Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera
The aim of this study was to evaluate the quantitative values of short-time scan (STS) of metastatic lesions compared with a standard scan (SS) when acquired by whole-body bone SPECT/CT with cadmium–zinc–telluride (CZT) detectors. We retrospectively reviewed 13 patients with bone metastases from pro...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8692318/ https://www.ncbi.nlm.nih.gov/pubmed/34934145 http://dx.doi.org/10.1038/s41598-021-03853-0 |
_version_ | 1784618935136550912 |
---|---|
author | Yamane, Tomohiko Takahashi, Masafumi Matsusaka, Yohji Fukushima, Kenji Seto, Akira Kuji, Ichiei Matsunari, Ichiro |
author_facet | Yamane, Tomohiko Takahashi, Masafumi Matsusaka, Yohji Fukushima, Kenji Seto, Akira Kuji, Ichiei Matsunari, Ichiro |
author_sort | Yamane, Tomohiko |
collection | PubMed |
description | The aim of this study was to evaluate the quantitative values of short-time scan (STS) of metastatic lesions compared with a standard scan (SS) when acquired by whole-body bone SPECT/CT with cadmium–zinc–telluride (CZT) detectors. We retrospectively reviewed 13 patients with bone metastases from prostate cancer, who underwent SPECT/CT performed on whole-body CZT gamma cameras. STSs were obtained using 75, 50, 25, 10, and 5% of the list-mode data for SS, respectively. Regions of interest (ROIs) were set on the increased uptake areas diagnosed as metastases. Intraclass correlation coefficients (ICCs) of standardized uptake values (SUVs) for the ROIs were calculated between the SS and each STS, and ICC ≥ 0.8 was set as a perfect correlation. Moreover, the repeatability coefficient (RC) was calculated, and RC ≤ 20% was defined as acceptable. A total of 152 metastatic lesions were included in the analysis. The ICCs between the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 0.999, 0.997, 0.994, 0.983, and 0.955, respectively. The RCs of the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 7.9, 12.4, 19.8, 30.8, and 41.3%, respectively. When evaluating the quality of CZT bone SPECT/CT acquired by a standard protocol, 25%-STS may provide adequate quantitative values. |
format | Online Article Text |
id | pubmed-8692318 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Nature Publishing Group UK |
record_format | MEDLINE/PubMed |
spelling | pubmed-86923182021-12-22 Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera Yamane, Tomohiko Takahashi, Masafumi Matsusaka, Yohji Fukushima, Kenji Seto, Akira Kuji, Ichiei Matsunari, Ichiro Sci Rep Article The aim of this study was to evaluate the quantitative values of short-time scan (STS) of metastatic lesions compared with a standard scan (SS) when acquired by whole-body bone SPECT/CT with cadmium–zinc–telluride (CZT) detectors. We retrospectively reviewed 13 patients with bone metastases from prostate cancer, who underwent SPECT/CT performed on whole-body CZT gamma cameras. STSs were obtained using 75, 50, 25, 10, and 5% of the list-mode data for SS, respectively. Regions of interest (ROIs) were set on the increased uptake areas diagnosed as metastases. Intraclass correlation coefficients (ICCs) of standardized uptake values (SUVs) for the ROIs were calculated between the SS and each STS, and ICC ≥ 0.8 was set as a perfect correlation. Moreover, the repeatability coefficient (RC) was calculated, and RC ≤ 20% was defined as acceptable. A total of 152 metastatic lesions were included in the analysis. The ICCs between the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 0.999, 0.997, 0.994, 0.983, and 0.955, respectively. The RCs of the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 7.9, 12.4, 19.8, 30.8, and 41.3%, respectively. When evaluating the quality of CZT bone SPECT/CT acquired by a standard protocol, 25%-STS may provide adequate quantitative values. Nature Publishing Group UK 2021-12-21 /pmc/articles/PMC8692318/ /pubmed/34934145 http://dx.doi.org/10.1038/s41598-021-03853-0 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Article Yamane, Tomohiko Takahashi, Masafumi Matsusaka, Yohji Fukushima, Kenji Seto, Akira Kuji, Ichiei Matsunari, Ichiro Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera |
title | Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera |
title_full | Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera |
title_fullStr | Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera |
title_full_unstemmed | Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera |
title_short | Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium–zinc–telluride gamma camera |
title_sort | satisfied quantitative value can be acquired by short-time bone spect/ct using a whole-body cadmium–zinc–telluride gamma camera |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8692318/ https://www.ncbi.nlm.nih.gov/pubmed/34934145 http://dx.doi.org/10.1038/s41598-021-03853-0 |
work_keys_str_mv | AT yamanetomohiko satisfiedquantitativevaluecanbeacquiredbyshorttimebonespectctusingawholebodycadmiumzinctelluridegammacamera AT takahashimasafumi satisfiedquantitativevaluecanbeacquiredbyshorttimebonespectctusingawholebodycadmiumzinctelluridegammacamera AT matsusakayohji satisfiedquantitativevaluecanbeacquiredbyshorttimebonespectctusingawholebodycadmiumzinctelluridegammacamera AT fukushimakenji satisfiedquantitativevaluecanbeacquiredbyshorttimebonespectctusingawholebodycadmiumzinctelluridegammacamera AT setoakira satisfiedquantitativevaluecanbeacquiredbyshorttimebonespectctusingawholebodycadmiumzinctelluridegammacamera AT kujiichiei satisfiedquantitativevaluecanbeacquiredbyshorttimebonespectctusingawholebodycadmiumzinctelluridegammacamera AT matsunariichiro satisfiedquantitativevaluecanbeacquiredbyshorttimebonespectctusingawholebodycadmiumzinctelluridegammacamera |