Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture

INTRODUCTION: The aims of study were to assess: 1) performance specifications of Atellica 1500, 2) comparability of Atellica 1500 and Iris, 3) the accuracy of both analysers in their ability to detect bacteria. MATERIALS AND METHODS: Carryover, linearity, precision, reproducibility, and limit of bla...

Descripción completa

Detalles Bibliográficos
Autores principales: Nikler, Ana, Čičak, Helena, Bejuk, Danijela, Radišić Biljak, Vanja, Šimundić, Ana-Maria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672386/
https://www.ncbi.nlm.nih.gov/pubmed/34955669
http://dx.doi.org/10.11613/BM.2022.010701
_version_ 1784615345471881216
author Nikler, Ana
Čičak, Helena
Bejuk, Danijela
Radišić Biljak, Vanja
Šimundić, Ana-Maria
author_facet Nikler, Ana
Čičak, Helena
Bejuk, Danijela
Radišić Biljak, Vanja
Šimundić, Ana-Maria
author_sort Nikler, Ana
collection PubMed
description INTRODUCTION: The aims of study were to assess: 1) performance specifications of Atellica 1500, 2) comparability of Atellica 1500 and Iris, 3) the accuracy of both analysers in their ability to detect bacteria. MATERIALS AND METHODS: Carryover, linearity, precision, reproducibility, and limit of blank (LoB) verification were evaluated for erythrocyte and leukocyte counts. ICSH 2014 protocol was used for estimation of carryover, CLSI EP15-A3 for precision, and CLSI EP17 for LoB verification. Comparison for quantitative parameters was evaluated by Bland-Altman plot and Passing-Bablok regression. Qualitative parameters were evaluated by Weighted kappa analysis. Sixty-five urine samples were randomly selected and sent for urine culture which was used as reference method to determine the accuracy of bacteria detection by analysers. RESULTS: Analytical specifications of Atellica 1500 were successfully verified. Total of 393 samples were used for qualitative comparison, while 269 for sediment urinalysis. Bland-Altman analysis showed statistically significant proportional bias for erythrocytes and leukocytes. Passing-Bablok analysis for leukocytes pointed to significant constant and minor proportional difference, while it was not performed for erythrocytes due to significant data deviation from linearity. Kappa analysis resulted in the strongest agreements for pH, ketones, glucose concentrations and leukocytes, while the poorest agreement for bacteria. The sensitivity and specificity of bacteria detection were: 91 (59-100)% and 76 (66-87)% for Atellica 1500 and 46 (17-77)% and 96 (87-100)% for Iris. CONCLUSION: There are large differences between Atellica 1500 and Iris analysers, due to which they are not comparable and can not be used interchangeably. While there was no difference in specificity of bacteria detection, Iris analyser had greater sensitivity.
format Online
Article
Text
id pubmed-8672386
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-86723862021-12-23 Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture Nikler, Ana Čičak, Helena Bejuk, Danijela Radišić Biljak, Vanja Šimundić, Ana-Maria Biochem Med (Zagreb) Original Articles INTRODUCTION: The aims of study were to assess: 1) performance specifications of Atellica 1500, 2) comparability of Atellica 1500 and Iris, 3) the accuracy of both analysers in their ability to detect bacteria. MATERIALS AND METHODS: Carryover, linearity, precision, reproducibility, and limit of blank (LoB) verification were evaluated for erythrocyte and leukocyte counts. ICSH 2014 protocol was used for estimation of carryover, CLSI EP15-A3 for precision, and CLSI EP17 for LoB verification. Comparison for quantitative parameters was evaluated by Bland-Altman plot and Passing-Bablok regression. Qualitative parameters were evaluated by Weighted kappa analysis. Sixty-five urine samples were randomly selected and sent for urine culture which was used as reference method to determine the accuracy of bacteria detection by analysers. RESULTS: Analytical specifications of Atellica 1500 were successfully verified. Total of 393 samples were used for qualitative comparison, while 269 for sediment urinalysis. Bland-Altman analysis showed statistically significant proportional bias for erythrocytes and leukocytes. Passing-Bablok analysis for leukocytes pointed to significant constant and minor proportional difference, while it was not performed for erythrocytes due to significant data deviation from linearity. Kappa analysis resulted in the strongest agreements for pH, ketones, glucose concentrations and leukocytes, while the poorest agreement for bacteria. The sensitivity and specificity of bacteria detection were: 91 (59-100)% and 76 (66-87)% for Atellica 1500 and 46 (17-77)% and 96 (87-100)% for Iris. CONCLUSION: There are large differences between Atellica 1500 and Iris analysers, due to which they are not comparable and can not be used interchangeably. While there was no difference in specificity of bacteria detection, Iris analyser had greater sensitivity. Croatian Society of Medical Biochemistry and Laboratory Medicine 2021-12-15 2022-02-15 /pmc/articles/PMC8672386/ /pubmed/34955669 http://dx.doi.org/10.11613/BM.2022.010701 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Nikler, Ana
Čičak, Helena
Bejuk, Danijela
Radišić Biljak, Vanja
Šimundić, Ana-Maria
Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture
title Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture
title_full Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture
title_fullStr Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture
title_full_unstemmed Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture
title_short Verification of Atellica 1500 and comparison with Iris urine analyser and urine culture
title_sort verification of atellica 1500 and comparison with iris urine analyser and urine culture
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672386/
https://www.ncbi.nlm.nih.gov/pubmed/34955669
http://dx.doi.org/10.11613/BM.2022.010701
work_keys_str_mv AT niklerana verificationofatellica1500andcomparisonwithirisurineanalyserandurineculture
AT cicakhelena verificationofatellica1500andcomparisonwithirisurineanalyserandurineculture
AT bejukdanijela verificationofatellica1500andcomparisonwithirisurineanalyserandurineculture
AT radisicbiljakvanja verificationofatellica1500andcomparisonwithirisurineanalyserandurineculture
AT simundicanamaria verificationofatellica1500andcomparisonwithirisurineanalyserandurineculture