Cargando…

Finite Element Analysis of Horizontal Screw-Screw Crosslink Used in C1–C2 Pedicle Screw-Rod Fixation

BACKGROUND: In the craniocervical junction, a C1–C2 pedicle screw-rod (PSR) fixation is applied to provide stability. The horizontal rod-rod crosslink (hR-R CL) is often used to enhance segmental posterior instrumentation. However, the biomechanics of the alternative horizontal screw-screw crosslink...

Descripción completa

Detalles Bibliográficos
Autores principales: Ouyang, Beiping, Zou, Xiaobao, Luo, Chunshan, Lu, Tingsheng, Xia, Hong, Ma, Xiangyang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Scientific Literature, Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8684241/
https://www.ncbi.nlm.nih.gov/pubmed/34903706
http://dx.doi.org/10.12659/MSM.932026
Descripción
Sumario:BACKGROUND: In the craniocervical junction, a C1–C2 pedicle screw-rod (PSR) fixation is applied to provide stability. The horizontal rod-rod crosslink (hR-R CL) is often used to enhance segmental posterior instrumentation. However, the biomechanics of the alternative horizontal screw-screw crosslink (hS-S CL) in the craniocervical junction are unclear. MATERIAL/METHODS: A nonlinear atlantoaxial instability 3-dimensional C1–C2 finite element model was constructed using computed tomography images. On this basis, 2 fixation models were established with C1–C2 PSR fixation using (1) a rod-rod crosslink (R-R CL), and (2) a screw-screw crosslink (S-S CL). Range of motion (ROM) of the atlantoaxial joint, stress distribution of the implants, and maximum stress value of the vertebral bodies were calculated and compared under 4 loading conditions, including flexion, extension, lateral bending, and axial rotation. RESULTS: Atlantoaxial joint ROM was reduced by 90.19% to 98.5% with the hR-R CL, and by 90.1% to 98.7% with the hS-S CL, compared with the instability model. During axial rotation, the total stress peak of the PSR fixation was smaller with hS-S CL than with hR-R CL. The peak stress values of the vertebral bodies were comparable between the 2 fixation models. CONCLUSIONS: The 2 tested crosslink models provided comparable stability. However, during axial rotation, the total stress peak of hS-S CL fixation was smaller than that of hR-R CL fixation. Since the atlantoaxial joint primarily functions as a rotational joint, our results suggested that the use of hS-S CL can provide a more stable environment for the implants.