Cargando…
Urgent care study of the LumiraDx SARS-CoV-2 Ag Test for rapid diagnosis of COVID-19
BACKGROUND: Testing individuals suspected of severe acute respiratory syndrome–like coronavirus 2 (SARS-CoV-2) infection is essential to reduce the spread of disease. The purpose of this retrospective study was to determine the false negativity rate of the LumiraDx SARS-CoV-2 Ag Test when utilized f...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8709539/ https://www.ncbi.nlm.nih.gov/pubmed/34952653 http://dx.doi.org/10.1186/s41512-021-00113-7 |
Sumario: | BACKGROUND: Testing individuals suspected of severe acute respiratory syndrome–like coronavirus 2 (SARS-CoV-2) infection is essential to reduce the spread of disease. The purpose of this retrospective study was to determine the false negativity rate of the LumiraDx SARS-CoV-2 Ag Test when utilized for testing individuals suspected of SARS-CoV-2 infection. METHODS: Concurrent swab samples were collected from patients suspected of SARS-CoV-2 infection by their healthcare provider within two different urgent care centers located in Easton, MA, USA and East Bridgewater, MA, USA. One swab was tested using the LumiraDx SARS-CoV-2 Ag Test. Negative results in patients considered at moderate to high risk of SARS-CoV-2 infection were confirmed at a regional reference laboratory by polymerase chain reaction (PCR) using the additional swab sample. The data included in this study was collected retrospectively as an analysis of routine clinical practice. RESULTS: From October 19, 2020 to January 3, 2021, a total of 2241 tests were performed using the LumiraDx SARS-CoV-2 Ag Test, with 549 (24.5%) testing positive and 1692 (75.5%) testing negative. A subset (800) of the samples rendering a negative LumiraDx SARS-CoV-2 Ag Test was also tested using a PCR-based test for SARS-CoV-2. Of this subset, 770 (96.3%) tested negative, and 30 (3.8%) tested positive. Negative results obtained with the LumiraDx SARS-CoV-2 Ag test demonstrated 96.3% agreement with PCR-based tests (CI 95%, 94.7–97.4%). A cycle threshold (C(T)) was available for 17 of the 30 specimens that yielded discordant results, with an average C(T) value of 31.2, an SD of 3.0, and a range of 25.2–36.3. C(T) was > 30.0 in 11/17 specimens (64.7%). CONCLUSIONS: This study demonstrates that the LumiraDx SARS-CoV-2 Ag Test had a low false-negative rate of 3.8% when used in a community-based setting. |
---|