Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement

INTRODUCTION: Haemorrhage is a leading cause of death following traumatic injury and the early detection of hypovolaemia is critical to effective management. However, accurate assessment of circulating blood volume is challenging when using traditional vital signs such as blood pressure. We conducte...

Descripción completa

Detalles Bibliográficos
Autores principales: Hutchings, Sam D., Watchorn, Jim, McDonald, Rory, Jeffreys, Su, Bates, Mark, Watts, Sarah, Kirkman, Emrys
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8699736/
https://www.ncbi.nlm.nih.gov/pubmed/34941918
http://dx.doi.org/10.1371/journal.pone.0261546
_version_ 1784620585533308928
author Hutchings, Sam D.
Watchorn, Jim
McDonald, Rory
Jeffreys, Su
Bates, Mark
Watts, Sarah
Kirkman, Emrys
author_facet Hutchings, Sam D.
Watchorn, Jim
McDonald, Rory
Jeffreys, Su
Bates, Mark
Watts, Sarah
Kirkman, Emrys
author_sort Hutchings, Sam D.
collection PubMed
description INTRODUCTION: Haemorrhage is a leading cause of death following traumatic injury and the early detection of hypovolaemia is critical to effective management. However, accurate assessment of circulating blood volume is challenging when using traditional vital signs such as blood pressure. We conducted a study to compare the stroke volume (SV) recorded using two devices, trans-thoracic electrical bioimpedance (TEB) and supra-sternal Doppler (SSD), against a reference standard using trans- thoracic echocardiography (TTE). METHODS: A lower body negative pressure (LBNP) model was used to simulate hypovolaemia and in half of the study sessions lower limb tourniquets were applied as these are common in military practice and can potentially affect some haemodynamic monitoring systems. In order to provide a clinically relevant comparison we constructed an error grid alongside more traditional measures of agreement. RESULTS: 21 healthy volunteers aged 18–40 were enrolled and underwent 2 sessions of LBNP, with and without lower limb tourniquets. With respect to absolute SV values Bland Altman analysis showed significant bias in both non-tourniquet and tourniquet strands for TEB (-42.5 / -49.6 ml), rendering further analysis impossible. For SSD bias was minimal but percentage error was unacceptably high (35% / 48%). Degree of agreement for dynamic change in SV, assessed using 4 quadrant plots showed a seemingly acceptable concordance rate for both TEB (86% / 93%) and SSD (90% / 91%). However, when results were plotted on an error grid, constructed based on expert clinical opinion, a significant minority of measurement errors were identified that had potential to lead to moderate or severe patient harm. CONCLUSION: Thoracic bioimpedance and suprasternal Doppler both demonstrated measurement errors that had the potential to lead to clinical harm and caution should be applied in interpreting the results in the detection of early hypovolaemia following traumatic injury.
format Online
Article
Text
id pubmed-8699736
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-86997362021-12-24 Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement Hutchings, Sam D. Watchorn, Jim McDonald, Rory Jeffreys, Su Bates, Mark Watts, Sarah Kirkman, Emrys PLoS One Research Article INTRODUCTION: Haemorrhage is a leading cause of death following traumatic injury and the early detection of hypovolaemia is critical to effective management. However, accurate assessment of circulating blood volume is challenging when using traditional vital signs such as blood pressure. We conducted a study to compare the stroke volume (SV) recorded using two devices, trans-thoracic electrical bioimpedance (TEB) and supra-sternal Doppler (SSD), against a reference standard using trans- thoracic echocardiography (TTE). METHODS: A lower body negative pressure (LBNP) model was used to simulate hypovolaemia and in half of the study sessions lower limb tourniquets were applied as these are common in military practice and can potentially affect some haemodynamic monitoring systems. In order to provide a clinically relevant comparison we constructed an error grid alongside more traditional measures of agreement. RESULTS: 21 healthy volunteers aged 18–40 were enrolled and underwent 2 sessions of LBNP, with and without lower limb tourniquets. With respect to absolute SV values Bland Altman analysis showed significant bias in both non-tourniquet and tourniquet strands for TEB (-42.5 / -49.6 ml), rendering further analysis impossible. For SSD bias was minimal but percentage error was unacceptably high (35% / 48%). Degree of agreement for dynamic change in SV, assessed using 4 quadrant plots showed a seemingly acceptable concordance rate for both TEB (86% / 93%) and SSD (90% / 91%). However, when results were plotted on an error grid, constructed based on expert clinical opinion, a significant minority of measurement errors were identified that had potential to lead to moderate or severe patient harm. CONCLUSION: Thoracic bioimpedance and suprasternal Doppler both demonstrated measurement errors that had the potential to lead to clinical harm and caution should be applied in interpreting the results in the detection of early hypovolaemia following traumatic injury. Public Library of Science 2021-12-23 /pmc/articles/PMC8699736/ /pubmed/34941918 http://dx.doi.org/10.1371/journal.pone.0261546 Text en © 2021 Hutchings et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Hutchings, Sam D.
Watchorn, Jim
McDonald, Rory
Jeffreys, Su
Bates, Mark
Watts, Sarah
Kirkman, Emrys
Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement
title Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement
title_full Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement
title_fullStr Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement
title_full_unstemmed Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement
title_short Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement
title_sort quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8699736/
https://www.ncbi.nlm.nih.gov/pubmed/34941918
http://dx.doi.org/10.1371/journal.pone.0261546
work_keys_str_mv AT hutchingssamd quantificationofstrokevolumeinasimulatedhealthyvolunteermodeloftraumatichaemorrhageacomparisonoftwononinvasivemonitoringdevicesusingerrorgridanalysisalongsidetraditionalmeasuresofagreement
AT watchornjim quantificationofstrokevolumeinasimulatedhealthyvolunteermodeloftraumatichaemorrhageacomparisonoftwononinvasivemonitoringdevicesusingerrorgridanalysisalongsidetraditionalmeasuresofagreement
AT mcdonaldrory quantificationofstrokevolumeinasimulatedhealthyvolunteermodeloftraumatichaemorrhageacomparisonoftwononinvasivemonitoringdevicesusingerrorgridanalysisalongsidetraditionalmeasuresofagreement
AT jeffreyssu quantificationofstrokevolumeinasimulatedhealthyvolunteermodeloftraumatichaemorrhageacomparisonoftwononinvasivemonitoringdevicesusingerrorgridanalysisalongsidetraditionalmeasuresofagreement
AT batesmark quantificationofstrokevolumeinasimulatedhealthyvolunteermodeloftraumatichaemorrhageacomparisonoftwononinvasivemonitoringdevicesusingerrorgridanalysisalongsidetraditionalmeasuresofagreement
AT wattssarah quantificationofstrokevolumeinasimulatedhealthyvolunteermodeloftraumatichaemorrhageacomparisonoftwononinvasivemonitoringdevicesusingerrorgridanalysisalongsidetraditionalmeasuresofagreement
AT kirkmanemrys quantificationofstrokevolumeinasimulatedhealthyvolunteermodeloftraumatichaemorrhageacomparisonoftwononinvasivemonitoringdevicesusingerrorgridanalysisalongsidetraditionalmeasuresofagreement