Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine
Rapid, accurate detection of serum amyloid A (SAA) is needed in equine practice. We validated a patient-side point-of-care (POC) assay (Stablelab; Zoetis) compared to the turbidimetric immunoassays LZ-SAA (TIA-Hum) and VET-SAA (TIA-Vet; both Eiken Chemical). Analytical performance was assessed at 3...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8688985/ https://www.ncbi.nlm.nih.gov/pubmed/34763564 http://dx.doi.org/10.1177/10406387211056029 |
_version_ | 1784618464587022336 |
---|---|
author | Kiemle, Julia Hindenberg, Sarah Bauer, Natali Roecken, Michael |
author_facet | Kiemle, Julia Hindenberg, Sarah Bauer, Natali Roecken, Michael |
author_sort | Kiemle, Julia |
collection | PubMed |
description | Rapid, accurate detection of serum amyloid A (SAA) is needed in equine practice. We validated a patient-side point-of-care (POC) assay (Stablelab; Zoetis) compared to the turbidimetric immunoassays LZ-SAA (TIA-Hum) and VET-SAA (TIA-Vet; both Eiken Chemical). Analytical performance was assessed at 3 different concentration ranges and with interferences. Inter-method comparison using 49 equine serum samples revealed a significant difference between median SAA results (p < 0.0001), with the strongest bias between the POC and TIA-Vet (median 1,093 vs. 578 mg/L). The median SAA value obtained with the TIA-Hum method was 752 mg/L. Correlation between POC/TIA-Hum and between POC/TIA-Vet was fair (r(s) = 0.77 and 0.69) and excellent between both TIAs (r(s) = 0.93). Bias between POC/TIA-Hum, POC/TIA-Vet, and TIA-Hum/TIA-Vet was −56.7%, –80.9%, and −28.2%, respectively. POC intra- and inter-assay CVs (16.1–30% and 19.8–35.5%) were higher than TIA CVs (generally <12%). Bilirubin and hemoglobin had a negative bias on POC and TIA-Vet results (−16.6 to −45.6%); addition of intralipid yielded a positive bias (35.9–77.4%). The POC had good linearity of SAA concentrations up to 10,312 mg/L (R(2) = 0.92). A hook effect was present at SAA >3,000 mg/L for the POC assay. Equine serum SAA was stable over a median period of 2.5 y when stored at −80°C. Overall, there was excellent-to-moderate correlation between tests, but imprecision and hook effect of the POC, as well as bias between the methods, must be considered. |
format | Online Article Text |
id | pubmed-8688985 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-86889852021-12-22 Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine Kiemle, Julia Hindenberg, Sarah Bauer, Natali Roecken, Michael J Vet Diagn Invest Full Scientific Reports Rapid, accurate detection of serum amyloid A (SAA) is needed in equine practice. We validated a patient-side point-of-care (POC) assay (Stablelab; Zoetis) compared to the turbidimetric immunoassays LZ-SAA (TIA-Hum) and VET-SAA (TIA-Vet; both Eiken Chemical). Analytical performance was assessed at 3 different concentration ranges and with interferences. Inter-method comparison using 49 equine serum samples revealed a significant difference between median SAA results (p < 0.0001), with the strongest bias between the POC and TIA-Vet (median 1,093 vs. 578 mg/L). The median SAA value obtained with the TIA-Hum method was 752 mg/L. Correlation between POC/TIA-Hum and between POC/TIA-Vet was fair (r(s) = 0.77 and 0.69) and excellent between both TIAs (r(s) = 0.93). Bias between POC/TIA-Hum, POC/TIA-Vet, and TIA-Hum/TIA-Vet was −56.7%, –80.9%, and −28.2%, respectively. POC intra- and inter-assay CVs (16.1–30% and 19.8–35.5%) were higher than TIA CVs (generally <12%). Bilirubin and hemoglobin had a negative bias on POC and TIA-Vet results (−16.6 to −45.6%); addition of intralipid yielded a positive bias (35.9–77.4%). The POC had good linearity of SAA concentrations up to 10,312 mg/L (R(2) = 0.92). A hook effect was present at SAA >3,000 mg/L for the POC assay. Equine serum SAA was stable over a median period of 2.5 y when stored at −80°C. Overall, there was excellent-to-moderate correlation between tests, but imprecision and hook effect of the POC, as well as bias between the methods, must be considered. SAGE Publications 2021-11-11 2022-01 /pmc/articles/PMC8688985/ /pubmed/34763564 http://dx.doi.org/10.1177/10406387211056029 Text en © 2021 The Author(s) https://creativecommons.org/licenses/by/4.0/This article is distributed under the terms of the Creative Commons Attribution 4.0 License (https://creativecommons.org/licenses/by/4.0/) which permits any use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access page (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Full Scientific Reports Kiemle, Julia Hindenberg, Sarah Bauer, Natali Roecken, Michael Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine |
title | Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine |
title_full | Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine |
title_fullStr | Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine |
title_full_unstemmed | Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine |
title_short | Comparison of a point-of-care serum amyloid A analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine |
title_sort | comparison of a point-of-care serum amyloid a analyzer frequently used in equine practice with 2 turbidimetric immunoassays used in human and veterinary medicine |
topic | Full Scientific Reports |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8688985/ https://www.ncbi.nlm.nih.gov/pubmed/34763564 http://dx.doi.org/10.1177/10406387211056029 |
work_keys_str_mv | AT kiemlejulia comparisonofapointofcareserumamyloidaanalyzerfrequentlyusedinequinepracticewith2turbidimetricimmunoassaysusedinhumanandveterinarymedicine AT hindenbergsarah comparisonofapointofcareserumamyloidaanalyzerfrequentlyusedinequinepracticewith2turbidimetricimmunoassaysusedinhumanandveterinarymedicine AT bauernatali comparisonofapointofcareserumamyloidaanalyzerfrequentlyusedinequinepracticewith2turbidimetricimmunoassaysusedinhumanandveterinarymedicine AT roeckenmichael comparisonofapointofcareserumamyloidaanalyzerfrequentlyusedinequinepracticewith2turbidimetricimmunoassaysusedinhumanandveterinarymedicine |