The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR
BACKGROUND: Q.Clear is a block sequential regularized expectation maximization penalized-likelihood reconstruction algorithm for Positron Emission Tomography (PET). It has shown high potential in improving image reconstruction quality and quantification accuracy in PET/CT system. However, the evalua...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8748582/ https://www.ncbi.nlm.nih.gov/pubmed/35006411 http://dx.doi.org/10.1186/s40658-021-00428-w |
_version_ | 1784631035982512128 |
---|---|
author | Tian, Defeng Yang, Hongwei Li, Yan Cui, Bixiao Lu, Jie |
author_facet | Tian, Defeng Yang, Hongwei Li, Yan Cui, Bixiao Lu, Jie |
author_sort | Tian, Defeng |
collection | PubMed |
description | BACKGROUND: Q.Clear is a block sequential regularized expectation maximization penalized-likelihood reconstruction algorithm for Positron Emission Tomography (PET). It has shown high potential in improving image reconstruction quality and quantification accuracy in PET/CT system. However, the evaluation of Q.Clear in PET/MR system, especially for clinical applications, is still rare. This study aimed to evaluate the impact of Q.Clear on the (18)F-fluorodeoxyglucose (FDG) PET/MR system and to determine the optimal penalization factor β for clinical use. METHODS: A PET National Electrical Manufacturers Association/ International Electrotechnical Commission (NEMA/IEC) phantom was scanned on GE SIGNA PET/MR, based on NEMA NU 2-2012 standard. Metrics including contrast recovery (CR), background variability (BV), signal-to-noise ratio (SNR) and spatial resolution were evaluated for phantom data. For clinical data, lesion SNR, signal to background ratio (SBR), noise level and visual scores were evaluated. PET images reconstructed from OSEM + TOF and Q.Clear were visually compared and statistically analyzed, where OSEM + TOF adopted point spread function as default procedure, and Q.Clear used different β values of 100, 200, 300, 400, 500, 800, 1100 and 1400. RESULTS: For phantom data, as β value increased, CR and BV of all sizes of spheres decreased in general; images reconstructed from Q.Clear reached the peak SNR with β value of 400 and generally had better resolution than those from OSEM + TOF. For clinical data, compared with OSEM + TOF, Q.Clear with β value of 400 achieved 138% increment in median SNR (from 58.8 to 166.0), 59% increment in median SBR (from 4.2 to 6.8) and 38% decrement in median noise level (from 0.14 to 0.09). Based on visual assessment from two physicians, Q.Clear with β values ranging from 200 to 400 consistently achieved higher scores than OSEM + TOF, where β value of 400 was considered optimal. CONCLUSIONS: The present study indicated that, on (18)F-FDG PET/MR, Q.Clear reconstruction improved the image quality compared to OSEM + TOF. β value of 400 was optimal for Q.Clear reconstruction. |
format | Online Article Text |
id | pubmed-8748582 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-87485822022-01-20 The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR Tian, Defeng Yang, Hongwei Li, Yan Cui, Bixiao Lu, Jie EJNMMI Phys Original Research BACKGROUND: Q.Clear is a block sequential regularized expectation maximization penalized-likelihood reconstruction algorithm for Positron Emission Tomography (PET). It has shown high potential in improving image reconstruction quality and quantification accuracy in PET/CT system. However, the evaluation of Q.Clear in PET/MR system, especially for clinical applications, is still rare. This study aimed to evaluate the impact of Q.Clear on the (18)F-fluorodeoxyglucose (FDG) PET/MR system and to determine the optimal penalization factor β for clinical use. METHODS: A PET National Electrical Manufacturers Association/ International Electrotechnical Commission (NEMA/IEC) phantom was scanned on GE SIGNA PET/MR, based on NEMA NU 2-2012 standard. Metrics including contrast recovery (CR), background variability (BV), signal-to-noise ratio (SNR) and spatial resolution were evaluated for phantom data. For clinical data, lesion SNR, signal to background ratio (SBR), noise level and visual scores were evaluated. PET images reconstructed from OSEM + TOF and Q.Clear were visually compared and statistically analyzed, where OSEM + TOF adopted point spread function as default procedure, and Q.Clear used different β values of 100, 200, 300, 400, 500, 800, 1100 and 1400. RESULTS: For phantom data, as β value increased, CR and BV of all sizes of spheres decreased in general; images reconstructed from Q.Clear reached the peak SNR with β value of 400 and generally had better resolution than those from OSEM + TOF. For clinical data, compared with OSEM + TOF, Q.Clear with β value of 400 achieved 138% increment in median SNR (from 58.8 to 166.0), 59% increment in median SBR (from 4.2 to 6.8) and 38% decrement in median noise level (from 0.14 to 0.09). Based on visual assessment from two physicians, Q.Clear with β values ranging from 200 to 400 consistently achieved higher scores than OSEM + TOF, where β value of 400 was considered optimal. CONCLUSIONS: The present study indicated that, on (18)F-FDG PET/MR, Q.Clear reconstruction improved the image quality compared to OSEM + TOF. β value of 400 was optimal for Q.Clear reconstruction. Springer International Publishing 2022-01-10 /pmc/articles/PMC8748582/ /pubmed/35006411 http://dx.doi.org/10.1186/s40658-021-00428-w Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Research Tian, Defeng Yang, Hongwei Li, Yan Cui, Bixiao Lu, Jie The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR |
title | The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR |
title_full | The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR |
title_fullStr | The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR |
title_full_unstemmed | The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR |
title_short | The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR |
title_sort | effect of q.clear reconstruction on quantification and spatial resolution of 18f-fdg pet in simultaneous pet/mr |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8748582/ https://www.ncbi.nlm.nih.gov/pubmed/35006411 http://dx.doi.org/10.1186/s40658-021-00428-w |
work_keys_str_mv | AT tiandefeng theeffectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT yanghongwei theeffectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT liyan theeffectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT cuibixiao theeffectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT lujie theeffectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT tiandefeng effectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT yanghongwei effectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT liyan effectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT cuibixiao effectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr AT lujie effectofqclearreconstructiononquantificationandspatialresolutionof18ffdgpetinsimultaneouspetmr |