Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study

Resuscitative endovascular balloon occlusion of the aorta (REBOA) for rapid hemorrhage control is increasingly being used in trauma management. Its beneficial hemodynamic effects on unstable patients beyond temporal hemostasis has led to growing interest in its use in other patient populations, such...

Descripción completa

Detalles Bibliográficos
Autores principales: Levis, Anja, Egli, Nives, Jenni, Hansjoerg, Hautz, Wolf E., Daley, James I., Haenggi, Matthias
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8674295/
https://www.ncbi.nlm.nih.gov/pubmed/34912008
http://dx.doi.org/10.1038/s41598-021-03502-6
_version_ 1784615620295262208
author Levis, Anja
Egli, Nives
Jenni, Hansjoerg
Hautz, Wolf E.
Daley, James I.
Haenggi, Matthias
author_facet Levis, Anja
Egli, Nives
Jenni, Hansjoerg
Hautz, Wolf E.
Daley, James I.
Haenggi, Matthias
author_sort Levis, Anja
collection PubMed
description Resuscitative endovascular balloon occlusion of the aorta (REBOA) for rapid hemorrhage control is increasingly being used in trauma management. Its beneficial hemodynamic effects on unstable patients beyond temporal hemostasis has led to growing interest in its use in other patient populations, such as during cardiac arrest from nontraumatic causes. The ability to insert the catheters without fluoroscopic guidance makes the technique available in the prehospital setting. However, in addition to correct positioning, challenges include reliably achieving aortic occlusion while minimizing the risk of balloon rupture. Without fluoroscopic control, inflation of the balloon relies on estimated aortic diameters and on the disappearing pulse in the contralateral femoral artery. In the case of cardiac arrest or absent palpable pulses, balloon inflation is associated with excess risk of overinflation and adverse events (vessel damage, balloon rupture). In this bench study, we examined how the pressure in the balloon is related to the surrounding blood pressure and the balloon's contact with the vessel wall in two sets of experiments, including a pulsatile circulation model. With this data, we developed a rule of thumb to guide balloon inflation of the ER-REBOA catheter with a simple disposable pressure-reading device (COMPASS). We recommend slowly filling the balloon with saline until the measured balloon pressure is 160 mmHg, or 16 mL of saline have been used. If after 16 mL the balloon pressure is still below 160 mmHg, saline should be added in 1-mL increments, which increases the pressure target about 10 mmHg at each step, until the maximum balloon pressure is reached at 240 mmHg (= 24 mL inflation volume). A balloon pressure greater than 250 mmHg indicates overinflation. With this rule and a disposable pressure-reading device (COMPASS), ER-REBOA balloons can be safely filled in austere environments where fluoroscopy is unavailable. Pressure monitoring of the balloon allows for recognition of unintended deflation or rupture of the balloon.
format Online
Article
Text
id pubmed-8674295
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-86742952021-12-16 Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study Levis, Anja Egli, Nives Jenni, Hansjoerg Hautz, Wolf E. Daley, James I. Haenggi, Matthias Sci Rep Article Resuscitative endovascular balloon occlusion of the aorta (REBOA) for rapid hemorrhage control is increasingly being used in trauma management. Its beneficial hemodynamic effects on unstable patients beyond temporal hemostasis has led to growing interest in its use in other patient populations, such as during cardiac arrest from nontraumatic causes. The ability to insert the catheters without fluoroscopic guidance makes the technique available in the prehospital setting. However, in addition to correct positioning, challenges include reliably achieving aortic occlusion while minimizing the risk of balloon rupture. Without fluoroscopic control, inflation of the balloon relies on estimated aortic diameters and on the disappearing pulse in the contralateral femoral artery. In the case of cardiac arrest or absent palpable pulses, balloon inflation is associated with excess risk of overinflation and adverse events (vessel damage, balloon rupture). In this bench study, we examined how the pressure in the balloon is related to the surrounding blood pressure and the balloon's contact with the vessel wall in two sets of experiments, including a pulsatile circulation model. With this data, we developed a rule of thumb to guide balloon inflation of the ER-REBOA catheter with a simple disposable pressure-reading device (COMPASS). We recommend slowly filling the balloon with saline until the measured balloon pressure is 160 mmHg, or 16 mL of saline have been used. If after 16 mL the balloon pressure is still below 160 mmHg, saline should be added in 1-mL increments, which increases the pressure target about 10 mmHg at each step, until the maximum balloon pressure is reached at 240 mmHg (= 24 mL inflation volume). A balloon pressure greater than 250 mmHg indicates overinflation. With this rule and a disposable pressure-reading device (COMPASS), ER-REBOA balloons can be safely filled in austere environments where fluoroscopy is unavailable. Pressure monitoring of the balloon allows for recognition of unintended deflation or rupture of the balloon. Nature Publishing Group UK 2021-12-15 /pmc/articles/PMC8674295/ /pubmed/34912008 http://dx.doi.org/10.1038/s41598-021-03502-6 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Article
Levis, Anja
Egli, Nives
Jenni, Hansjoerg
Hautz, Wolf E.
Daley, James I.
Haenggi, Matthias
Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study
title Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study
title_full Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study
title_fullStr Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study
title_full_unstemmed Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study
title_short Use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study
title_sort use of a disposable vascular pressure device to guide balloon inflation of resuscitative endovascular balloon occlusion of the aorta: a bench study
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8674295/
https://www.ncbi.nlm.nih.gov/pubmed/34912008
http://dx.doi.org/10.1038/s41598-021-03502-6
work_keys_str_mv AT levisanja useofadisposablevascularpressuredevicetoguideballooninflationofresuscitativeendovascularballoonocclusionoftheaortaabenchstudy
AT eglinives useofadisposablevascularpressuredevicetoguideballooninflationofresuscitativeendovascularballoonocclusionoftheaortaabenchstudy
AT jennihansjoerg useofadisposablevascularpressuredevicetoguideballooninflationofresuscitativeendovascularballoonocclusionoftheaortaabenchstudy
AT hautzwolfe useofadisposablevascularpressuredevicetoguideballooninflationofresuscitativeendovascularballoonocclusionoftheaortaabenchstudy
AT daleyjamesi useofadisposablevascularpressuredevicetoguideballooninflationofresuscitativeendovascularballoonocclusionoftheaortaabenchstudy
AT haenggimatthias useofadisposablevascularpressuredevicetoguideballooninflationofresuscitativeendovascularballoonocclusionoftheaortaabenchstudy