A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods

INTRODUCTION: Here we report estimates of glomerular basement membrane (GBM) thickness in the Brazilian population performed using direct (DM) and orthogonal interception methods (OIM), and comment on potential sources of variation among estimates made by different laboratories. METHODOLOGY: A total...

Descripción completa

Detalles Bibliográficos
Autores principales: Viana, Débora Leal, Alladagbin, Dona Jeanne, dos-Santos, Washington L. C., Figueira, Claudio Pereira
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8750857/
https://www.ncbi.nlm.nih.gov/pubmed/35012461
http://dx.doi.org/10.1186/s12882-021-02634-1
_version_ 1784631555572891648
author Viana, Débora Leal
Alladagbin, Dona Jeanne
dos-Santos, Washington L. C.
Figueira, Claudio Pereira
author_facet Viana, Débora Leal
Alladagbin, Dona Jeanne
dos-Santos, Washington L. C.
Figueira, Claudio Pereira
author_sort Viana, Débora Leal
collection PubMed
description INTRODUCTION: Here we report estimates of glomerular basement membrane (GBM) thickness in the Brazilian population performed using direct (DM) and orthogonal interception methods (OIM), and comment on potential sources of variation among estimates made by different laboratories. METHODOLOGY: A total of 38 patients, ranging from 3 to 78 years of age, 26 (68%) males and 12 (32%) females, were submitted to kidney biopsy procedures for renal disease diagnosis. Glomeruli were diagnosed with minor histological changes by conventional, immunofluorescence and electron microscopy. GBM thickness was estimated using both DM and OIM methods. RESULTS: Estimates of GBM thickness obtained using DM were higher than those obtained by OIM. However, the application of a correction for non-perpendicular membrane sectioning to DM estimates yielded similar results to those obtained under OIM. The estimated GMB thickness using DM after correction was 289 + 44 nm, versus 287 + 48 nm by OIM. No statistically significant differences were detected in GMB thickness, nor with respect to patient age or sex. CONCLUSIONS: GBM thickness in the studied Brazilian population measured approximately 290 nm. The application of criteria for estimating the shortest distance between the endothelial and podocyte cell membranes with correction for non-perpendicular membrane sectioning can increase the accuracy of GBM thickness estimates using DM and OIM. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12882-021-02634-1.
format Online
Article
Text
id pubmed-8750857
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-87508572022-01-11 A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods Viana, Débora Leal Alladagbin, Dona Jeanne dos-Santos, Washington L. C. Figueira, Claudio Pereira BMC Nephrol Research INTRODUCTION: Here we report estimates of glomerular basement membrane (GBM) thickness in the Brazilian population performed using direct (DM) and orthogonal interception methods (OIM), and comment on potential sources of variation among estimates made by different laboratories. METHODOLOGY: A total of 38 patients, ranging from 3 to 78 years of age, 26 (68%) males and 12 (32%) females, were submitted to kidney biopsy procedures for renal disease diagnosis. Glomeruli were diagnosed with minor histological changes by conventional, immunofluorescence and electron microscopy. GBM thickness was estimated using both DM and OIM methods. RESULTS: Estimates of GBM thickness obtained using DM were higher than those obtained by OIM. However, the application of a correction for non-perpendicular membrane sectioning to DM estimates yielded similar results to those obtained under OIM. The estimated GMB thickness using DM after correction was 289 + 44 nm, versus 287 + 48 nm by OIM. No statistically significant differences were detected in GMB thickness, nor with respect to patient age or sex. CONCLUSIONS: GBM thickness in the studied Brazilian population measured approximately 290 nm. The application of criteria for estimating the shortest distance between the endothelial and podocyte cell membranes with correction for non-perpendicular membrane sectioning can increase the accuracy of GBM thickness estimates using DM and OIM. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12882-021-02634-1. BioMed Central 2022-01-10 /pmc/articles/PMC8750857/ /pubmed/35012461 http://dx.doi.org/10.1186/s12882-021-02634-1 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Viana, Débora Leal
Alladagbin, Dona Jeanne
dos-Santos, Washington L. C.
Figueira, Claudio Pereira
A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods
title A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods
title_full A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods
title_fullStr A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods
title_full_unstemmed A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods
title_short A comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods
title_sort comparative study of human glomerular basement membrane thickness using direct measurement and orthogonal intercept methods
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8750857/
https://www.ncbi.nlm.nih.gov/pubmed/35012461
http://dx.doi.org/10.1186/s12882-021-02634-1
work_keys_str_mv AT vianadeboraleal acomparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods
AT alladagbindonajeanne acomparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods
AT dossantoswashingtonlc acomparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods
AT figueiraclaudiopereira acomparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods
AT vianadeboraleal comparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods
AT alladagbindonajeanne comparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods
AT dossantoswashingtonlc comparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods
AT figueiraclaudiopereira comparativestudyofhumanglomerularbasementmembranethicknessusingdirectmeasurementandorthogonalinterceptmethods