Corneal Biomechanics for Ocular Hypertension, Primary Open-Angle Glaucoma, and Amyloidotic Glaucoma: A Comparative Study by Corvis ST

BACKGROUND: To evaluate biomechanical parameters of the cornea provided by Corvis ST in patients with ocular hypertension, primary open-angle glaucoma, and amyloidotic glaucoma and to compare with healthy controls. METHODS: This was a cross-sectional study of patients with ocular hypertension, prima...

Descripción completa

Detalles Bibliográficos
Autores principales: Silva, Nisa, Ferreira, André, Baptista, Pedro Manuel, Figueiredo, Ana, Reis, Rita, Sampaio, Isabel, Beirão, João, Vinciguerra, Riccardo, Menéres, Pedro, Menéres, Maria João
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8754459/
https://www.ncbi.nlm.nih.gov/pubmed/35035215
http://dx.doi.org/10.2147/OPTH.S350029
Descripción
Sumario:BACKGROUND: To evaluate biomechanical parameters of the cornea provided by Corvis ST in patients with ocular hypertension, primary open-angle glaucoma, and amyloidotic glaucoma and to compare with healthy controls. METHODS: This was a cross-sectional study of patients with ocular hypertension, primary open-angle glaucoma, and amyloidotic glaucoma that underwent Corvis ST imaging. Primary outcome was the comparison of corneal biomechanical parameters between study groups after adjusting for age, gender, Goldmann intraocular pressure (GAT-IOP), and prostaglandin analogues medication. Secondary outcome was the comparison of different IOP measurements in each group. RESULTS: One hundred and eighty-three eyes from 115 patients were included: 61 with primary open-angle glaucoma, 32 with amyloidotic glaucoma, 37 with ocular hypertension and 53 were healthy controls. Amyloidotic glaucoma group had smaller radius (p=0.025), lower deflection amplitude at highest concavity (p=0.019), and higher integrated radius (p=0.014) than controls. Ocular hypertension group had higher stiffness parameter at first applanation (p=0.043) than those with primary open-angle glaucoma, and higher stress–strain index (p=0.049) than those with amyloidotic glaucoma. Biomechanically corrected intraocular pressure was significantly lower than Goldmann intraocular pressure in group with primary open-angle glaucoma (p=0.005) and control group (p=0.013), and Goldmann intraocular pressure adjusted for pachymetry in group with primary open-angle glaucoma (p=0.01). CONCLUSION: Eyes with amyloidotic glaucoma have more deformable corneas, while eyes with ocular hypertension have less deformable corneas. These findings may be linked to the susceptibility to glaucomatous damage and progression. There were significant differences between Goldmann applanation tonometry and biomechanically corrected intraocular ocular pressure provided by Corvis ST.