Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation
Atrial fibrillation (AF) is the most common cardiac arrhythmia and catheter mapping has been proved to be an effective approach for detecting AF drivers to be targeted by ablation. Among drivers, the so-called rotors have gained the most attention: their identification and spatial location could hel...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8696082/ https://www.ncbi.nlm.nih.gov/pubmed/34955872 http://dx.doi.org/10.3389/fphys.2021.732161 |
_version_ | 1784619727133343744 |
---|---|
author | Bartolucci, Chiara Fabbri, Claudio Tomasi, Corrado Sabbatani, Paolo Severi, Stefano Corsi, Cristiana |
author_facet | Bartolucci, Chiara Fabbri, Claudio Tomasi, Corrado Sabbatani, Paolo Severi, Stefano Corsi, Cristiana |
author_sort | Bartolucci, Chiara |
collection | PubMed |
description | Atrial fibrillation (AF) is the most common cardiac arrhythmia and catheter mapping has been proved to be an effective approach for detecting AF drivers to be targeted by ablation. Among drivers, the so-called rotors have gained the most attention: their identification and spatial location could help to understand which patient-specific mechanisms are acting, and thus to guide the ablation execution. Since rotor detection by multi-electrode catheters may be influenced by several structural parameters including inter-electrode spacing, catheter coverage, and endocardium-catheter distance, in this study we proposed a tool for testing the ability of different catheter shapes to detect rotors in different conditions. An approach based on the solution of the monodomain equations coupled with a modified Courtemanche ionic atrial model, that considers an electrical remodeling, was applied to simulate spiral wave dynamics on a 2D model for 7.75 s. The developed framework allowed the acquisition of unipolar signals at 2 KHz. Two high-density multipolar catheters were simulated (Advisor™ HD Grid and PentaRay(®)) and placed in a 2D region in which the simulated spiral wave persists longer. The configuration of the catheters was then modified by changing the number of electrodes, inter-electrodes distance, position, and atrial-wall distance for assessing how they would affect the rotor detection. In contact with the wall and at 1 mm distance from it, all the configurations detected the rotor correctly, irrespective of geometry, coverage, and inter-electrode distance. In the HDGrid-like geometry, the increase of the inter-electrode distance from 3 to 6 mm caused rotor detection failure at 2 mm distance from the LA wall. In the PentaRay-like configuration, regardless of inter-electrode distance, rotor detection failed at 3 mm endocardium-catheter distance. The asymmetry of this catheter resulted in rotation-dependent rotor detection. To conclude, the computational framework we developed is based on realistic catheter shapes designed with parameter configurations which resemble clinical settings. Results showed it is well suited to investigate how mapping catheter geometry and location affect AF driver detection, therefore it is a reliable tool to design and test new mapping catheters. |
format | Online Article Text |
id | pubmed-8696082 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-86960822021-12-24 Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation Bartolucci, Chiara Fabbri, Claudio Tomasi, Corrado Sabbatani, Paolo Severi, Stefano Corsi, Cristiana Front Physiol Physiology Atrial fibrillation (AF) is the most common cardiac arrhythmia and catheter mapping has been proved to be an effective approach for detecting AF drivers to be targeted by ablation. Among drivers, the so-called rotors have gained the most attention: their identification and spatial location could help to understand which patient-specific mechanisms are acting, and thus to guide the ablation execution. Since rotor detection by multi-electrode catheters may be influenced by several structural parameters including inter-electrode spacing, catheter coverage, and endocardium-catheter distance, in this study we proposed a tool for testing the ability of different catheter shapes to detect rotors in different conditions. An approach based on the solution of the monodomain equations coupled with a modified Courtemanche ionic atrial model, that considers an electrical remodeling, was applied to simulate spiral wave dynamics on a 2D model for 7.75 s. The developed framework allowed the acquisition of unipolar signals at 2 KHz. Two high-density multipolar catheters were simulated (Advisor™ HD Grid and PentaRay(®)) and placed in a 2D region in which the simulated spiral wave persists longer. The configuration of the catheters was then modified by changing the number of electrodes, inter-electrodes distance, position, and atrial-wall distance for assessing how they would affect the rotor detection. In contact with the wall and at 1 mm distance from it, all the configurations detected the rotor correctly, irrespective of geometry, coverage, and inter-electrode distance. In the HDGrid-like geometry, the increase of the inter-electrode distance from 3 to 6 mm caused rotor detection failure at 2 mm distance from the LA wall. In the PentaRay-like configuration, regardless of inter-electrode distance, rotor detection failed at 3 mm endocardium-catheter distance. The asymmetry of this catheter resulted in rotation-dependent rotor detection. To conclude, the computational framework we developed is based on realistic catheter shapes designed with parameter configurations which resemble clinical settings. Results showed it is well suited to investigate how mapping catheter geometry and location affect AF driver detection, therefore it is a reliable tool to design and test new mapping catheters. Frontiers Media S.A. 2021-12-09 /pmc/articles/PMC8696082/ /pubmed/34955872 http://dx.doi.org/10.3389/fphys.2021.732161 Text en Copyright © 2021 Bartolucci, Fabbri, Tomasi, Sabbatani, Severi and Corsi. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Physiology Bartolucci, Chiara Fabbri, Claudio Tomasi, Corrado Sabbatani, Paolo Severi, Stefano Corsi, Cristiana Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation |
title | Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation |
title_full | Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation |
title_fullStr | Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation |
title_full_unstemmed | Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation |
title_short | Computational Analysis of Mapping Catheter Geometry and Contact Quality Effects on Rotor Detection in Atrial Fibrillation |
title_sort | computational analysis of mapping catheter geometry and contact quality effects on rotor detection in atrial fibrillation |
topic | Physiology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8696082/ https://www.ncbi.nlm.nih.gov/pubmed/34955872 http://dx.doi.org/10.3389/fphys.2021.732161 |
work_keys_str_mv | AT bartoluccichiara computationalanalysisofmappingcathetergeometryandcontactqualityeffectsonrotordetectioninatrialfibrillation AT fabbriclaudio computationalanalysisofmappingcathetergeometryandcontactqualityeffectsonrotordetectioninatrialfibrillation AT tomasicorrado computationalanalysisofmappingcathetergeometryandcontactqualityeffectsonrotordetectioninatrialfibrillation AT sabbatanipaolo computationalanalysisofmappingcathetergeometryandcontactqualityeffectsonrotordetectioninatrialfibrillation AT severistefano computationalanalysisofmappingcathetergeometryandcontactqualityeffectsonrotordetectioninatrialfibrillation AT corsicristiana computationalanalysisofmappingcathetergeometryandcontactqualityeffectsonrotordetectioninatrialfibrillation |