Cargando…
Development of a high-throughput differential mobility separation–tandem mass spectrometry (DMS-MS/MS) method for clinical urine drug testing
INTRODUCTION: Differential mobility separation (DMS) is an analytical technique used for rapid separation of ions and isomers based on gas phase mobility prior to entering a mass spectrometer for analysis. The entire DMS process is accomplished in fewer than 20 ms and can be used as a rapid alternat...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8753179/ https://www.ncbi.nlm.nih.gov/pubmed/35036987 http://dx.doi.org/10.1016/j.jmsacl.2021.12.008 |
Sumario: | INTRODUCTION: Differential mobility separation (DMS) is an analytical technique used for rapid separation of ions and isomers based on gas phase mobility prior to entering a mass spectrometer for analysis. The entire DMS process is accomplished in fewer than 20 ms and can be used as a rapid alternative to chromatographic separation. OBJECTIVE: The primary objective was to evaluate the utility of DMS-tandem mass spectrometry (DMS-MS/MS) as a replacement for immunoassay-based clinical toxicology testing. METHODS: A sensitive DMS-MS/MS method was developed and validated for simultaneous identification of 33 drugs and metabolites in human urine samples. After DMS optimization, the method was validated and used to screen 56 clinical urine samples. These results were compared to results obtained by immunoassay. RESULTS: The DMS-MS/MS method achieved limits of detection ranging from 5 to 100 ng/mL. Moreover, the total analysis time was 2 min per sample. For the method performance evaluation, DMS-MS/MS results were compared with previously obtained urine toxicology immunoassay results. DMS-MS/MS showed higher sensitivity and identified 20% more drugs in urine, which were confirmed by LC-MS/MS. CONCLUSION: The DMS-MS/MS as applied in our lab demonstrated the capability for rapid drug screening and provided better analytical performance than immunoassay. |
---|