least two LVADs and a static reference. Here, we validated this mini test loop for standardized assessment of blood damage with one 450-mL single donation of fresh human blood. Blood damage was assessed for HeartMate 3 and BPX-80 in 9 experiments with heparinized human blood for 6 hours. We analyzed plasma free hemoglobin, von Willebrand factor (vWF) concentration and collagen-binding functionality and calculated indices of hemolysis and vWF-ratios. Overall, we observed less blood damage compared to our previous study; however, the differences in mean indices of hemolysis and in mean normalized vWF-ratio between BPX-80 and HeartMate 3 were consistent for human blood. Thus, our mini test loop proved to be valid for preclinical standardized assessment of blood damage with only 450 mL of fresh human blood. Springer International Publishing 2021-08-24 2021 /pmc/articles/PMC8671281/ /pubmed/34431015 http://dx.doi.org/10.1007/s10439-021-02849-1 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Woelke, Eva
Mager, Ilona
Schmitz-Rode, Thomas
Steinseifer, Ulrich
Clauser, Johanna C.
Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
title Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
title_full Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
title_fullStr Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
title_full_unstemmed Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
title_short Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
title_sort validation of a miniaturized test loop for the assessment of human blood damage by continuous-flow left-ventricular assist devices
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8671281/
https://www.ncbi.nlm.nih.gov/pubmed/34431015
http://dx.doi.org/10.1007/s10439-021-02849-1
work_keys_str_mv AT woelkeeva validationofaminiaturizedtestloopfortheassessmentofhumanblooddamagebycontinuousflowleftventricularassistdevices
AT magerilona validationofaminiaturizedtestloopfortheassessmentofhumanblooddamagebycontinuousflowleftventricularassistdevices
AT schmitzrodethomas validationofaminiaturizedtestloopfortheassessmentofhumanblooddamagebycontinuousflowleftventricularassistdevices
AT steinseiferulrich validationofaminiaturizedtestloopfortheassessmentofhumanblooddamagebycontinuousflowleftventricularassistdevices
AT clauserjohannac validationofaminiaturizedtestloopfortheassessmentofhumanblooddamagebycontinuousflowleftventricularassistdevices