Accuracy and Reliability of Single-Camera Measurements of Ankle Clonus and Quadriceps Hyperreflexia
OBJECTIVE: To evaluate the accuracy and reliability of a simple, single-camera smartphone-based method, named the Reflex Tracker (RT) system, for measuring reflex threshold angles related to ankle clonus and quadriceps hyperreflexia. DESIGN: A prospective comparison study using a high-fidelity refer...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8683842/ https://www.ncbi.nlm.nih.gov/pubmed/34977536 http://dx.doi.org/10.1016/j.arrct.2021.100153 |
Sumario: | OBJECTIVE: To evaluate the accuracy and reliability of a simple, single-camera smartphone-based method, named the Reflex Tracker (RT) system, for measuring reflex threshold angles related to ankle clonus and quadriceps hyperreflexia. DESIGN: A prospective comparison study using a high-fidelity reference standard was constructed employing a 2 × 2 × 2 factorial design, with factors of rater (tester) type (student and experienced physical therapist), joint (ankle and knee), and repetition (2 per condition). SETTING: This multicenter study was conducted at 4 outpatient rehabilitation clinics. PARTICIPANTS: A convenience sample of 14 individuals with a neurologic condition presented with 20 lower limbs that exhibited ankle clonus and/or quadriceps hyperreflexia and were included in the study. Also participating in the study were 8 student and 8 experienced physical therapist raters (testers) (N=16). INTERVENTIONS: Not applicable. MAIN OUTCOME MEASURES: The plantar flexor reflex threshold angle (PFRTA) related to ankle clonus and the quadriceps reflex threshold angle (QRTA) related to quadriceps hyperreflexia were quantified. RESULTS: PFRTA and QRTA results were compared between the smartphone RT method and synchronous 3-dimensional inertial measurement unit (IMU) sensor motion capture. Mean difference (bias) was minimal between RT and IMU measurements for PFRTA (bias≤0.2°) and QRTA (bias≤1.2°). Intrarater reliability for PFRTA ranged from 0.85-0.90 using RT and from 0.85-0.87 using IMU; QRTA ranged from 0.97-0.98 using RT and from 0.96-0.99 using IMU. Intersensor reliability for PFRTA and QRTA was 0.97 and 0.99, respectively. Minimum detectable change for PFRTA ranged from 7.1°- 8.7° and for QRTA ranged from 6.1°-8.3°. CONCLUSIONS: RT performed comparable to IMU for accurate and reliable measurement of PFRTA and QRTA to quantify ankle clonus and quadriceps hyperreflexia in clinical settings. |
---|