Cargando…
Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery
CATEGORY: Other; Ankle INTRODUCTION/PURPOSE: Fluoroscopy is critical for orthopaedic surgery. The mini C-arm is popular among foot and ankle surgeons owing ease of easy, maneuverability, and cost effectiveness. However, there are concerns that the mini C-arm may represent a potential source of conta...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8705516/ http://dx.doi.org/10.1177/2473011420S00062 |
_version_ | 1784621965992001536 |
---|---|
author | Momenzadeh, Kaveh Williams, Caroline Czerwonka, Natalia Alemi, Mohammad Mehdi Kwon, John Y. Nazarian, Ara Miller, Christopher P. |
author_facet | Momenzadeh, Kaveh Williams, Caroline Czerwonka, Natalia Alemi, Mohammad Mehdi Kwon, John Y. Nazarian, Ara Miller, Christopher P. |
author_sort | Momenzadeh, Kaveh |
collection | PubMed |
description | CATEGORY: Other; Ankle INTRODUCTION/PURPOSE: Fluoroscopy is critical for orthopaedic surgery. The mini C-arm is popular among foot and ankle surgeons owing ease of easy, maneuverability, and cost effectiveness. However, there are concerns that the mini C-arm may represent a potential source of contamination and increasing the risk of postoperative infection. Evaluation of large c-arm drapes have shown high rates of contamination. However, the mini C-arm may be even more concerning due to how it is handled. The mini c-arm is continuously manipulated by the surgical team who then return to the surgical field carrying with them any potential cross-contamination.Objective: We hypothesize that there is significant contamination of the drape, and that specific locations of the drape are at higher risk. METHODS: 50 foot and ankle surgery cases which required the use of mini C-arm fluoroscopy were included. Eight locations on the mini C-arm drape were chosen for sampling. Locations were defined based on commonly touched areas on the drape by the surgical team, and their proximity to the non-sterile field. Negative controls were obtained by separately sampling an additional 20 mini C-arm drapes after draping c-arm and waiting for average surgery time. Positive controls were obtained by sampling non- sterile surfaces within the OR.Culture Q-swabs were used for sampling defined locations, using an established semiquantitative technique. Plates were incubated at 37 °C for 48 hours. Based on the number of quadrants on plate where colony growth was observed, bacterial growth was graded on a scale of 0, 1+, 2+, 3+, or 4+. Contamination threshold was defined as any growth pattern of 1+ or higher. RESULTS: Figure 1 shows the contamination rates of the sampled areas of the c-arm drape. Contamination was noted in 70% of the cases for at least one area, and in two or more areas in 38% of cases.Contamination rates were significantly higher on defined locations in comparison with their relative negative control, except for locations 4 and 5. (P<0.05)The most frequently contaminated location was location #2 (30%) followed by Location 1 (26%), and then location 8 (22%). (P<0.05) CONCLUSION: Bacterial contamination of the drape was found to be common (70% of cases) after foot and ankle surgery. Contamination was more common on the superior and inferior aspects of the mini c-arm. This correlates to the area that is most commonly manipulated by the surgical team to position the arms of the c-arm. This information can be utilized by surgeons to identify and avoid high risk areas and, in high risk cases, they may consider changing gloves or re-draping the mini c-arm part way through the case. |
format | Online Article Text |
id | pubmed-8705516 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-87055162022-01-28 Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery Momenzadeh, Kaveh Williams, Caroline Czerwonka, Natalia Alemi, Mohammad Mehdi Kwon, John Y. Nazarian, Ara Miller, Christopher P. Foot Ankle Orthop Article CATEGORY: Other; Ankle INTRODUCTION/PURPOSE: Fluoroscopy is critical for orthopaedic surgery. The mini C-arm is popular among foot and ankle surgeons owing ease of easy, maneuverability, and cost effectiveness. However, there are concerns that the mini C-arm may represent a potential source of contamination and increasing the risk of postoperative infection. Evaluation of large c-arm drapes have shown high rates of contamination. However, the mini C-arm may be even more concerning due to how it is handled. The mini c-arm is continuously manipulated by the surgical team who then return to the surgical field carrying with them any potential cross-contamination.Objective: We hypothesize that there is significant contamination of the drape, and that specific locations of the drape are at higher risk. METHODS: 50 foot and ankle surgery cases which required the use of mini C-arm fluoroscopy were included. Eight locations on the mini C-arm drape were chosen for sampling. Locations were defined based on commonly touched areas on the drape by the surgical team, and their proximity to the non-sterile field. Negative controls were obtained by separately sampling an additional 20 mini C-arm drapes after draping c-arm and waiting for average surgery time. Positive controls were obtained by sampling non- sterile surfaces within the OR.Culture Q-swabs were used for sampling defined locations, using an established semiquantitative technique. Plates were incubated at 37 °C for 48 hours. Based on the number of quadrants on plate where colony growth was observed, bacterial growth was graded on a scale of 0, 1+, 2+, 3+, or 4+. Contamination threshold was defined as any growth pattern of 1+ or higher. RESULTS: Figure 1 shows the contamination rates of the sampled areas of the c-arm drape. Contamination was noted in 70% of the cases for at least one area, and in two or more areas in 38% of cases.Contamination rates were significantly higher on defined locations in comparison with their relative negative control, except for locations 4 and 5. (P<0.05)The most frequently contaminated location was location #2 (30%) followed by Location 1 (26%), and then location 8 (22%). (P<0.05) CONCLUSION: Bacterial contamination of the drape was found to be common (70% of cases) after foot and ankle surgery. Contamination was more common on the superior and inferior aspects of the mini c-arm. This correlates to the area that is most commonly manipulated by the surgical team to position the arms of the c-arm. This information can be utilized by surgeons to identify and avoid high risk areas and, in high risk cases, they may consider changing gloves or re-draping the mini c-arm part way through the case. SAGE Publications 2020-11-06 /pmc/articles/PMC8705516/ http://dx.doi.org/10.1177/2473011420S00062 Text en © The Author(s) 2020 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Article Momenzadeh, Kaveh Williams, Caroline Czerwonka, Natalia Alemi, Mohammad Mehdi Kwon, John Y. Nazarian, Ara Miller, Christopher P. Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery |
title | Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery |
title_full | Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery |
title_fullStr | Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery |
title_full_unstemmed | Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery |
title_short | Contamination Profile of the Mini C-Arm Fluoroscopy Unit in Orthopedic Surgery |
title_sort | contamination profile of the mini c-arm fluoroscopy unit in orthopedic surgery |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8705516/ http://dx.doi.org/10.1177/2473011420S00062 |
work_keys_str_mv | AT momenzadehkaveh contaminationprofileoftheminicarmfluoroscopyunitinorthopedicsurgery AT williamscaroline contaminationprofileoftheminicarmfluoroscopyunitinorthopedicsurgery AT czerwonkanatalia contaminationprofileoftheminicarmfluoroscopyunitinorthopedicsurgery AT alemimohammadmehdi contaminationprofileoftheminicarmfluoroscopyunitinorthopedicsurgery AT kwonjohny contaminationprofileoftheminicarmfluoroscopyunitinorthopedicsurgery AT nazarianara contaminationprofileoftheminicarmfluoroscopyunitinorthopedicsurgery AT millerchristopherp contaminationprofileoftheminicarmfluoroscopyunitinorthopedicsurgery |