Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors

This study aimed at establishing native T1 reference values for a Canon Vantage Galan 3T system and comparing them with previously published values from different vendors. A total of 20 healthy volunteers (55% Women; 33.9 ± 11.1 years) underwent left ventricular T1 mapping at 3T MR. A MOLLI 5(3)3 se...

Descripción completa

Detalles Bibliográficos
Autores principales: Tribuna, Liliana, Oliveira, Pedro Belo, Iruela, Alba, Marques, João, Santos, Paulo, Teixeira, Tiago
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8699831/
https://www.ncbi.nlm.nih.gov/pubmed/34943571
http://dx.doi.org/10.3390/diagnostics11122334
_version_ 1784620608695304192
author Tribuna, Liliana
Oliveira, Pedro Belo
Iruela, Alba
Marques, João
Santos, Paulo
Teixeira, Tiago
author_facet Tribuna, Liliana
Oliveira, Pedro Belo
Iruela, Alba
Marques, João
Santos, Paulo
Teixeira, Tiago
author_sort Tribuna, Liliana
collection PubMed
description This study aimed at establishing native T1 reference values for a Canon Vantage Galan 3T system and comparing them with previously published values from different vendors. A total of 20 healthy volunteers (55% Women; 33.9 ± 11.1 years) underwent left ventricular T1 mapping at 3T MR. A MOLLI 5(3)3 sequence was used, acquiring three short-axis slices. Native T1 values are shown as means (±standard deviation) and Student’s independent samples t-test was used to test gender differences in T1 values. Pearson’s correlation coefficient analysis was used to compare two processes of T1 analysis. The results show a global native T1 mean value of 1124.9 ± 55.2 ms (exponential analysis), that of women being statistically higher than men (1163 ± 30.5 vs. 1077.9 ± 39.5 ms, respectively; p < 0.001). There were no specific tendencies for T1 times in different ventricular slices. We found a strong correlation (0.977, p < 0.001) with T1 times derived from parametric maps (1136.4 ± 60.2 ms). Native T1 reference values for a Canon 3T scanner were provided, and they are on par with those already reported from other vendors for a similar sequence. We also found a correlation between native T1 and gender, with higher values for women.
format Online
Article
Text
id pubmed-8699831
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-86998312021-12-24 Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors Tribuna, Liliana Oliveira, Pedro Belo Iruela, Alba Marques, João Santos, Paulo Teixeira, Tiago Diagnostics (Basel) Article This study aimed at establishing native T1 reference values for a Canon Vantage Galan 3T system and comparing them with previously published values from different vendors. A total of 20 healthy volunteers (55% Women; 33.9 ± 11.1 years) underwent left ventricular T1 mapping at 3T MR. A MOLLI 5(3)3 sequence was used, acquiring three short-axis slices. Native T1 values are shown as means (±standard deviation) and Student’s independent samples t-test was used to test gender differences in T1 values. Pearson’s correlation coefficient analysis was used to compare two processes of T1 analysis. The results show a global native T1 mean value of 1124.9 ± 55.2 ms (exponential analysis), that of women being statistically higher than men (1163 ± 30.5 vs. 1077.9 ± 39.5 ms, respectively; p < 0.001). There were no specific tendencies for T1 times in different ventricular slices. We found a strong correlation (0.977, p < 0.001) with T1 times derived from parametric maps (1136.4 ± 60.2 ms). Native T1 reference values for a Canon 3T scanner were provided, and they are on par with those already reported from other vendors for a similar sequence. We also found a correlation between native T1 and gender, with higher values for women. MDPI 2021-12-11 /pmc/articles/PMC8699831/ /pubmed/34943571 http://dx.doi.org/10.3390/diagnostics11122334 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Tribuna, Liliana
Oliveira, Pedro Belo
Iruela, Alba
Marques, João
Santos, Paulo
Teixeira, Tiago
Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors
title Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors
title_full Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors
title_fullStr Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors
title_full_unstemmed Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors
title_short Reference Values of Native T1 at 3T Cardiac Magnetic Resonance—Standardization Considerations between Different Vendors
title_sort reference values of native t1 at 3t cardiac magnetic resonance—standardization considerations between different vendors
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8699831/
https://www.ncbi.nlm.nih.gov/pubmed/34943571
http://dx.doi.org/10.3390/diagnostics11122334
work_keys_str_mv AT tribunaliliana referencevaluesofnativet1at3tcardiacmagneticresonancestandardizationconsiderationsbetweendifferentvendors
AT oliveirapedrobelo referencevaluesofnativet1at3tcardiacmagneticresonancestandardizationconsiderationsbetweendifferentvendors
AT iruelaalba referencevaluesofnativet1at3tcardiacmagneticresonancestandardizationconsiderationsbetweendifferentvendors
AT marquesjoao referencevaluesofnativet1at3tcardiacmagneticresonancestandardizationconsiderationsbetweendifferentvendors
AT santospaulo referencevaluesofnativet1at3tcardiacmagneticresonancestandardizationconsiderationsbetweendifferentvendors
AT teixeiratiago referencevaluesofnativet1at3tcardiacmagneticresonancestandardizationconsiderationsbetweendifferentvendors