Verification of electron beam parameters in an intraoperative linear accelerator using dosimetric and radiobiological response methods

BACKGROUND: The availability of linear accelerators (linac) for research purposes is often limited and therefore alternative radiation sources are needed to conduct radiobiological research. The National Centre for Radiation Research in Poland recently developed an intraoperative mobile linac that e...

Descripción completa

Detalles Bibliográficos
Autores principales: Dróżdż, Agnieszka, Waluś, Martyna, Zieliński, Marcin, Malesa, Bożena, Kruszyna-Mochalska, Marta, Kulcenty, Katarzyna, Adamczyk, Beata, Nowaczyk, Piotr, Malicki, Julian, Pracz, Jacek
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Via Medica 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8726448/
https://www.ncbi.nlm.nih.gov/pubmed/34992877
http://dx.doi.org/10.5603/RPOR.a2021.0128
Descripción
Sumario:BACKGROUND: The availability of linear accelerators (linac) for research purposes is often limited and therefore alternative radiation sources are needed to conduct radiobiological research. The National Centre for Radiation Research in Poland recently developed an intraoperative mobile linac that enables electron irradiation at energies ranging from 4 to 12 MeV and dose rates of 5 or 10 Gy/min. The present study was conducted to evaluate the electron beam parameters of this intraoperative linac and to verify the set-up to evaluate out-of-field doses in a water phantom, which were determined through dosimetric and biological response measurements. MATERIALS AND METHODS: The distribution of radiation doses along and across the radiation beam were measured in a water phantom using a semiconductor detector and absolute doses using an ionisation chamber. Two luminal breast cancer cell lines (T-47D and HER2 positive SK-BR-3) were placed in the phantom to study radiation response at doses ranging from 2 to 10 Gy. Cell response was measured by clonogenic assays. RESULTS AND CONCLUSION: The electron beam properties, including depth doses and profiles, were within expected range for the stated energies. These results confirm the viability of this device and set-up as a source of megavoltage electrons to evaluate the radiobiological response of tumour cells.